Application News No. M298 Gas Chromatography Mass Spectrometry Analysis of Mixed Polymer Sample as Microplastics using Pyrolysis GC/MS Fine plastic particles with sizes of several μm to about 5 mm are called microplastics (MPs), and their effect on pollution and ecosystems in marine environments is a serious concern. It has been pointed out that bioaccumulation may occur by a process of adsorption of harmful substances on MPs, and ingestion of those MPs by marine organisms in marine environments. Due to that, many surveys of the actual condition of MPs and evaluation of their toxicity are being conducted. As analytical instruments, the Fourier transform infrared spectrometer (FTIR) is used in qualitative analysis of comparatively large MPs, while the FTIR microscope employing an infrared microscope is mainly used for fine MPs which cannot be analyzed by the attenuated total reflectance (ATR) method of FTIR. The FTIR microscope enables highly sensitive analysis of fine MPs with a size of approximately 10 μm. However, in cases where it is difficult to distinguish the MPs containing multiple types of fine particles, it has been reported that the pyrolysis gas chromatography/mass spectrometry (Py-GC/MS) method is effective. Using Py-GC/MS method, qualitative analysis of individual polymers contained in mixed samples is possible by highly sensitive detection of the distinctive pyrolysis products of each polymer. This article reports the results of a qualitative analysis of a sample, prepared by mixing multiple polymers to simulate MPs, using Py-GC/MS method. Y. Kudo, Y. Ikezawa, Y. Aoyama Fig. 1 Appearance of Py-GC/MS System Sample and Analysis Conditions Fig. 1 shows the Py-GC/MS system (GCMS-QP2020 NX and EGA/PY-3030D pyrolyzer) used in the analysis. Approximately 0.05 mg each of fragments of four types of commercially-available standard polymer sample materials (polyethylene (PE), polypropylene (PP), polystyrene (PS), and polyvinyl chloride (PVC)) were introduced into a sample cup, and about 1 mg of wool was placed in the cup to prevent scattering. This sample was then set in the auto-shot sampler of the pyrolyzer. Table 1 shows the instruments and the analytical conditions. According to References (1) and (2), the contents of the polymers were judged by detecting the pyrolysis products shown in Table 2 (however, some of these compounds were selected independently). Table 2 Pyrolysis Products and Analytical Conditions Used in Detection of Polymers Polymer Pyrolysis product Retention time (min) SIM monitoring ion PE C20, alkane 20.937 99, 85 C20, Į-alkene 20.879 97, 83 C20, Į, Ȧ-alkene 20.817 95, 82 PP 2,4-dimethylhept-1-ene 5.231 126, 70 2,4,6,8-tetramethyl-1-undecene 12.908 111, 69 2,4,6,8-tetramethyl-1-undecene 13.027 111, 69 2,4,6,8-tetramethyl-1-undecene 13.145 111, 69 PS Styrene 6.17 104, 78 3-butene-1,3-diyldibenzene 18.136 208, 91 5-hexene-1,3,5-triyltribenzene 25.032 312, 207 PVC Benzene 2.498 78, 51 1-Chloroindan 8.874 116, 115 Dihydronaphthalene 10.835 130, 115 Azulene 11.145 128, 102 Table 1 Analytical Conditions Instruments Pyrolyzer : EGA/PY-3030D multi-shot pyrolyzer, AS-1020E auto-shot sampler (Frontier Laboratories Ltd.) GC-MS : GCMS-QP™2020 NX Column : UA-5 (MS/HT)-30M-0.25 (Length 30 m, I.D. 0.25 mm, df = 0.25 μm) (Frontier Laboratories Ltd.) Pyrolyzer conditions Pyrolyzer furnace temp. : 600 °C Interface temp. : 300 °C (manual) GC conditions MS conditions Vaporizing chamber temp. : 300 °C Interface temp. : 300 °C Column oven temp. : 40 °C (2 min) o 10 °C/min o 320 °C (16 min) Ion source temp. : 230 °C Carrier gas : Helium Ionization method : El Control mode : Constant linear velocity Measurement mode : Scan/SIM (m/z 29 - 700) Injection mode : Split (1 : 50) Scan event time : 0.3 s Linear velocity : 36.1 cm/min SIM event time : 0.15 s Purge flow rate : 3 mL/min
