The ICP Emission Spectrometer having the following features is applied for a wide field of elemental analysis.
In the meantime, with the expansion of the application field, however, the following problems have emerged.
For the settlement of the above problems, advanced functions as mentioned below have been embodied in the compact body of the ICPS-7000.
The ICPS-7000, which embodies a large capability in a compact body, characteristic of space saving and energy saving, is an instrument commensurate with the 21st century.
| Optical system | Czerny-Turner mounting | |
|---|---|---|
| Focal length | 750mm | |
| Number of grooves of the grating | 2,160 grooves/mm | |
| Wavelength range | 190 - 380 nm (2nd spectra) 380-850 nm (1st spectra) |
|
| Reciprocal linear dispersion | 0.22 nm/mm (2nd spectra) 0.42 nm/mm (1st spectra) | |
| Torch section | Observation direction selectable(axial direction, radial direction) Cyclone chamber: Made of Pyrex glass Plasma torch : Made of quartz, mini torch Nebulizer : Coaxial type, made of Pyrex glass |
|
|---|---|---|
| Gas flow rate | Setting range Coolant gas 20 L/min max. Auxiliary gas 2 L/min max. Carrier gas 2 L/min max. Purge gas 3 L/min max. |
|
| Oscillator | Crystal oscillator | |
|---|---|---|
| Frequency | 40.68 MHz | |
| Circuit | Power MOS FET | |
| OS | Windows® compatible | |
|---|---|---|
| Measuring wavelength to be set | Qualitative analysis 1 : 72 elements x 1 wavelength Quantitative analysis 2 : 216 wavelengths max. for 72 elements Quantitative analysis : 72 wavelengths max. for 72 elements |
|
| Database | Analytical wavelength data : 72 elements x 16 wavelengths max. Registration to data permitted. |
|
| Analytical card | 100 cards. | |
| Quantitative analysis | Standard samples for calibration curve creation : 16 samples max. / card Drift correction Internal standard correction (sequential correction of internal standard) Background correction Blank signal elimination |
|
*Windows® is a trade mark of Microsoft Corp., U.S.A.