Following their introduction in 1995, the world's highest performance, easy-to-use Shimadzu UV-2401/2501 spectrophotometers are now even more powerful with the addition of the world's most advanced software. These spectrophotometers have made their entrance to greet the new century.
The UV-2550, with its DDM (double blazed grating, double monochromator) design, provides a high energy throughput optical system with ultra low stray light. The UV-2450, the single monochromator version of the UV-2550, achieves 0.015% stray light even with its simpler optical system.
Both instruments offer outstanding cost performance in their ability to provide UV-VIS spectral analysis of wide ranging samples, including organic and inorganic compounds, biological samples like DNA and enzymes, as well as special measurements on optical materials.
Operating under the Windows XP Pro/2000 Pro environment, the new generation UV Probe software is provided as standard. UVProbe incorporates the latest multitasking and report generation functionality, including network support, enabling measurement assistance among colleagues and effective presentation of precious data.
The 8/16 micromulti-cell permits selection among 4 possible combinations with sample capacities of 50mL and 100mL. And, either the normal type or the constant-temperature controlled type cell holder may be selected.
(NOTE) Both the cell holder and the multi-cell are optional accessories.
|
Item
|
Description
|
||||
|
Setting wavelength range
|
190 ~ 1100nm
|
||||
|
Measurement wavelength range
|
190 ~ 900nm
(up to 1100nm with special detector)
|
||||
|
Wavelength accuracy
|
Ẕ0.3nm
with auto wavelength correction included
|
||||
|
Wavelength repeatability
|
Ẕ0.1nm
|
||||
|
Wavelength scanning speed
|
Wavelength slew rate: about 3200nm/min
Wavelength scan rate: about 900 ~ 160nm/min Monitor scan rate: about 2500nm/min |
||||
|
Wavelength setting
|
At 1nm units for scan start and scan
end wavelengths, and 0.1nm units for other wavelengths
|
||||
|
Lamp interchange wavelength
|
Auto switching synchronized with wavelength,
switching range selectable between 282 ~
393nm (0.1nm units)
|
||||
|
Spectral bandwidth
|
6-step switching among 0.1/0.2/0.5/1/2/5nm
|
||||
|
Response
|
Optimum response speed automatically
set depending on bandwidth, minimum 0.1sec
|
||||
|
Resolution
|
0.1nm
|
||||
|
Stray light
|
UV-2450
|
UV-2550
|
|||
|
Less than 0.015%
Less than 0.0003% (220nm, NaI 10g/L solution) Less than 0.015% Less than 0.0003% (340nm, UV-39 filter) |
|||||
|
Photometric system
|
Double-beam, direct ratio system with
dynode feedback
|
||||
|
Photometric modes
|
Absorbance (Abs.), transmittance (%),
reflectance (%), energy (E)
|
||||
| Photometric range | Absorbance: -4 ~
5 Abs Transmittance, reflectance: 0.0 ~ 999.9% |
||||
| Recording range | Absorbance: -9.999 ~
9.999 Abs Transmittance, reflectance: -999.9 ~ 999.9% |
||||
| Photometric accuracy |
|
||||
* If the detector has been replaced with a near infrared sensitive photomultiplier, the standard optical specifications of the above instrument will not be satisfied.
|
Item
|
Description
|
|
|
Photometric repeatability
|
Ẕ0.001
Abs (0 ~ 0.5 Abs)
Ẕ0.002 Abs (0.5 ~ 1.0 Abs) Ẕ0.1%T |
|
|
Baseline flatness
|
Ẕ0.001
Abs (excluding noise, using 2nm slit, and slow wavelength
scanning speed)
|
|
| Baseline correction | Auto correction using PC (Stored baseline is automatically loaded when power is switched on, re-correction is possible) | |
| Drift | 0.0004Abs/h (after power is on for 2 hours) | |
| Temperature and humidity requirements | 15 ~ 35°C, 45 ~ 80% (no condensation, less than 70% above 30°C) | |
| Light source | 50W halogen lamp (2,000 hours life), deuterium lamp (socket type), light source auto position adjustment built in | |
| Monochromator | UV-2450 | |
| Single monochromator, high-performance blazed holographic grating in aberration-corrected Czerny-Turner mounting | ||
| UV-2550 | ||
| Grating/Grating type double monochromator, Pre-monochromator: double-blazed holographic grating Main monochromator: high-performance blazed holographic grating in aberration-corrected Czerny-Turner mounting | ||
| Detector | Photomultiplier R-928 | |
| Sample compartment | Internal dimensions: 150W x 260D x 120H
(mm) Distance between light beams: 100mm Maximum light path length of cell: 100 mm |
|
| Power requirements | AC100, 120, 220, 240 V, switch selectable 50/60Hz; 250 VA |
|
| Dimensions | 570W x 660D x 275H (mm) | |
| Weight | About 36 kg | |
(PC and printer are not included.)
| Operating System | Windows XP Pro/2000 Pro |
| Instruments | UV-1601 Series, UV-2401/2501 Series |
| Data Acquisition Modes | Spectrum, Kinetics and Photometric |
| General |
|
| Spectrum Mode |
|
| Data Processing in Spectrum Mode |
|
| Photometric (Quantitation) Mode |
|
| Kinetics (Time Course) Mode |
|
| Report Generator |
|