First Edition: Jun. 2020 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. Shimadzu disclaims any proprietary interest in trademarks and trade names used in this publication other than its own. See http://www.shimadzu.com/about/trademarks/index.html for details. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. © Shimadzu Corporation, 2020 www.shimadzu.com/an/ Application News No. M298 Simultaneous Analysis of Polymers in Mixed Polymer Sample Fig. 2 shows the total ion current chromatogram (TIC chromatogram) obtained by analyzing the mixed sample containing multiple polymers. Since the result of this analysis takes the form of a complex chromatogram in which the pyrolysis products of the respective polymers are intermixed, it was difficult to specify the individual polymers contained in the sample from this chromatogram alone. Fig. 3 shows the SIM chromatograms of the pyrolysis products of the respective polymers. By tracing the distinctive pyrolysis products of the respective polymers, the contents of the polymers could be specified accurately, even multiple polymers were intermixed in a sample. Fig. 2 TIC Chromatogram Obtained by Analysis of Mixed Polymer Sample by Py-GC/MS Method Fig. 3 SIM Chromatograms of Pyrolysis Products of Polymers in Mixed Polymer Sample Conclusion In this experiment, it was found that qualitative analysis of various polymers is possible, even the sample contains multiple polymers, by monitoring the pyrolysis products of the respective polymers using the Py-GC/MS method. Although the object of analysis in this experiment consisted of four substances (polymers), it is thought that this technique is also applicable to larger numbers of polymers. Thus, this technique is expected as a new analysis method of MPs. <References> (1) M. Fisher and B. M. Scholz-Böttcher, Environ. Sci. Technol., 51,5052-5060 (2017) (2) S. Tsuge, H. Ohtani, C. Watanabe: Pyrolysis-GC/MS Data Book of Synthetic Polymers -Pyrograms, Thermorgams and MS of Pyrolyzers-,1st Edition, Elsevier, 420 (2011) GCMS-QP is a trademark of Shimadzu Corporation in Japan and/or other countries. Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “£”. 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0 45.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 (x1,000,000) 1:TIC (1.00) min C20, alkane C20, α-alkene Pyrolysis products of PE Pyrolysis products of PP Pyrolysis products of PS Pyrolysis products of PVC C20, α,ω-alkene 20.50 20.75 21.00 21.25 0.5 1.0 1.5 2.0 2.5 3.0 3.5 (x10,000) 82 95 20.50 20.75 21.00 21.25 1.0 2.0 3.0 4.0 5.0 6.0 7.0 (x10,000) 83 97 20.50 20.75 21.00 21.25 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 (x10,000) 85 99 12.75 13.00 13.25 13.50 0.2 0.4 0.6 0.8 1.0 1.2 (x100,000) 111 69 12.50 12.75 13.00 13.25 0.2 0.4 0.6 0.8 1.0 1.2 (x100,000) 111 69 12.25 12.50 12.75 13.00 0.2 0.4 0.6 0.8 1.0 1.2 (x100,000) 111 69 4.75 5.00 5.25 5.50 0.0 1.0 2.0 3.0 4.0 5.0 6.0 (x100,000) 126 70 5.75 6.00 6.25 6.50 1.0 2.0 3.0 4.0 5.0 6.0 7.0 (x1,000,000) 78 104 17.75 18.00 18.25 18.50 0.25 0.50 0.75 1.00 1.25 (x1,000,000) 91 208 24.75 25.00 25.25 25.50 0.00 0.50 1.00 1.50 2.00 2.50 (x100,000) 207 312 2.00 2.25 2.50 2.75 0.50 1.00 1.50 2.00 2.50 (x1,000,000) 51 78 10.50 10.75 11.00 11.25 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 (x10,000) 115 130 2,4-dimethylhept-1-ene 2,4,6,8-tetramethyl-1-undecene 2,4,6,8-tetramethyl-1-undecene 2,4,6,8-tetramethyl-1-undecene Styrene 3-butene-1,3-diyldibenzene 5-hexene-1,3,5-triyltribenzene Benzene Dihydronaphthalene 10.75 11.00 11.25 11.50 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 (x100,000) 102 128 Azulene 8.50 8.75 9.00 9.25 0.25 0.50 0.75 1.00 1.25 1.50 1.75 (x100,000) 115 116 1-Chloroindan
Related Products Some products may be updated to newer models. GCMS-QP2020 NX Gas Chromatograph Mass Spectrometer Related Solutions Hydrocarbon Processing Industry (Petrochemical, Ch Microplastics Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)