Application News Pharmaceuticals / HS-20 GC-2030 Analysis of Residual Solvents in Pharmaceutical Products by Headspace-GC-FID with Nitrogen Carrier Gas Following USP<467> - Procedure B No. AD-0228 ❑ Introduction USP <467> monograph describes Procedure A and B for analysing residual solvents in pharmaceutical products by Headspace Gas Chromatography (HS- GC) [1]. Procedure A is the main method, while Procedure B is to be carried out as verification when certain criteria is not met in Procedure A result. In Application News AD-0209, analysis of residual solvents in pharmaceutical products was done following Procedure A which was performed using nitrogen carrier gas, an inexpensive and abundant alternative to helium [2]. Here, we describe HS-GC method using nitrogen as the carrier gas to analyse Class 1 and Class 2 solvents following USP<467> Procedure B criteria. Table 1: HS-GC analytical conditions for residual solvent analysis following USP <467> Class 1 Standard Class 1 Standard (five solvents) was analysed for 7 times to determine the peak area percentage relative standard deviation (%RSD) and signal to noise ratio (S/N). The S/N ratios were calculated using USP method (Table 2). The chromatogram of Class 1 Standard is displayed in Figure 1. Carbon tetrachloride was co-eluted with 1,1,1-Trichloroethane when Procedure B was used (Figure 1). The lowest S/N value for benzene (Peak 4) obtained was 93, which is greater than the requirement stated in USP<467> (i.e., S/N ratio is not less than 5). The S/N ratio for the rest of the Class 1 Standards were greater than 3 which conformed to the requirement stated in USP<467>. The repeatability of peak areas, %RSD (n=7), for the five solvents (2 solvents co-elute) obtained ranges ❑ Results and Discussion Instruments and Column information GC-FID Nexis GC-2030 Auto Injector HS-20 Column SH-Stabilwax TM 30 m x 0.32 mm ID x 0.25 μm df HS parameter Oven Temperature 80 o C Sample Line Temperature 110 o C Transfer Line Temperature 120 o C Injection Time 1 min Pressurizing Gas Pressure 75 kPa Equilibrating Time 60 min Shaking Level 2 GC-FID parameter Injection Mode Split mode Split ratio 10 Carrier Gas Nitrogen Gas Flow Condition Constant linear velocity mode Linear velocity 35 cm/s Oven Temperature Programming 50 o C (20 min) →6 o C/min to 165 o C (20 min) Detector Temperature 250 o C Hydrogen Flow 32 mL/min Synthetic Air Flow 200 mL/min Make-up Gas Flow 24 mL/min ❑ Experimental Analytical conditions and sample preparation HS-20 headspace autosampler and Nexis TM GC-2030 (Shimadzu Corporation, Japan) were used in this work. The analytical conditions following Procedure B under water-soluble article section in USP<467> are shown in Table 1. Certified USP<467> Class 1 and 2 Standard solutions were purchased from Restek. The standards were prepared according to USP<467> Procedure B before analysis. Figure 1: HS-GC-FID chromatogram of Class 1 Standard following Procedure B in USP<467>. Peak labelling refers to Table 2. 2.0 2.5 3.0 3.5 4.0 min uV 1 2 & 3 4 5 from 1.8% to 5.1%. These results indicate that when using nitrogen carrier gas as a substitute of helium carrier gas, Procedure B criteria can still be fulfilled for Class 1 Standard.
