Product

Laboratory Instruments

Mass Spectrometer Systems

Gas Chromatograph-Mass Spectrometers (GC/MS)

GCMS-QP2010 Plus

Specifications

[1] Ion source
Ionization modes EI (standard); CI, NCI (options)
Ionization voltage 10 eV to 200 eV
Emission current 5 µA to 250µA
Filament Dual rhenium coil type
Ion source temperature 100 °C to 300 °C
[2] Mass Spectrometer
Mass range m/z 1.5 to m/z 1090
Mass filter High accuracy quadrupole with pre-rods
SIM mode 64 channels x 128 groups or FASST
Resolution R>2M (FWHM)
Stability ±0.1 u/48 hours (constant temperature)
MS interface 100 °C to 350 °C
[3] Gas chromatograph
GC unit Shimadzu GC-2010
[4] Detector
Detector unit Secondary electron multiplier tube
(-10kV conversion dynode, with overdrive lens)
Detector ACTIVE FILM Multiplier®
Dynamic range 8 x 106 (electrical)
[5] Vacuum System
Main pump Dual inlet differential vacuum system
(179 L/sec + 185 L/sec, He)
Backup pump 30 L/min (60Hz) oil rotation pump
Vacuum gauges Pirani gauge, ion gauge
Column flow He, 15 mL/min maximum
[6] Scan speed
Maximum scan speed 10,000 u/sec
From m/z 50 to m/z 900, measurement at 0.1 sec intervals in 0.1 u steps Minimum interval 20 ms (maximum 50 scans per second possible)
[7] DI Specifications (Option)
Temperature measurement range Ambient to 500 °C
Heating rate 5, 10, 20, 40, 80 °C/min. (at maximum output)
Program Temperature program up to 5 steps
Sample vial size 1.9 mm I.D. x 8 mm depth
[8] Installation conditions (See installation Requirements Manual for details)
Power requirements Frequency 50 / 60 Hz
GC AC single phase 110V,115V 1800 VA
AC single phase 230V 2600 VA(high power oven)
MS AC single phase 110V to 240V 1000VA
Voltage regulation operational power voltage range
±10%

specified power voltage range
±5%
Set-up environment Temperature constant temperature between 18 to 28 °C
Humidity 40 to 70% RH (no condensation)

Specifications described above are subject to change without notice.

Example Installation

(PC and printer dimensions will differ according to model.)
The following gives cable and piping lengths:

Example Installation

MS rotary-pump
vacuum piping:
approx. 1.5 m
Power cable : approx. 1.5 m
PC-MS cable: approx. 2 m
Rotary-pump
evacuation piping:
approx. 4 m