UV accessory selection guide for liquids

Liquid Samples

Sample Analytical form and analytical condition Accessory
Transparent Sample Sample volume: 2.5 mL min. Standard Sample Compartment + 10 mm Cell
Micro-volume measurement 1 mL min. Semi-Micro Cell + Micro Cell Holder with Mask
500 µL min. Micro Cell + Micro Cell Holder with Mask
50 µL min. Super-micro Cell + Super-micro Cell Holder
For automatically measuring samples in multiple cells MMC-1600 8/16 Series Micro Multi-Cell Holders and Cells
Samples with high absorbance, but that are difficult to dilute (short optical path measurement) Short-Path Cell (1, 2, 5 mm) + Spacer for Short-Path Cell
Samples with low absorbance, but that are difficult to concentrate (long optical path measurement) Long-Path Cell (20, 30, 50, 100 mm) + Long-Path Rectangular Cell Holder
For automatically measuring samples in multiple cells Normal measurement Multi-Cell Sample Compartment (sample volume: 2.5 mL min.)
Small sample volumes (50 µL min.) MMC-1600 8/16 Series Micro Multi-Cell Holders and Cells
Requires temperature control CPS-100 Six-Cell Thermoelectrically Temperature-Controlled Cell Positioner
(sample volume: 2.5 mL min.)
For temperaturecontrolled measurements
(constanttemperature measurement)
Temperature-controlled with water circulation Constant-Temperature Cell Holder + NTT-2200P Constant-Temperature Water Circulator
Thermoelectrically temperature controlled Normal measurement TCC-100 Thermoelectrically Temperature-Controlled Cell Holder
Multi Samples CPS-100 Six-Cell Thermoelectrically Temperature-Controlled Cell Positioner
Tm Analysis / Temperature control S-1700 Thermoelectric Single Cell Holder
Automatically supplies sample to flow cells
(automatic analysis)
Requires temperature control (constant-temperature water circulation) 160C Sipper Unit + NTT-2200P
Temperature control not necessary 160L/160T/160U Sipper Unit (Select type based on liquid volume.)
Requires accurate control of aspiration volume. Requires temperature control
(constant-temperature water circulation)
Syringe Sipper CN + NTT-2200P (Select flow cell based on liquid volume.)
Temperature control not necessary Syringe Sipper N (Select flow cell based on liquid volume)
For automating measurement of multiple samples Sipper Unit or Syringe Sipper + ASC-5 Auto Sample Changer
Suspension Sample Absorption measurement of suspension samples Wavelength range: 240 nm min. Integrating Sphere Attachment (ISR-2600, ISR-2600Plus, ISR-603)
For measuring UV region above 190 nm SolidSpec-3700DUV
Turbidity measurement Light transmitted light turbidity measurement
(commonly used measurement method)
10/50 mm Cell + Long-Path Rectangular Cell Holder
(Optical path length of cell depends on test method.)
Integrating sphere turbidity measurement Integrating Sphere Attachment (ISR-2600, ISR-2600Plus, ISR-603)

For Research Use Only. Not for use in diagnostic procedures.

This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.

Top of This Page