Application
News
User Benefits
No. X278
Elemental Analysis in Active Pharmaceutical
Ingredients (APIs) Development Using EDX:
Residual Catalysts, Drug Salts, Contaminants
Chemical pretreatment of APIs is not necessary and measurement of samples as little as 0.1 g is possible.
Unlike ICP-MS which requires complicated sample preparation, EDX requires minimal sample preparation.
Various applications such as analysis of residual catalysts, drug salts, and contaminants are possible using EDX.
T. Nakao, H. Moriya
Shimadzu EDX-7000 Energy Dispersive X-ray Fluorescence Spectrometer
Introduction
Elemental analyses are carried out for various reasons during
API development and manufacturing processes. For example, in
the development stage, quick and simple analysis of target
elements in drug salts and elements regulated in ICH Q3D
(1)
and
Japanese Pharmacopoeia
(2)
is required. During acceptance and
pre-shipment testing, there is also a need to check for
contamination that may have occurred during the manufacturing
process. Analysis by EDX is simple, does not require special
sample preparation skills, and can easily be analyzed by anyone.
Startup of the instrument is fast, and analysis can be carried out
by just placing the sample as-is into a dedicated container. EDX
can non-destructively analyze a small amount of sample with
sufficient sensitivity. Hence it is especially useful for API
development where samples are little and valuable.
As examples of analysis during API development, this article
introduces the following examples of EDX analysis:
1. Residual metal catalysts
It is shown that quantitative analysis of Ir, Pt, Ru, Rh, Pd, and Os is
possible with only 0.1 g of sample.
2. Drug Salts
Cl, Br, and S were quantified by assuming the presence of counter
ions.
3. Contamination
Contaminants that could enter during the manufacturing process
were analyzed using the EDXIR- Analysis™ software for EDX and
FTIR.
Fig. 1 Samples
Standard Samples for Calibration
Sample solutions with 6 concentration levels were prepared using
the following two commercially available reagents. Table 1 shows
the concentrations.
(A) USP-TXM4 (SPEX CertiPrep): Ir, Pt, Ru, Rh, Pd
(B) Os 1000 (Osmium standard stock solution) (Kanto
Chemical Co.,Inc.): Os
Table 1 Concentrations of Standard Samples
1. Residual Catalysts
Elements
Under the ICH Q3D guidelines, the following Class 2B elements
must be measured if they are intentionally added into
pharmaceutical products. These elements were evaluated by
the calibration curve method.
77
Ir,
78
Pt,
44
Ru,
45
Rh,
46
Pd,
76
Os
Samples
The following two commercially available API powders and
cellulose powder were used. Considering the possibility of
reaction with the APIs, cellulose was used for the analysis of Os,
as it has a composition similar to the drug material.
① Benazepril hydrochloride
(3)
② Captopril
(3)
③ Cellulose powder
[μg/g]
Ultrapure
water
STD1 STD2 STD3 STD4 STD5
Ir, Pt, Ru, Rh, Pd 0 5 10 20 50 100
Os 0 5 10 20 50 100
Verification Samples
The verification samples were prepared by adding standard reagent
(A) to samples ① and ② and adding standard reagent (B) to
sample ③, and mixing the solutions until uniform
(3)
. Verification
samples with the following three concentration levels were
prepared.
10, 5, 0 (blank; no addition) [μg/g]
Sample Preparation
A constant volume of the samples, as shown below, was
introduced into a sample container lined with polypropylene
film and measured as-is. Fig. 1 shows the appearance of the
samples.
Standard samples: 8 mL
Verification samples: 2.0 g (sufficient to assume bulk thickness)
0.5, 0.3, 0.1 g (small amounts)
Verification sample (2.0 g) Standard sample