LCMS-8060
Liquid Chromatograph Mass Spectrometer
Bringing together ultra fast scanning technologies with high sensitivity pushes the limits of what we can see. The LCMS-8060 is built for high data quality and drives confidence and consistency 24/7.
High-Sensitivity Quantitation of Verapamil in Plasma
The triple quadrupole LCMS-8060 has an advanced ion guide design to increase ion production and detection and makes a meaningful impact on high sensitivity detection.

Actual Conc. (ng/mL) |
Calculated Conc. (ng/mL) |
Area RSD (%, n=3) |
Accuracy (%) |
0.0001 | 0.000100 | 2.97 | 100.3 |
0.0005 | 0.000508 | 6.15 | 101.7 |
0.001 | 0.000942 | 5.32 | 94.3 |
0.005 | 0.00491 | 3.54 | 98.9 |
0.01 | 0.00949 | 2.94 | 95.0 |
0.05 | 0.0511 | 2.14 | 102.4 |
0.1 | 0.0996 | 1.18 | 99.8 |
0.5 | 0.522 | 0.63 | 104.5 |
1 | 1.01 | 0.26 | 100.8 |
5 | 5.28 | 0.43 | 105.6 |
10 | 10.0 | 0.60 | 100.0 |
50 | 48.8 | 0.33 | 97.8 |
- Verapamil spiked into a plasma crash sample can be quantified at 100 ag on column with a precision of 5.08% for multiple injections (n=10).
- The LCMS-8060 also delivers a linear dynamic range from 0.1 pg/mL to 50 ng/mL.
High sensitivity quantitation of steroid hormones