First Edition: Apr. 2020
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
Shimadzu disclaims any proprietary interest in trademarks and trade names used in this publication other than its own.
See http://www.shimadzu.com/about/trademarks/index.html for details.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
to change without notice.
© Shimadzu Corporation, 2020
www.shimadzu.com/an/
Application
News
No. G319
Analysis of Standard Gas (Moisture in N
2
)
The calibration curve was prepared by the absolute calibration curve method using nitrogen standard gas containing
approximately 5 ppm, 25 ppm, or 100 ppm of moisture. Fig. 2 shows the chromatogram, Fig. 3 shows the calibration curve, and
Table 2 shows the peak area and repeatability (N = 5) at each moisture concentration level. Although trace moisture analysis by
GC had been considered difficult until now, a trace moisture analysis was carried out at the ppm level, and good separation,
linearity, and repeatability were obtained. The peak of water separated by the ionic liquid capillary column was sharper than those
obtained with conventional columns and could also be obtained without tailing. Under these analysis conditions, a shorter analysis
cycle time was achieved by setting a constant oven temperature (120 °C) without GC oven cooling time. (Although a sharper peak
shape is possible by programmed temperature GC, the analysis cycle time becomes long.)
Fig. 2 Chromatogram of Nitrogen Standard Gas Containing Trace Moisture
Fig. 3 Calibration Curve
Table 2 Peak Area and Repeatability at
Each Moisture Concentration
Moisture concentration (ppm)
5.994 24.44 109.3
Peak area 1 42431 85912 272693
Peak area 2 43249 85077 274031
Peak area 3 42832 83731 272912
Peak area 4 42395 82438 274313
Peak area 5 42373 82360 272680
Average peak area 42656 83904 273326
%RSD 0.894 1.881 0.287
Nexis is a trademark of Shimadzu Corporation in Japan and/or other countries.
Watercol is a trademark of Sigma-Aldrich Co LLC.
Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not
they are used with trademark symbol “TM” or “£”.
Water
N
2
5.994 ppm (S/N ratio : 851)
25.44 ppm (S/N ratio : 1969)
109.3 ppm (S/N ratio : 6106)
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 min
0.0 25.0 50.0 75.0 100.0 Concentration
0
25000
50000
75000
100000
125000
150000
175000
200000
225000
250000
275000
Peak area
1
2
3
R
2
= 0.9999