TOC-V series Options

Other Options

Gas Sample Injection Kit

Incorporating this kit with TOC-VC** allows manual sample injection from a micro liter syringe. Gas or aqueous samples can be injected. The TC, carbon dioxide, and TOC of a gas sample can be measured. (TOC is determined as (TC - CO2). CO is included in TOC.)

Small Vial Rack Set for ASI-V

This set consists of a vial rack and 9mL vials, and enables measurements of small sample volume. Applicable models are: TOC-VCSH/CPH/CSN/CPN.

  • ASI-V Magnetic Stirrer or External Sparge Kit cannot be used when this set is used.

Nitrogen Carrier Gas Kit

Using this kit, which consists of an oxygen-permeating tube and a CO2 absorber, nitrogen gas can be used as a carrier gas instead of high-purity air. This kit is available for TOC-VCSH/CPH/CSN/CPN, TOC-VE, and ON-LINE TOCVCSH. Purity of nitrogen gas should be 99.999% or higher. When the kit is used, the measurement range of both TOC and IC is limited to 0 to 100mg/L.

  • This kit cannot be combined with the TNM-1
  • This kit cannot be used with the Carrier Gas Purification Kit.

Carrier Gas Purification Kit

Compressed air supplied as instrument air or compressor air, for example, can be used as the carrier gas after elimination of the carbon-containing impurities, such as carbon dioxide and hydrocarbons. This kit is suitable for samples with TOC of 0.5 mg/L (0.5 ppm), or higher.

  • This kit is available for TOC-VCSN/CPN.
  • Compressed air supply pressure: 300 to 700 kPa
  • Air consumption: approx. 300 mL/min
  • Dust, mist, drain liquid to be removed in advance.