
Sensitivity
High sensitivity and expanded functionality for leading edge Research & Development.
Easy software and Automatic Accessory Recognition for User Friendly operation.
ADA (Advanced Dynamic Alignment) for optical system reliability
Status Monitoring and Validation Program
Near IR/Middle IR/Far IR
Wide variety of Accessories for a large range of applications
Powerful software - IRsolution - to enhance performance of IRPrestige-21.
Considering the environment, the electric power consumption while measuring is 25% reduced(NOTE 3).
IRPrestige-21 and IRsolution opens a new world for FTIR users.
(NOTE 1) 4cm-1 resolution, 1 minute accumulation, around 2,100cm-1, peak to peak
(NOTE 2) IRsolution Agent software is needed to fully comply to FDA 21 CFR Part 11 regulation.
(NOTE 3) Comparison with the previous Shimadzu FTIR.
| Interferometer | Michelson interferometer (30 degree incident angle) Advanced Dynamic alignment system Sealed and desiccated interferometer with an automatic dryer |
|
|---|---|---|
| Optical system | Single beam optics | |
| Beam splitter | Germanium-coated KBr plate for Middle IR (Standard) Germanium-coated CsI plate for Middle/Far IR (Optional) Silicon-coated CaF2 plate for Near IR (Optional) |
|
| Light source | Air Cooled Ceramic for Middle/Far IR with 3 years guarantee (Standard) Tungsten lamp for Near IR (Optional) |
|
| Detector | DLATGS detector with Temperature Control for Middle/Far IR (Standard) MCT(Hg-Cd-Te) detector with Liquid Nitrgen Cooling for Middle IR(Optional) InGaAs detector for Near IR (Optional) |
|
| Wavenumber range | 7,800 - 350 cm-1 12,500 - 240 cm-1 (Optional. See Figure for detail) |
|
| Resolution | 0.5 cm-1, 1 cm-1, 2 cm-1, 4 cm-1, 8 cm-1, 16 cm-1 (Middle/Far IR) 2 cm-1, 4 cm-1, 8 cm-1, 16 cm-1 (Near IR) |
|
| S/N ratio | 40,000: 1 and higher (Peak-to-peak, 4 cm-1 resolution, around. 2100 cm-1, 1 minute scan) |
|
| Mirror speed | 3-step selection of 2.8, 5, or 9 mm/sec Scanning at 4 cm-1 takes from 2-3 sec |
|
| Data sampling | He-Ne laser with 30 months guarantee | |
| Gain control | Automatic or manual from x1 - x128 | |
| Sample compartment | Automatic Accessory Recognition 200 (W) x 230 (L) x 170 (H) mm Center Focus |
|
| Dimensions | 600 (W) x 680 (L) x 290 (H) mm | |
| Weight | 54Kg | |
| Light Source | Beam Splitter | Detector | Scan Range(cm-1) | Necessary Parts |
| Tungsten | CaF2 | InGaAs | 12,500-3,800 | NIR Kit (206-72015-91) |
| Ceramic | KBr | DLATGS | 7,800-350 | Standard |
| MCT | 5,000-720 | MCT Kit (206-72017-91) | ||
| CsI | DLATGS | 5,000-240 | FIR Kit (206-72016-91) |
(Caution) The beam splitters must be kept inside of IRPrestige-21 or a desiccator.
The orange bar expresses the wavenumber range which can be measured with the standard elements. The red, yellow and brown bars express the wavenumber range which can be measured with the optional elements.