Cyanic Analysis System

Cyanic Analysis System

Analysis Examples

1. Standard Sample

1. Standard Sample

Chromatograms of 1 µg/L (1 ppb) standard cyan and cyanogen chloride samples are shown above. They show that adequate reproducibility can be obtained even at 1/10 the concentration of the criteria value of 0.01 mg/L. In addition, good linearity is obtained in the concentration range from 0.5 µg/L to 100 µg/L.

2. Tap Water Spiked with Potassium Cyanide

2. Tap Water Spiked with Potassium Cyanide

A chromatogram of water spiked with potassium cyanide (added concentration: 1 µg/L) is shown above. Since the added cyan ions react with the residual chlorine in the tap water, most is detected as cyanogen chloride.

Application Data L301A Analysis of Cyanide Ion and Cyanogen Chloride by Post-Column Ion Chromatography