Product

Laboratory Instruments

Total Organic Carbon Analyzer TOC-V Series

Total Organic Carbon Analyzer TOC-V Series

  Total Organic
Carbon Analyzer
PC-controlled
Total Organic
Carbon Analyzer
Basic model Total Organic Carbon Analyzer Wet
oxidation
Total
Organic
Carbon Analyzer
PC-controlled Total Organic Carbon Analyzer
high-sensitivity model standard model high-sensitivity model standard model
Model TOC-VCSH TOC-VCSN TOC-VCPH TOC-VCPN TOC-
VE
TOC-
VWS
TOC-
VWP
Measurement method 680 degC combustion catalytic oxidation/NDIR method wet oxidation/NDIR
Operation method standalone PC-controlled standalone standalone PC-controlled
Measured items TC,IC,TOC,NPOC
(optional POC,TN)
TC,IC,
TOC,
NPOC (optional TN)
TC,IC,TOC,NPOC
Applicable samples aqueous sample (optional solid/gas samples) aqueous sample aqueous sample
Measurement range (mg/L) TC:
0 to 25000
IC:
0 to 30000
TC:
0 to 25000
IC:
0 to 3000
TC:
0 to 25000
IC:
0 to 30000
TC:
0 to 25000
IC:
0 to 3000
TC:
0 to 20000
IC:
0 to 20000
TC:
0 to 3000
IC:
0 to 2500
Detection
limit
4mg/L 50mg/L 4mg/L 50mg/L ---- 0.5mg/L
Measurement accuracy (reprodu
cibility)
CV 1.5% max. CV2% max.
(CV3% max. at 8000mg/L or higher)
CV1.5% max.
(CV2% max.
at 1000mg/L
or higher)
Measuring
time
TC:
approx
.3mins.
TC:
approx
.3mins.
TC:
approx
.3mins.
TC:
approx
.3mins.
TC:approx
.3mins.
TC:approx.4mins.
IC:
approx
.3mins.
IC:
approx
.4mins.
IC:
approx
.3mins.
IC:
approx
.4mins.
IC:approx
.3mins.
IC:approx.4mins.
Sample injection automatic injection manual
injection
automatic injection
Sample injection volume 10 to 2000mL
variable
10 to 150
mL
variable
10 to 2000mL
variable
10 to 150
mL
variable
1 to 150mL
(requires
change of syringe)
350 to 20400
mL variable
IC pre-treatment Automatic internal acidification and sparging Sparge gas supply Automatic internal acidification
and sparging
Automatic dilution dilution factor 2 to 50 none dilution factor
2 to 50
Gas consumption approx. 1440 L/month approx. 2210 L/month approx. 1440 L/month approx. 2210 L/month approx. 2210 L/month approx. 3000L/month
(operating conditions: 8 hours/day x 5days/week)
Operating
keys
built-in use PC built-in built-in use PC
Display built-in LCD use PC built-in LCD built-in LCD use PC
Printer (CENTRONICS, ESC/P) PC printer Optional (CENT-
RONICS,
ESC/P)
PC printer
Ambient temperature range 5 to 35degC
Power supply AC100 ~ 127V ± 10%, MAX800VA
AC220 ~ 240V ± 10%, MAX1200VA
AC100
~
127V
± 10%,
MAX350VA
AC220
~
240V
± 10%,
MAX350VA
Dimensions approx. (W)440 x (D)560 x (H)460mm(excluding protrusions)
Weight approx. 40 kg approx
. 38 kg
approx. 40kg

Autosampler ASI-V

Vial types Select from three types: 24mL,40mL,125mL

Number of vials


Vial septum
24mL : 93
40mL : 68
125mL : 24

with dedicated septum
Sample sparging OK (if the external sparging kit is used)
Dimensions approx. (W)370 x (D)540 x (H)490mm (excluding protrusions)
Weight approx. 14 kg

TN (Total Nitrogen) unit TNM-1

Measurement method chemiluminescence
Measured items TN (Total Nitrogen)
Measurement range 0 to 4000mg/L
Measurement accuracy (reproducibility) CV3% max.
Measuring time approx. 4 mins.
Dimensions approx. (W)160 x (D)235 x (H)400mm (excluding protrusions)
Weight approx. 8 kg

Solid sample combustion unit SSM-5000A

Method TC:Catalytically aided combution oxidation at 900治C
IC:Pre-acidification, oven temperature:900ޡC
Measuring Range* TC:0.1mg to 30mg carbon
(1-20mg carbon in high sensitivity measurement)
IC:0.1mg to 20mg carbon
Sample Amount* 1gram-aqueous content < 0.5g
Analysis Time* 5 to 6 minutes at a gas rate of 500mL/min.
Carrier Gas 99.9% O2 at a 500mL/min.
Power Requirements 100-127V or 220-240V as ordered.
700VA. 50/60Hz.
Weight Approx. 30kg
*Will vary with sample type and measurement condition.