Powders & Particle Size Analyzers
Single Nano Particle Size Analyzer
Specifications
| IG-1000 |
| Measurement Principle |
Induced grating (IG) method |
| Measurement Range |
0.5 to 200 nm |
| Measurement Time |
30 sec (from the start of measurement to the display of results) |
| Sample Liquid Volume |
250 to 300 µL |
| Measurement Unit |
Light Source |
Semiconductor laser (wavelength: 785 nm; output: 3 mW) |
| Light-receiving Unit |
Photodiode |
| Cell |
Batch cell (material: Pyrex glass*1) |
| Output Terminal |
Serial output (connector type: D-Sub, 25 pin, female) |
| Temperature Setting Range |
7°C to 40°C (cell holder temperature setting range*2) |
| Operating Environment |
Temperature: 15°C to 35°C |
| Humidity: 20% to 80% (without condensation) |
Humidity: 20% to 80% (without condensation) |
| Power Supply |
AC 115 or 230 V ±10% |
| Size and Weight |
600 (W) x 400 (D) x 200 (H) mm, approx. 15 kg |
*1: Do not use the kind of solvents that may damage Pyrex glass.
* : Measurement is possible as long as the conductivity of the sample liquid does not exceed 400 μS/cm (microsiemens per centimeter). (For example, saline and seawater cannot be used for measurement unless they are greatly diluted.)
Parts Required for Installation
| Micropipette |
"Required to inject samples into the cell. Provide a micropipette that can measure out 200 µL with an
accuracy of 5%."
|