This UHPLC-like model is the new standard for the Shimadzu LC series. It is compatible with a wide range of analysis conditions from conventional HPLC analysis to ultra-high separation analysis using core-shell columns with a particle size of 3 μm and below.

 Nexera XR High Pressure Gradient 
  Model Name Features
System controller SCL-40 Color screen with touch panel and system controller with reservoir tray
Number of connectable units : 8 (Using option: 12)
Connectable unit : Solvent delivery unit: max. 4, Autosampler: 1, Column oven: max. 4, Detector: max 2, etc.
Solvent Delivery Pump LC-40B XR
Pumping method : Parallel-type double plunger (approx. 10 μL/1 stroke)
Allowable maximum pressure : 70 MPa
Flow rate settings range : 0.0001 – 3.0000 mL/min (1.0 – 70 MPa)
3.0001 – 5.0000 mL/min (1.0 – 44 MPa) 5.0001 – 10.0000 mL/min
(1.0 – 22 MPa)
Flow rate accuracy :  ≦± 1% or ± 2 μL/min, whichever greater
(under specified conditions)
Gradient mode : High-pressure gradient (2 solvents (LC-40B XR standard) or 3 solvents) Quaternary low-pressure gradient (Only available for LC-40D XR)
Gradient Mixer
Degassing Unit DGU-403 Number of degassed solvents : 3
Degassed flow line capacity : 400 μL/1 line
Operating temperature range : 4 to 35°C
Autosampler SIL-40C XR
Injection method : Total-volume Injection (standard), loop injection (optional)
Allowable maximum pressure : 80 MPa
Injection volume : 0.1 to 50 μL, 0.1 to 2000 μL (optional)
Injection volume accuracy :  ≦± 1% (5 μL injection, n = 20)
Injection cycle time :  ≦6.7 seconds (under specified conditions)
Column Oven CTO-40S Temperature control type : Forced air circulation
Cooling Method : Electronic cooling
Temperature control range : Room temperature − 10°C to 85°C
Temperature accuracy : ± 0.8°C
Temperature precision : ± 0.1°C
UV-VIS Detector SPD-M40 Light source :Deuterium (D2) lamp, Tungsten lamp
Number of diode elements : 1024
Wavelength range : 190 to 800 nm
Wavelength accuracy :  ≦± 1 nm
Wavelength reproducibility :  ≦± 0.1 nm
Mobile phase monitor
Software Labsolutions LC/GC Multi LC-PDA


 
 