EDX-7000/8000/8100

Energy Dispersive X-ray Fluorescence Spectrometer

Measurement principle X-ray fluorescence spectrometry
Measurement method Energy dispersion
Target samples Solids, liquids, powders
Measuring range 11Na to 92U (EDX-7000) 6C to 92U (EDX-8000/8100)
Sample size W 300 x D 275 x approx.H 100 mm (excluding radiuses)
Maximum sample mass 5kg (200g per sample when using turret, Gross mass 2.4kg)

 X-ray generator
X-ray tube Rh target
Voltage 4 kV to 50 kV
Current 1 μA to 1000 μA
Cooling method Air-cooled (with fan)
Irradiated area Automatic switching in four stages: 1, 3, 5, and 10 mm diameter
Automatic switching in four stages: 0.3, 1, 3, and 10 mm diameter*1
Primary filters Five types (six, including the open position), automatic replacement

 Detector
Type Silicon drift detector (SDD)
Liquid nitrogen Not required (electronic cooling)

 Sample chamber
Measurement atmosphere Air, vacuum*1, helium (He)*2
Sample replacement* 12-sample turret
Sample observations Semiconductor camera

 Data processor
Memory 2 GB min. (32-bit), 4 GB min. (64-bit)
HDD 250 GB min.
Optical drive Super multi drive
OS Windows 7 (32-bit/64-bit)*3

 Software
Qualitative analysis Measurement/analysis software
Quantitative analysis
Calibration curve method, correction for coexistent elements, FP method,
film FP method, background FP method
Matching software Intensity/content
Utilities Automatic calibration functions (energy calibration, FWHM calibration)
Instrument status monitoring function  
Analysis results tabulation
function
 

 Installation
Temperature
10 °C to 30 °C (temperature fluctuation rate 2 °C/hour max.,
temperature fluctuation range: 10 °C max.)
Relative humidity 40 % to 70 % (no condensation)
Power supply 100-240 V AC ±10 %, 2 A earthed socket
Dimensions W 460 x D 590 x H 360 mm
Weight Approx. 45 kg

*1 Option for EDX-7000/8000/8100
*2 Option for EDX-7000
*3 Microsoft Office is not included.

Installation Example

Vacuum measurement unit (optional) consists of a control switch box and rotary pump.

For Research Use Only. Not for use in diagnostic procedures.

This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.

Top of This Page