Application News No.J106A Inductively Coupled Plasma Atomic Emission Spectrometry Analysis of Heavy Metals in Toys and Accessories by ICPE-9800 Series LAAN-A-CP-E019A n Introduction n Samples n Sample Preparation n Instrument and Analytical Conditions n Analysis Table 1 Analytical Conditions In recent years, safety issues related to inexpensive metal toys and accessories have become a social problem. In particular, the presence of heavy metals such as lead (Pb) and cadmium (Cd) in toys poses a health risk to infants if such contaminated toys are licked, placed in the mouth, or even swallowed. Therefore, toys and accessories are now required to be evaluated for safety with respect to their content using an appropriate test method. These test methods are defined according to various international standards, including ISO 8124 1) (Safety of toys) and the European standard EN71-Part3 2) (Safety of toys – Migration of certain elements). Both are dissolution tests based on the assumption that such toys and accessories will be ingested by infants, resulting in the absorption of toxic elements into the body. Such an analysis of toys and accessories requires that, in addition to being able to measure harmful trace elements with high sensitivity, measurement can be conducted for multiple samples quickly and at low cost. Here, using the Shimadzu ICPE-9800 series multi-type ICP atomic emission spectrometer, we conducted measurement of commercially available toys and metallic accessories as samples. The ICPE-9800 series, with its mini-torch plasma and spectrometer capable of simultaneous analysis of all elements at all wavelengths, can conduct high-throughput, low-cost analysis with high sensitivity and high accuracy. Processing was conducted according to the method specified in EN71-Part3. (1) Paint coated on toy Very fine paint scrapings are removed from the surface of the toy, and transferred to vials. Then, 5 mL of 0.07 mol/L hydrochloric acid is added for each 0.1 g of the scraped fragments, and after shaking for 1 hour in the dark, the vials are left standing at 37 °C for 1 hour. The eluates are then filtered. (2) Accessory A weighed sample is placed in a beaker 40 mm in diameter, and after immersing the sample in 0.07 mol/L hydrochloric acid heated at 37 °C, it is left standing at 37 °C for 2 hours, shielded from the light. The eluate is then filtered. Then, 0.07 mol/L hydrochloric acid is added to the respective filtrates of (1) and (2) to adjust their volumes to 20 mL, and these are used as sample solutions. Quantitative analysis was conducted using the calibration curve method. Commercially available toys (building block, vehicle, etc.) Commercially available metallic accessories (section of key ring, etc.) (including second-hand items such as those found at a flea market) [References] 1) ISO 8124-1: 2009 (Safety of toys – Part 1) 2) BS EN71-3: 2013, Safety of toys – Migration of certain elements The Shimadzu ICPE-9800 series multi-type ICP atomic emission spectrometer was used to conduct the measurements. Table 1 shows the measurement conditions. The ICPE-9800 series, with its echelle spectrometer and CCD detector, can conduct simultaneous measurement of all elements at all wavelengths, permitting high-throughput measurement even when there are many target elements and samples. Also, the plasma gas flowrate-suppressing mini-torch, the Eco mode which limits gas usage and power consumption during wait time, and the use of a vacuum spectrometer, which does not require purge gas, all greatly reduce running costs compared to conventional ICP instruments. Instrument : ICPE-9800 series Radio Frequency Power : 1.2 kW Plasma Gas Flowrate : 10 L/min Auxiliary Gas Flowrate : 0.6 L/min Carrier Gas Flowrate : 0.7 L/min Sample Introduction : Nebulizer 10 Misting Chamber : Cyclone chamber Plasma Torch : Mini Torch Observation : Axial (AX) Measurement Time : 2.5 min/sample (Including rinse time)
Application News No. For Research Use Only. Not for use in diagnostic procedures. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. © Shimadzu Corporation, 2015 www.shimadzu.com/an/ J106A Second Edition: Apr. 2015 First Edition: Sep. 2014 Table 2 shows the analytical results. Fig. 1 shows the spectral profiles. The detection limit shows clearly the sufficient detection sensitivity to measure less than 1/10 reference value. Use of the ICPE-9800 series for analysis of toys and accessories clearly demonstrated that high-sensitivity measurement can be conducted quickly and at low cost. Table 2 Analytical Results for Toys and Metallic Accessories (Unit: mg/kg) Element Toy Standard Detection Limit Sample Name EN ISO Toy 1 Toy 2 Toy 3 Toy 4 Toy 5 Accessory 1 Accessory 2 Al 70000 0.2 493 209 1080 188 1860 41 14 Sb 560 60 1 < 2 < < < < < As 47 25 1 < < 4 < < < < Ba 56000 1000 0.003 404 353 12 11 21 0.7 0.1 B 15000 0.03 0.6 1.0 0.6 27.7 33.3 0.6 0.7 Cd 17 75 0.02 < 0.3 0.8 < 1.5 0.1 0.1 Cr 460 * 60 0.05 22 34 1.0 103 174 0.1 < Co 130 0.04 < < < < < 0.2 2.2 Cu 7700 0.1 3 9 4 72 35 93 377 Pb 160 90 0.5 161 290 1 477 24 < 289 Mn 15000 0.006 7.4 14.1 1.0 2.4 10.4 0.1 0.2 Hg 94 60 0.3 < < < < < < < Ni 930 0.06 1.8 1.4 2.7 20 9 179 361 Se 460 500 1 < < < < < < < Sr 56000 0.002 20 9 29 2.3 1.2 0.19 0.13 Sn 180000 0.2 < < < 13.8 7.4 < 17.9 Zn 46000 0.04 1140 576 16900 407 18600 343 649 Concentrations in table = Measurement solution concentration × Dilution factor (20 mL/sample weight (g)) EN: EN71-Part3 Category Ⅱ ISO: ISO8124-3: 2010 *: Cr 3+ (trivalent chromium) standard value <: Below the detection limit Detection limit: 3 times the concentration of the standard deviation obtained from 10 repeated measurements of the calibration curve blank × Dilution factor (200) n Analytical Results n Conclusion Fig. 1 Spectral Profiles of Toys 0 500 1000 1500 2000 2500 3000 3500 Intensity Pb 220.353 nm STD 2 mg/L 0 500 1000 1500 2000 2500 3000 3500 Intensity Cd 228.802 nm STD 0.2 mg/L Toy-5 Toy-5 Toy-2 Toy-4
Related Products Some products may be updated to newer models. ICPE-9800 Series Multitype ICP Emission Spectrometers Related Solutions Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)