Shimadzu was the first company to offer dynamic alignment as a standard feature in an affordable FTIR spectrometer.With state-of-the-art technology,the Shimadzu dynamic alignment system continuously maintains optimal alignment during data acquisition,ensuring consistently reproducible spectra without the need for tedious mechanical adjustments.
FTIR-8400S is combined with the IRsolution - a 32 bit high performance FTIR software - to analyze your samples easily and securely.
Highly precise control of the moving mirror is required to stabilize the interferogram of the FTIR instrument. The FTIR-8400S includes the patented FJS system - very smooth and precise moving mirror unit - and the Dynamic Alignment system to optimize and stabilize the inter-ferometer unit. FTIR-8400S requires only a short stabilization time and is very secure.
This System monitors the interferometer condition of He-Ne laser and compares it with that of optimized conditions. Any detected discrepancies by DSP system are corrected by automatic alignment of the piezo actuators at the fixed mirror.
Optimal interferometer alignment is maintained automatically for continuous optimization and exceptional reproducibility.
Peak-to-Peak S/N ratio is 20,000:1 or better, guaranteed!
New FTIR operators will love the quick-start function that prompts the user through every action from setting scan parameters and acquiring the spectrum, to detecting peaks and printing.
Powerful IRsolution software offers a myriad of standard data processing functions.The management of User group, protection by user name/password, Operation log recording, and Electric Signature supports the FDA 21 CFR Part 11.note)
The software validates instrument performance to ensure compliance based on Japanese/European Pharmacopoeia and ASTM. The instrument and data reliability are ensured by GLP/GMP and FDA 21 CFR Part 11 compliance.
note) IRsolution Agent software is needed to fully comply to FDA 21 CFR Part 11 regulation.
| Interferometer | Michelson type with 30° incident angle,dynamic alignment,sealed and desiccated | |
|---|---|---|
| Optical system | Single beam optics | |
| Beam splitter | Germanium-coated KBr plate | |
| Light source | High brightness ceramic | |
| Detector | Temperature controlled high sensitivity detector(DLATGS detector) | |
| S/N ratio | Greater than 20,000:1(KRS-5 window,4cm-1,1 minute,2200cm-1,P-P) | |
| Wavenumber range | Greater than 20,000:1(KRS-5 window,4cm-1,1 minute,2200cm-1,P-P) | |
| Resolution | 0.85,1,2,4,8,16cm-1 | |
| Mirror speed | 3 steps;2.8,5,9mm/sec | |
| Data sampling | He-Ne laser | |
| Sample compartment | W200mm x D230mm x H170mm | |
| Data processing functions | Arithmetic operation between spectrum and constant,Arithmetic operation among spectra,Spectral subtraction,Peak detection,Smoothing,Derivative,Baseline correction,Data correction,Normalization,Kubelka-Munk conversion,Kramers-Kronig analysis,ATR correction | |
| Optional programs | PLS quantitation,Curve fitting,Mapping,Macro platform | |
| Validation programs | Compliance to Japanese/European Pharmacopoeia and ASTM | |
| GLP/GMP | Operation log recording,Software protection by user name/password, Management of user level | |
| FDA 21 CFR Part 11(note 1) | Electronic signature,Data manipulation history recording,Saving of measured interferogram/background spectrum/manipulated spectra | |
| Ambient conditions | 15-30°C Less than 70% |
|
| Power requirement | AC100/120/220/230/240V 50/60Hz 160VA(note2) | |
(note1)IRsolution Agent software is needed to fully comply to FDA 21 CFR part11.
(note2)PC requires additional power.