Application
News
No. A596
First Edition: Jul. 2019
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
Shimadzu disclaims any proprietary interest in trademarks and trade names used in this publication other than its own.
See http://www.shimadzu.com/about/trademarks/index.html for details.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
to change without notice.
© Shimadzu Corporation, 2018
www.shimadzu.com/an/
First, the spectral evaluation item needs to be set. This function allows
setting of multiple items that can be evaluated at the same time.
Fig. 4 shows the detailed settings window of the spectral evaluation
function and Fig. 5 shows an enlarged view. The red film was regarded as
the acceptable product and evaluation was performed using "Cutoff -
Over" in the spectral evaluation function to differentiate the red film from
the yellow film. When "Cutoff - Over" is selected, judgment is made based
on the wavelength at the point when the measured transmittance has
exceeded the threshold in the set wavelength range. In this example, we
set the criterion of a pass judgment to the condition that the wavelength
at which transmittance exceeds 50 % is 550 nm or higher.
First under [Kind], select [Cutoff] and then [Over] ① . Next, set
[Wavelength Range] ② under [Parameter] to the wavelength range for
detection and [Threshold] ③ to the threshold value. In this example, we
set 450 to 600 nm for the wavelength range and 50 for the threshold
(since the judgment criterion is 50 % transmittance). Also select either
option ④to set whether a judgment is to be made when the threshold
value is exceeded from the long-wavelength side or short-wavelength
side (in this case we select [From short end]). Next, select the [Perform
pass/fail judgment] check box ⑤. Clicking ⑥ displays the drop-down
list shown in Fig. 5, and here we select [Pass if the evaluation is equal to or
more than the threshold]. Lastly, set the wavelength that serves as the
judgment criterion to 550 nm using [Threshold 1] ⑦.
Fig. 4 Evaluation Detailed Settings Window
Fig. 5 Enlarged View of Evaluation Detailed Settings Window
① Evaluation Type ⑥ Evaluation Conditions
Click the [OK] button to set the spectral evaluation items shown in Fig. 6.
Fig. 6 List Showing Evaluation Item for Colored Film
Click the [Others] tab in Fig. 6 and select the [After measurement,
data is automatically added to the evaluation table] check box to
perform pass/fail judgment automatically after spectrum acquisition
(Fig. 7). Also, saving an evaluation template allows you to perform the
same test at any time simply by loading the template.
Fig. 7 Others Tab
Results
Fig. 8 shows the obtained transmittance spectrum of each sample
and the spectral evaluation results. (Fig. 9 shows an enlarged view of
the evaluation results.) "PASS" is indicated for the red film since the
evaluation value is 579 nm. However, the yellow film received a
"FAIL" and is colored red in the table due to having a cutoff
wavelength below 550 nm.
The column color in the evaluation result table changes when a result
deviates from the set evaluation criteria and this allows you to
determine at a glance whether a test has passed or failed. Moreover,
this table information can be copied into other applications using the
copy and paste function or the [Excel Export] function to facilitate
creation of various reports.
Fig. 8 Measurement Results and
Spectral Evaluation Results
Fig. 9 Enlarged View of Spectral Evaluation Results
Conclusion
This article introduced an example of automatically performing
transmittance measurement, analysis, and pass/fail judgment of
colored film using the spectral evaluation function of the
LabSolutions UV-Vis software. The use of this function enables
various quality inspections with great efficiency.
LabSolutions is a trademark of Shimadzu Corporation in Japan and/or
other countries.
Third-party trademarks and trade names may be used in this publication
to refer to either the entities or their products/services, whether or not
they are used with trademark symbol “TM” or “®”.