Application News No. AD-0228 Peak No. Solvent %RSD (n=7) of peak area S/N ratio data 1 S/N ratio data 2 S/N ratio data 3 S/N ratio data 4 S/N ratio data 5 S/N ratio data 6 S/N ratio data 7 1 1,1-dichloroethene 5.1 23 54 36 37 32 32 79 2 & 3 1,1,1-trichloroethane and Carbon Tetrachloride 2.2 38 86 70 84 56 65 86 4 Benzene 1.8 119 127 108 138 93 107 104 5 1,2-dichloroethane 2.1 40 48 46 60 50 58 40 Table 2: Peak area repeatability (n=7) and signal to noise ratio (S/N) for Class 1 Standard Class 2 Standard Class 2 Standard consists of two groups, Class 2A and Class 2B. The chromatograms are shown in Figures 2 and 3, respectively. The repeatability results of Class 2A and Class 2B are compiled into Table 3. The peak area %RSD (n=7) values obtained for all the solvents ranges from 0.7% to 9.1%. The average (n=7) specific resolution between cis-1,2- dichloroethene and acetonitrile obtained with nitrogen carrier gas was 2.9. This meets the requirement of USP<467> that R s of these 2 compounds must not be less than 1. In summary, the results for Class 1 Standard and Class 2 Standard indicate that the HS-GC analysis method with nitrogen as carrier gas can achieve required sensitivity (S/N) and peak resolution as stated in the criteria of USP<467> Procedure B. Class 2A standard No. Compound %RSD (n=7) of peak area A1 Cyclohexane 5.7 A2 Methylcyclohexane 5.7 A3 & A4 trans-1,2-Dichloroethene and THF 4.7 A5 Methanol 1.3 A6 DCM 5.7 A7 Cis-1,2-Dichloroethene 5.8 A8 Acetonitrile 2.2 A9 Toluene 5.6 A10 1,4-Dioxane 1.6 A11 EthylBenzene 5.5 A12 p-xylene 5.7 A13 m-xylene 5.6 A14 Cumene 5.5 A15 o-xylene 5.4 A16 Chlorobenzene 5.6 Class 2B standard No. Compound %RSD (n=7) of peak area B1 Hexane 9.1 B2 1,2-dimethoxyethane 1.1 B3 Trichloroethene 7.1 B4 chloroform 5.1 B5 Methylbutyylketone 0.7 B6 Nitromethane 4.6 B7 Pyridine 1.1 B8 Tetralin 4.2 Table 3: Peak area repeatability (n=7) for Class 2A and 2B Standards 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 22.5 25.0 27.5 min 0 1000 2000 3000 uV Figure 3: Chromatogram of Class 2B Standard analysed using Procedure B B1 B2 B3 B4 B5 B6 B7 B8 Datafile Name:20190916 batch split 10_9172019_Class 2A std_006.gcd Sample Name:Class 2A std 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 min 0 25000 50000 75000 uV A1 A2 A3 & A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 Figure 2: Chromatogram of Class 2A Standard analysed using Procedure B A15 3.25 min 0 2500 5000 uV A7 A8 R s = 2.9
Application News No. AD-0228 ❑ Conclusions This study demonstrates the feasibility of using nitrogen as carrier gas in HS-GC-FID analysis of residual solvents in pharmaceuticals. The results of Class 1 and Class 2 standards fulfilled the criteria stated in USP<467> Procedure B from water-soluble article section. ❑ References 1. The United States Pharmacopeia, USP <467> RESIDUAL SOLVENTS. 2. Shimadzu Application News AD-0209, Analysis of Residual Solvents in Pharmaceutical Products by Headspace-GC-FID with Nitrogen Carrier Gas Following USP<467> -Procedure A www.shimadzu.com/an/ Shimadzu Corporation For Research Use Only. Not for use in diagnostic procedures. Contents and/or instrumentations in this application may not available in some countries under each regulation. Please contact the local representatives in details. SHIMADZU (Asia Pacific) Pte. Ltd 79 Science Park Drive, #02-01/08 Cintech IV, Singapore 118264, www.shimadzu.com.sg ; Tel: +65-6778 6280 Fax: +65-6778 2050 Copyright © 2020 SHIMADZU (Asia Pacific) Pte. Ltd. All rights reserved. No part of this document may be reproduced in any form or by any means without permission in writing from SHIMADZU (Asia Pacific) Pte. Ltd. Nexis is a trademark of Shimadzu Corporation in Japan and/or other countries. Stabilwax is either a trademark or a registered trademark of Restek Corporation in the United States and/or other countries.
Related Products Some products may be updated to newer models. Nexis™ GC-2030 Gas Chromatograph Related Solutions Small Molecule Pharmaceutical Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)