Thermal Analysis Instruments
Differential Thermal Analyzer
High Temperature (ambient to 1500 C) heat-flux Differential Thermal Analyzer
This DTA uses a dumbbell-shaped detector. A temperature controller, gas flow adjuster, and transmission interface have all been incorporated into a slim, 173 cm-wide body. A high-temperature DSC function is also included.
- Temperature range: Room temperature to 1500°C
- Measurement range: ±0.2 to ±1000 µV (from ±0.2 mW)
- Heating speed: 0 to +50°C/min
High Temperature Heat Flux DTA, Provides Quantitative Calorimetry Measurements
Quick Response and High Sensitivity
Accurate Temperature Control
High Temperature DSC Type Performance
Rapid Atmosphere Purge
Specifications
|
Temperature range
|
Ambient to 1500°C
|
Temperature program format
|
99 steps maximum
|
|
(accurate type use up to 1100°C)
|
Temperature program file
|
Up to 100 files
|
|
Measuring range (DTA/DSC)
|
0.2 to 1000 V/0.2mW
|
Sample
|
Solid or liquid
|
|
Signal output
|
Analog and digital
|
Dimensions
|
W173x D540xH400mm
|
|
Programmable heating rate
|
0.1 to 50.0°C/min, and 0.1 to 50.0°C /hour
|
Power supply
|
AC100V, 120V, 220V,
|
|
Hold time
|
0 to 999min, 0 to 999 hour
|
240V, 1.2kVA, 50/60Hz
|
|
Atmosphere control
|
Built-in gas flowmeter (250ml/min maximum)
|
|
|