Atomic Absorption Spectrophotometers(AAS)

AA-6300

Next Generation Double Beam
Atomic Absorption Spectrophotometer

AA-6300

The newly developed,high hroughput dynamic beam maneger system achieves measurements with the world's highest level of sensitivity (Direct determination - Flame:Pb, 0.1ppm; Furnace:Pb, 0.3ppb)

  • A next generation double beam system which drastically reduces the noise of a double beam optical system through the adoption of a high throughput dynamic beam manager system and a high speed operation digital filter.

The newly developed "WizAArd" (Windows-based 32bit)
software - a match for this electronic records era - is much more easier to understand and use, as well as being more secure

  • In order to strengthen the security of electronic records, the ability to manage using login ID names and passwords, management of usage authority restriction according to user level, electronic signatures, and audit trail functions have been equipped as standard.
  • QA/QC functionality as required by the U.S.
    Environmental Protection Agency (EPA) has been improved even further.
  • Hardware validation software is also standard equipment.

Switch smoothly between Flame and Furnace

  • All that is involved in switching from flame to furnace is to remove the burner head and put in place the furnace section. These steps require no tools, and can be done easily and speedily.
  • Furthermore, the single autosampler can be used for both flame and furnace, which negates the need to prepare the autosampler for each analysis method.

Adoption of Automatic Burner Head Height Control Mechanism

  • In addition to the automated control of the flame gas flow rate conditions, the burner head height can be automatically controlled as well. The optimal parameters for each element/ sample can be set automatically to enable continuous measurement.

Specifications

Photometric system Wavelength range 185 to 900nm
Mounting Aberration-corrected Czerny-Turner mounting, Number of grating grooves: 1800lines/mm, Forcal length: 298mm
Bandwidth 0.2 0.7 0.7 (Low) 2.0 (Low) nm (4 step automatic switching)
Detector Photomultiplier (short wavelengths), semiconductor (long wavelengths)
No. of lamps mounted 6 lamps, 2 user-selected lamps can be illuminated at the same time (1 for measurement,1warming up for next measurement)
Photometric method Flame: Optical double beam. Furnace: Electrical double beam
Background correction High-speed self-reversal method(BGC-SR), high-speed deuterium lamp method(BGC-D2)
Lamp mode EMISSION, NON-BGC, BGC-SR, BGC-D2,
Data processing Software requirments MS-WindowsTM2000
Parameter settings Wizard method
Measurement mode Flame suction method, flame microsampling method, furnace method
Concentration conversion mode Calibration curve method(can be selected from 1st, 2nd, and 3rd orders)
Standard addition method(1st order), simple standard addition method(1st order)
Repeat analysis Max. 20 times Display of mean values,deviation(SD), coefficient of variation(RSD)
Automatic exclusion of deviant value by setting SD and %RSD
Baseline correction Automatic correction of the baseline drift through offset processing in peak height/peak area mode
Signal processing interval settings Signal processing interval setting in peak height/peak area mode
Sensitivity correction Automatic calibration curve correction with sensitivity monitoring
Tabular data processing Calculation of final concentration from sample volume, lab bench dilutions, autosampler dilutions, and weight factors
Multi-tasking Software such as word processors can be used while the measurements are running
Parameters recall Template function available
Procedure/ results display MRT worksheet(MRT : Measured Result Table)
Result printout Summary report
QA/QC-related It is possible to choose between aborting or continuing the measurement in the event of the reference values for the correlation coefficient, LCS, SPK, DUP, %RSD etc. being exceeded.
Electronic records
Management using login IDs/ passwords
Access restrictions according to user level
Recording of the log
Audit trail
Electronic signatures
Power requirements AC100 ±10% (no sudden voltage fluctuations)
230VA, 50/60Hz
Dimensions & Weight Width 835 x Depth 545 x Height 460mm, 70kg(not including the flue, protruding parts)
Ambient temp. 10°C to 35°C
Ambient humidity 20% to 80%(however,below 70% at over 30°C)
Burner unit Type Air-cooled premix type
Burner head Made of titanium,10cm slot (the 5cm slot for high temperature flames is optional)
Nebulizer Pt-Ir capillary with PCTFE orifice and ceramic impact bead (Hydrofluoric acids can be used)
Chamber Polypropylene
Positioning control Automatic search for the optimal burner height, automatic adjustment of vertical position, manual adjustment of back and forth position
Gas control unit Flow rate control Fuel gas, support gas, automatic flow rate setting (0.1L/min step), automatic search for the optimal gas flow rate
Safety measures
Safe ignition/ extinction sequence,
Automatic extinction in a power failure
Flashback prevention through gas pressure monitoring,
Prevention of raw gas release when the flame dies out,
Burner misuse prevention
Support for safety inspections through software, system safety mechanisms self-diagnostics function
Graphite Furnace Atomizer GFA-EX7i Heating control system Drying: Current control system through automatic temperature calibration
Ashing, atomization; digital temperature control via optical sensor system
Heating temp.range Room temperature to 3000°C
Heating conditions setting
Max. 20 stages,
2 line switching of the inner gas type
Setting of the high sensitivity mode,
Enrichment inside furnace Max. 20 times
Optimal temperature program search function
Safety measures
Cooling water monitor,
Gas pressure (2 line)monitor,
Overcurrent protection,
Furnace block cooling check
Cooling Water,Ar gas Cooling water: 0.6 to 1.5L/min (supply pressure 0.08 to 0.15MPa), water temp.10 to 30°C
Ar gas: Max. 3.5L/min, supply pressure 0.32 to 0.38MPa
Power requirements AC200V±5%, max. 6000VA, both 50/60Hz
Dimensions & weight of power supply unit 255 W mm x 428 D mm x 460 H mm, 41kg

