Product

Laboratory Instruments

Gas Chromatograph

Natural Gas Systems

Natural Gas Systems

Shimadzu's Natural Gas Systems are specially-modified GC-8As designed for the measurement of natural gas components. Determining the composition of natural gas is important because its heating value (and price) is determined by its BTU per cubic foot.

Shimadzu's two systems conform to ASTM D1945 and D3588. Natural Gas System #1 separates C1 - C5 paraffins and isoparaffins, O2, N2, CO and C6+ by backflush. System #2 analyzes the same components, but it produces a single air peak rather than separating O2 and N2. After analysis, the mole% of each component is automatically calculated, along with the BTUs per cubic foot, compressibility, specific gravity, and relative density.

Each system is reliable and easy to operate. The PRG-102A event programmer switches the valves automatically, while the optional Chromatopac data processor controls the system operation and performs the required calculations automatically. Calibration is based on a Certified Calorific Natural Gas Standard.

System #1

Gas Chromatography GC-8AIT
Detector TCD for detection of C1-C5 Paraffins and Isoparaffins, O2, N2, CO2, and C6+ by backflush.
Valves One automated, air actuated, 10-port valve
Two automated, air actuated 6-port valves
Columns 6 stainless steel columns
Matrix Natural Gas
Minimum sample volume 1 mL
Requires PRG-102A, PC-17N

System #2

Gas Chromatography GC-8AIT
Detector TCD for detection of C1-C5 Paraffins and Isoparaffins, Air, CO2, and C6+ by backflush.
Valve One automated, air actuated, 10-port valve
Columns 4 stainless steel columns
Carrier Gas Helium, (99.999% purity or better)
Requires PRG-102A, PC-17N