Name of Element As Cd Hg Pb
PDE (μg/day) 15 5 15 10
Maximum Permitted Concentration
*1
(μg/g)
1.5 0.5 1.5 1
Maximum Permitted Concentration in
Measurement Solution
*2
(μg/g)
3 1 3 2
Detection Limit
*2
(μg/L) 0.009 0.003 0.003 0.001
*1: When the maximum intake per day is 10 g
*2: When a 0.2 g sample is pretreated and diluted to a volume of 100 mL
Name of Element
Sample name
As Cd Hg Pb
① 0.21 0.029 N.D. 0.035
② 0.08 0.018 N.D. 0.009
③ 0.11 0.008 N.D. 0.037
④ 0.29 0.071 N.D. 0.095
Recovery rate (%) 97 98 92 101
N.D.: Not detected
Spike recovery rate (%) = { (Analysis value for the spike recovery test sample − Analysis value) / Spike concentration} × 100
Application
News
No.
© Shimadzu Corporation, 2016
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
Company names, product/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation or its
affiliates, whether or not they are used with trademark symbol “TM” or “®”. Third-party trademarks and trade names may be used in this
publication to refer to either the entities or their products/services. Shimadzu disclaims any proprietary interest in trademarks and trade names
other than its own.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
to change without notice.
www.shimadzu.com/an/
J123
First Edition: Aug. 2016
p Analysis
The calibration curve method was used for
quantitative analysis of the elements arsenic,
cadmium, mercury, and lead.
To verify the analysis results, a spike recovery test
sample was created by adding a standard solution
of measurement elements after the sample
decomposition. Quantitative analysis was performed
in the same way using this sample.
p Analytical Results
Table 3 shows the permitted daily exposure (PDE) levels,
maximum permitted concentrations, and detection
limits. The analysis results and spike recovery rates are
shown in Table 4. These show favorable recovery rates
and confirm that the results obtained were appropriate.
The sensitivity was also evidently sufficient, even at the
maximum intake of 10 g per day.
p Conclusion
Using the ICPMS-2030, it is possible to perform an
analysis for controlling the heavy metals arsenic,
cadmium, mercury, and lead in supplements.
[Reference]
USP 2232 Elemental Contaminants in Dietary Supplements
Table 3 PDE Levels, Maximum Permitted Concentrations, and Detection Limits for Elemental Impurities Specified in USP 2232
Table 4 Analytical Results for Supplements (concentration in solid sample, μg/g)