Application
News
No. L572
High Performance Liquid Chromatography
High Speed Analysis of Monosaccharides and
Disaccharides in Soft Drinks by ELSD-LT III
Saccharides show very narrow UV absorption
wavelength range, from 190 nm to 195 nm because
hydroxyl group is the only major functional group that
show UV absorption in the structure. Therefore, a
refractive detector (RID) is commonly used for this
analysis. However, gradient elution cannot be used
with RID because the baseline drifting derived from the
change of mobile phase composition during gradient
elution is practically unacceptable. So, RID is not
suitable for a simultaneous separation of compounds
that show widely different retention behaviors due to
expected long analysis time without gradient elution.
Evaporative light scattering detector (ELSD) is one of
universal detector that detects the scattering light from
the target compounds after nebulizing and evaporating
the mobile phase. ELSD provides reduced analysis time
and simultaneous separation of compounds that show
widely different retention due to applicability to
gradient elution.
Monosaccharides and disaccharides are mainly analyzed
by anion-exchange chromatography or ligand
exchange chromatography. However it takes a long
time to analyze. So, in this article, monosaccharides and
disaccharides were separated in short time by
hydrophilic interaction chromatography (HILIC) and
were detected by ELSD .
“Wide function”, a new feature of ELSD-LT III used in
this article, automatically optimizes a parameter that is
related to sensitivity and a single method file can be
used for data acquisition regardless of sample
concentration, from low to high.
Here, high speed analysis of monosaccharides and
disaccharide in soft drinks were carried out by ELSD-LT
III.
K. Matsuoka, N. Iwata
Analysis of 5 Standard Compounds
Table 1 shows the analytical conditions for 5 standard
compounds of different saccharides (Fructose, Glucose,
Sucrose, Maltose, Lactose). Fig. 1 shows the obtained
chromatogram (1000 mg/L each). It is recommended
to wash the column well before use. 5 saccharides
eluded within 5 minutes by gradient elution.
Table 1 Analytical Conditions
System : Nexera™ XR
Column : Shinwa Chemical Industries Ltd.
ULTRON AF-HILIC-CD (HT)
(100 mm × 3 mm I.D., 2 μm)
ULTRON AF-HILIC-CD (HT) Guard Cartridge
(5 mm × 2 mm I.D., 2.0 μm)
Mobile Phase : A) 10 mmol/L Ammonium Acetate
B) Acetonitrile
Time Program : B. Conc. 85% (0-0.5 min) ڀ82% (3-5 min)
ڀ85% (5.01-8 min)
Flow Rate : 0.8 mL/min
Column Temp. : 45 °C
Injection Vol. : 2 μL
Vial : LabTotal Vial for LC 1.5 mL, Glass
*1
Detection : ELSD-LT III
Gain : Wide
Filter : 4 sec
Drift Tube Temp. : 40 °C
Nebulizer Gas : N2
Gas Pressure : 350 kPa
*1: P/N: 227-34001-01
Fig. 1 Chromatogram of 5 Standard Sacchrides
(1000 mg/L each)
Repeatability
Table 2 shows the repeatability confirmed by repeated
analyses at 250 mg/L (n=6).
Table 2 Repeatability of 250 mg/L(n=6)
Compounds Retention Time (%RSD) Area (%RSD)
Fructose 0.05 2.33
Glucose 0.03 1.66
Sucrose 0.04 3.18
Maltose 0.03 2.28
Lactose 0.02 3.56
Linearity
Fig. 2 shows the calibration curves. Table 3 shows the
concentration ranges of the calibration curves. The
response of ELSD was plotted on double logarithmic
axes because the logarithm of ELSD response is in
proportion to the logarithm of concentration.
The calibration curves of Fructose and Glucose were
created using 6 different concentrations.
1.0 2.0 3.0 4.0 min
0
1000
2000
3000
4000
5000
6000
mV
1
2 3
5
4
ٹPeaks
1. Fructose
2. Glucose
3. Sucrose
4. Maltose
5. Lactose
0.0