  • Auto sampler ASC-6100F
  • Furnace kit ASK-6100
    ASK-6300

Specifications common to flame and furnace

Function
Zero point detection
Auto rinse
Auto diagnosis
Random access
Max. no. of samples/ reagents
For reagents - 8 positions
60 samples (both can be accessed randomly)
Sample Container
16mL sample vial (test tube for flame suction), or
2mL (sample cup for flame micro-sampling/furnace)
40mL sample vial (glass container for flame suction), or
20mL (sumple cup for flame micro-sampling/furnace)
Nozzle rinse
Solvent discharge method (during flame micro-sampling/ furnace measurement)
Solvent aspiration method (during flame measurement)
Rinse water bottle 2L
Power requirements AC100V, 100VA, both 50/60Hz
ASC-6100F : 330 W x 285 H x 280 D mm, 7.5kg
If the ASC/ASK is combined: 360 W x 360 H x 280 D mm, 15kg

Furnace/ (flame micro-sampling) specifications

Sampling function
Automatic dilution
Automatic reagent addition
Syringe 250 µL
Injection volume 2 to 100 µL
Repeatablity 1%R.S.D. (20 µL)
Carryover
Rinse port Below 0.00001
Mixing port Below 0.00001
Mixing port cleaning
Solvent discharge method
Solvent discharge and rinse with next sample
Mixing function
Uses the mixing port
Max. mixable volume 600 µL
No. of reagents added Max. 4 solutions The injection order of the sample and reagent can be set(when not mixing)
The number of solutions that can be mixed are:
Calibration curve method :Max. 5 solutions
Standard addition method :Max. 6 solutions
Automatic dilution reanalysis With the calibration curve created by the measurement results for an unknown sample.
Diluted by the factor that is automatically calculated so that it falls within the calibration curve range.
To be set to 10 times in all cases when the factor cannot be calculated.

Note

  • The PC, display, and printer are not included in the standard configuration. Make arrangements to obtain these separately.
  • The ASC-6100F, ASK-6100, ASK-6300, GFA-EX7i are special accessories.
  • The separate micro-sampling kit is required for flame micro-sampling measurement.