| Photometric system |
Wavelength range |
185 to 900nm |
| Mounting |
Aberration-corrected Czerny-Turner mounting, Number of grating grooves: 1800lines/mm, Forcal length: 298mm |
| Bandwidth |
0.2 0.7 0.7 (Low) 2.0 (Low) nm (4 step automatic switching) |
| Detector |
Photomultiplier (short wavelengths), semiconductor (long wavelengths) |
| No. of lamps mounted |
6 lamps, 2 user-selected lamps can be illuminated at the same time
(1 for measurement,1warming up for next measurement) |
| Photometric method |
Flame: Optical double beam. Furnace: Electrical double beam |
| Background correction |
High-speed self-reversal method(BGC-SR), high-speed deuterium lamp method(BGC-D2) |
| Lamp mode |
EMISSION, NON-BGC, BGC-SR, BGC-D2, |
| Data processing |
Software requirments |
MS-WindowsTM2000 |
| Parameter settings |
Wizard method |
| Measurement mode |
Flame suction method, flame microsampling method, furnace method |
| Concentration conversion mode |
Calibration curve method(can be selected from 1st, 2nd, and 3rd orders)
Standard addition method(1st order), simple standard addition method(1st order)
|
| Repeat analysis |
Max. 20 times Display of mean values,deviation(SD), coefficient of variation(RSD)
Automatic exclusion of deviant value by setting SD and %RSD |
| Baseline correction |
Automatic correction of the baseline drift through offset processing in peak height/peak area mode |
| Signal processing interval settings |
Signal processing interval setting in peak height/peak area mode |
| Sensitivity correction |
Automatic calibration curve correction with sensitivity monitoring |
| Tabular data processing |
Calculation of final concentration from sample volume, lab bench dilutions, autosampler dilutions, and weight factors |
| Multi-tasking |
Software such as word processors can be used while the measurements are running |
| Parameters recall |
Template function available |
| Procedure/ results display |
MRT worksheet(MRT : Measured Result Table) |
| Result printout |
Summary report |
| QA/QC-related |
It is possible to choose between aborting or continuing the measurement in the event of the reference values for the correlation coefficient, LCS, SPK, DUP, %RSD etc. being exceeded. |
| Electronic records |
| • |
Management using login IDs/ passwords |
| • |
Access restrictions according to user level |
| • |
Recording of the log |
| • |
Audit trail |
| • |
Electronic signatures |
|
| Power requirements |
AC100 ±10% (no sudden voltage fluctuations)
230VA, 50/60Hz |
| Dimensions & Weight |
Width 835 x Depth 545 x Height 460mm, 70kg(not including the flue, protruding parts) |
| Ambient temp. |
10°C to 35°C |
| Ambient humidity |
20% to 80%(however,below 70% at over 30°C) |
| Burner unit |
Type |
Air-cooled premix type |
| Burner head |
Made of titanium,10cm slot (the 5cm slot for high temperature flames is optional) |
| Nebulizer |
Pt-Ir capillary with PCTFE orifice and ceramic impact bead (Hydrofluoric acids can be used) |
| Chamber |
Polypropylene |
| Positioning control |
Automatic search for the optimal burner height, automatic adjustment of vertical position, manual adjustment of back and forth position |
| Gas control unit |
Flow rate control |
Fuel gas, support gas, automatic flow rate setting (0.1L/min step), automatic search for the optimal gas flow rate |
| Safety measures |
| • |
Safe ignition/ extinction sequence, |
| • |
Automatic extinction in a power failure |
| • |
Flashback prevention through gas pressure monitoring, |
| • |
Prevention of raw gas release when the flame dies out, |
| • |
Burner misuse prevention |
| • |
Support for safety inspections through software, system safety mechanisms self-diagnostics function |
|
| Graphite Furnace Atomizer GFA-EX7i |
Heating control system |
Drying: Current control system through automatic temperature calibration
Ashing, atomization; digital temperature control via optical sensor system |
| Heating temp.range |
Room temperature to 3000°C |
| Heating conditions setting |
| • |
Max. 20 stages, |
| • |
2 line switching of the inner gas type |
| • |
Setting of the high sensitivity mode, |
| • |
Enrichment inside furnace Max. 20 times |
| • |
Optimal temperature program search function |
|
| Safety measures |
| • |
Cooling water monitor, |
| • |
Gas pressure (2 line)monitor, |
| • |
Overcurrent protection, |
| • |
Furnace block cooling check |
|
| Cooling Water,Ar gas |
Cooling water: 0.6 to 1.5L/min (supply pressure 0.08 to 0.15MPa), water temp.10 to 30°C
Ar gas: Max. 3.5L/min, supply pressure 0.32 to 0.38MPa |
| Power requirements |
AC200V±5%, max. 6000VA, both 50/60Hz |
| Dimensions & weight of power supply unit |
255 W mm x 428 D mm x 460 H mm, 41kg |