Application News No. M294 Gas Chromatography Mass Spectrometry Analysis of Double Bond Position in Unsaturated Fatty Acid Methyl Esters by SMCI Method The compounds called fatty acids are measured in wide- ranging areas from foods to the biological samples. Analysis using the gas chromatograph (GC) is a suitable measurement technique, as GC has a high separation capability in the chromatogram. In this technique, the fatty acids are first derivatized by methyl esterification to produce inactivated fatty acid methyl esters (FAME) with ample volatility. However, because localization of the double bond position in unsaturated FAMEs is difficult, even when using mass spectrometry is used, research on this issue was limited. To overcome this difficulty, an analysis technique based on chemical ionization utilizing acetonitrile was developed (1) , and expanded the possibilities of double bond localization. Solvent mediated chemical ionization (SMCI) is a soft ionization method using an organic solvent, and use of the new Shimadzu SMCI unit makes it possible to introduce acetonitrile, which is necessary in this double bond localization technique, into the analytical instrument. This article introduces the results of an analysis of unsaturated FAMEs by the SMCI method. For details of the analytical method and real sample analysis, refer to Shimadzu Technical Reports C146-0396 and C146-0397. R. Kitano Samples and Analysis Conditions A 50 ng/mL mixed standard solution of methyl oleate, methyl linoleate, and methyl linolenate was prepared, and measurement was carried out under the conditions in Table 1 using the SMCI method. Table 1 Instrument and Analysis Conditions Instrument GCMS : GCMS-TQ™ 8040 NX Autosampler : AOC-20i+s Column : BPX-70 (L: 25 m, df: 0.25 μm, ID: 0.22 mm) Insert liner : Split-less deactivated liner w/low wool GC conditions Injection temp. : 250 °C Injection volume : 1 μL Injection mode : Split-less Carrier gas control : Linear velocity (46.2 cm/s) Column oven temp. : 80 °C Æ (15 °C/min) Æ 170 °C (4 min) Æ (7 °C/min) Æ 240 °C (10 min) MS conditions Interface temp. : 240 °C Ion source temp. : 230 °C Ionization method : SMCI (acetonitrile) Measurement mode : Scan, product ion scan (CE = 6 V) Event time : 0.5 s Scan Mass Spectra With the conventional CI method, mainly the peak [M+1] originating from the addition of protons to the molecular weight is observed. However, with the SMCI method using acetonitrile, the peak [M+54] is obtained by addition of the reactant ions produced by the interaction with acetonitrile. Since the reactant ions are added selectively to the double bond, the adduct ions can be confirmed from mass spectra of unsaturated FAMEs, as shown in Fig. 1. Double bond localization is done by conducting a product ion scan analysis for this [M+54] ion. Fig. 1 SMCI Scan Mass Spectra Methyl oleate Methyl linoleate Methyl linolenate M+54 M+54 M+54 100.0 125.0 150.0 175.0 200.0 225.0 250.0 275.0 300.0 325.0 350.0 375.0 400.0 0 25 50 75 100 % 297 265 336 350 318 247 276 222 111 125 165 137 180 151 199 100.0 125.0 150.0 175.0 200.0 225.0 250.0 275.0 300.0 325.0 350.0 375.0 400.0 0 25 50 75 100 % 263 295 348 334 245 316 109 137 151 199 123 179 213 373 243 277 100.0 125.0 150.0 175.0 200.0 225.0 250.0 275.0 300.0 325.0 350.0 375.0 400.0 0 25 50 75 100 % 293 346 261 108 332 123 243 137 223 151 165 314 291 191 244 209
First Edition: May 2020 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. Shimadzu disclaims any proprietary interest in trademarks and trade names used in this publication other than its own. See http://www.shimadzu.com/about/trademarks/index.html for details. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. © Shimadzu Corporation, 2020 www.shimadzu.com/an/ Application News No. M294 Product Ion Scan Mass Spectra Fig. 2 shows the results of the product ion scan for the precursor ions, the [M+54] confirmed in Fig. 1. Two distinctive ions can be confirmed in each of the unsaturated FAMEs, and it is possible to trace back to the double bond position from these values. The reactant ions originating from acetonitrile are added to the double bond, and cleavage due to CID occurs at the corresponding locations. Because the reactant ions are added from two directions, two distinctive ions can be confirmed. These correspond to the peaks indicated by the red numbers in the mass spectra. In mono- and bivalent unsaturated FAMEs, ions generated by cleavage on each side of the double bond can be observed, and in a trivalent unsaturated FAME, the two end double bonds existing with methylene group, cleavage occurs between the two double bonds. Thus, designated ions can be observed corresponding to the number of double bonds and their coordination sites. Fig. 2 SMCI Product Ion Scan Mass Spectra (The red numbers (upper numbers) in the structural formulas are the values with the reactant ion observed in mass spectra, and the black numbers (lower numbers) are the values without the reactant ion.) Conclusion By using the SMCI method, it is possible not only to confirm the molecular weight of unsaturated fatty acid methyl esters, but also to identify the positions of their double bonds, which had been difficult to analyze and required special derivatization when using the conventional techniques. SMCI can also be applied to compounds for which standards do not exist, and is useful in structural analysis of unknown compounds. <Reference> (1) Van Pelt, C.K. and J.T. Brenna, Acetonitrile chemical ionization tandem mass spectrometry to locate double bonds in polyunsaturated fatty acid methyl esters. Anal Chem, 1999, 71(10): p. 1981-9. GCMS-TQ is a trademark of Shimadzu Corporation in Japan and/or other countries. 208 252 192 278 148 276 Methyl oleate Methyl linoleate Methyl linolenate 208 153 252 197 192 137 278 223 148 95 276 223 50 100 150 200 250 300 350 0 1 2 3 4 5 6 7 % 350 318 194 222 301 93 71 125 57 234 280 151 359 331 50 100 150 200 250 300 350 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 % 348 316 67 150 131 292 220 83 250 178 164 236 106 307 55 357 50 100 150 200 250 300 350 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 % 346 314 159 204 136 109 190 290 81 95 226 254 55 235 318 356
Please fill out the survey Related Products Some products may be updated to newer models. GCMS-TQ RX Series Triple Quadrupole Gas Chromatograph Mass Spectrome… AOC-20i/AOC-20s Auto Injector/Auto Sampler for GC/GC- MS Related Solutions Food and Beverages Food and Nutrition Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)