Rapid Analysis of Immuno-suppressants Using Triple Quadrupole LC-MS/MS
This report exemplifies an ultra high speed analysis of 4 immuno-supressants (Tacrolimus, Everolimus, Rapamycin, Cyclosporin A) in less than 120 seconds by Shimadzu UFMS Triple Quadrupole Mass Spectrometer LCMS-8050 with negative ionization mode. Combination of Nexera X2 and LCMS-8050 provides you much faster run time than ever you experienced without sacrificing the quality of results.
Including two internal standrads such as Ascomycin and Cyclosporin D, total 6 compounds were measured in 1.8 minutes. The structure of each compounds is illustrated below. In spite of low injection volume (1.5 uL) was utilized, this ultra high speed method achieves 1 ng / mL as the limit of quantitation for all compounds.
- Content Type:
- Application
- Document Number:
- LAAN-J-LM-E010
- Product Type:
- Liquid Chromatograph-Mass Spectrometry, Mass Spectrometry
- Keywords:
- Immuno-suppressants, Tacrolimus, Everolimus, Rapamycin, Cyclosporin A, Clinical research, Forensics, Pharmaceutical, Life Science, LCMS-8050
- Language:
- English
- File Name:
- jpo114032.pdf
- File Size:
- 608kb