NOA-7100

Transportable Gas Analyzer

Specifications

Standard type

Two standard types are available
  Type 1 Type 2
Measurement components NOx, O2 NOx, NO, NO2 (*1)
Measurement range NOx: 
0-25/ 50/ 100/ 250/ 500/ 1000/ 2500/4000 ppm
O2: 0-5/ 10/ 25 vol%
NOx, NO, NO2
0-100/ 250/ 500/ 1000/ 2500/ 4000 ppm
Measurement principle NOx: Atmospheric pressure chemiluminescence method; O2: Limiting current zirconia type method
Repeatability Within ±0.5% of full scale
Zero drift Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day)
Span drift Within ±1.5% of full scale per day
Linearity Within ±2% of full scale
Response time (*2) NOx: Approx. 20/30/60 sec
       (Switching is configurable.)
O2: Approx. 30/60 sec
       (Switching is configurable.)
NOx/NO: Within 90 sec
Gas sample amount Approx. 2 L/min 100 mL/min
Display Liquid crystal display, able to display 3 items at a time
Transmission output 0-1VDC, 3 channels, insulated output (individual channels not insulated)
Wireless data transmission Equipped
Data output to external media (*3) Measurement data stored in the internal memory of the analyzer can be written to USB flash drives in CSV file format; can also be output to a computer via Wi-Fi or LAN cable connection.
Permissible ambient temperature (*4) 5-40°C; must not be exposed to direct sunlight or radiant heat
Permissible ambient humidity (*4). Relative humidity 90%RH or less
Power requirements

Choose from 110, 115, 120, 127, 220, 230, 240 VAC 50-60 Hz,

100 V series: 410 VA, 200 V series: 700 VA (steady 200 VA)

Dimensions W260 x H452 x D420 mm (does not include protrusions)
Weight (analyzer) Approx. 16 kg
External drain separator Fitted Not fitted
Sample gas conditions (*5)

[ At NOA-7100 inlet port ] Pressure: -980 to + 980 Pa;

Temperature: Ambient temperature; Dust: 0.1 g/Nm3 or less

External signal input 0-1VDC, 1 channel, Signal input from an external analyzer, display (Data can be saved)

 

Special specifications

 

In addition to the standard model, it is possible to specify constituents, and if the small flow rate set, communication set, or NOx-NO switching solenoid valve set is provided.
Measurement
components
1 constituent (NOx) or 2 constituents (NOx, O2) (*5)
O2 cannot be measured if small flow rate set is selected. See small flow rate set page.
NO and NO2 can be measured as well as NOx if NOx-NO switching solenoid valve set is selected.
See NOx-NO switching solenoid valve set page.
Measurement
range
NOx: 0-25/50/100/250/500/1000/2500/4000 ppm
O2: 0-5/10/25vol%
NOx measurement range differs if small flow rate set is selected.
See small flow rate set page.
Small flow rate set Not fitted Fitted
Response time (*2) NOx: Approx. 20/30/60 sec
       (Switching is configurable.)
O2: Approx. 30/60 sec
       (Switching is configurable.)
NOx/NO: Within 90 sec
Gas sample amount Approx. 2 L/min 100 mL/min
Repeatability Within ±0.5% of full scale
Span drift Within ±1.5% of full scale
Measurement
components
NOx or NOx/O2 NOx
NO and NO2 (*1) can be measured as well as NOx if NOx-NO switching solenoid valve set is selected.
Measurement
range
NOx: 0-25/ 50/ 100/ 250/ 500/ 1000/ 2500/ 4000 ppm
O2: 0-5/ 10/ 25 vol%
NOx: 0-100/ 250/ 500/ 1000/ 2500/ 4000 ppm
Communication set Not fitted Fitted
Wireless data transmission Not equipped Equipped
Data output to
external media (*3)
Not equipped Measurement data stored in the internal memory of the analyzer can be output to USB flash drives in CSV file format; can also be output to computer via Wi-Fi or LAN cable connection.
NOx-NO switching solenoid valve set Not fitted Fitted
Measurement components NOx NOx, NO, NO2 (*1)
This special accessory switches the flow path inside the analyzer between a path that flows through the converter (NOx flow path) and one that does not flow through the converter (NO path). The switch cycle is selectable from auto (configurable from 1 to 99 minutes), manual, and fixed. If there is significant variation over time in concentrations of NOx, NO, or NO2 relative to the switching cycle, NO2 concentration measurement accuracy diminishes.
Basic specifications are as follows.

Basic specifications

Measurement principle NOx: Atmospheric pressure chemiluminescence method
O2: Limiting current type zirconia method
Repeatability Within ±0.5% of full scale
Zero drift Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day)
Span drift Within ±1.5% of full scale
Linearity Within ±2% of full scale
 
The following specifications are the same as for the standard model.
Display, transmission output, permissible ambient temperature, permissible ambient humidity, power requirements, dimensions, weight (body), sample gas conditions, external signal input.
*1: NO2 is measured by calculation (NOx-NO).
*2: Time that indicates 90% response from gas flowing from sample gas inlet (flowing gas switching time + electrical system response time). With Type 2, this will be the response when NOx-NO switching is fixed.
*3: Check the Shimadzu website for smart devices and USB flash drives for which operation has been confirmed.
*4: Permissible ambient temperature and humidity are indicated by the shaded part of the figure.
*5: Sample gas conditions and coexisting gases are summarized below. Please contact your Shimadzu representative if the conditions are not satisfied.
 
 

For Research Use Only. Not for use in diagnostic procedures.