Application
News
www.shimadzu.com/an/
Shimadzu Corporation
Analytical & Measuring Instruments Division
Global Application Development Center
© Shimadzu Corporation, 2021
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
See http://www.shimadzu.com/about/trademarks/index.html for details.
Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they
are used with trademark symbol “TM” or “”.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or
completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this
publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change
without notice.
First Edition: Oct. 2021 01-00029-EN
Fig. 5 Chromatogram of Roxatidine Acetate Hydrochloride
Extended-Release Capsules (Dissolution Time of 8 Hours)
Table 4 Dissolution Conditions
Conclusion
Online dissolution test of roxatidine acetate hydrochloride
extended-release capsules was carried out. The dissolution test
was able to be carried out more easily than conventional
methods. In this case, about 35% of the total time required for
the entire processes was able to be reduced.
Fig. 6 Dissolution Curve
Nexera, LabSolutions and Shim-pack are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries.
Dissolution Test
Fig. 5 shows the chromatogram of the dissolution medium from
commercially available roxatidine acetate hydrochloride
extended-release capsules (labeled amount: 75 mg per capsule,
dissolution time: 8 hours). Table 4 shows the dissolution
conditions. HPLC analytical conditions are the same as those in
Table 1. The conditions for dissolution and HPLC analysis were
determined by referring to the test conditions listed in the
Japanese Pharmacopoeia, 18th edition, "Roxatidine Acetate
Hydrochloride Extended-Release Capsules". Fig. 6 shows the
dissolution curve, Table 5 shows the results of the dissolution
test, and Table 6 shows the tolerances for the dissolution test of
roxatidine acetate hydrochloride extended-release capsules
listed in the Japanese Pharmacopoeia, 18th edition, as reference
values. In this test, we used sinkers for dissolution. The
dissolution medium was automatically sampled and filtered at
specified timing, automatically diluted 2 times with water using
Nexera FV, and then analyzed.
mAU
Roxatidine Acetate Hydrochloride
12.5
10.0
7.5
5.0
2.5
0.0
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 min
System : NTR-6600AST (TOYAMA SANGYO CO., LTD.)
Dissolution method : Paddle
Dissolution media : Pure water
Media volume : 900 mL
Rotation speed : 50 rpm
Bath temperature : 37 ˚C
Total time : 480 min
Sampling time : 15, 30, 60, 90, 120, 300 and 480 min
100
80
60
40
20
0
0 100 400 500
Dissolution rate* (%)
1
Dissolution time (min)
4 6
* Dissolution rate (%) = Concentration (mg/L) × Media volume 0.9 (L) / Labeled
amount 75 (mg) × 100
Vessel No.
200 300
5 3 2
Table 6 Tolerances of the Dissolution Test
(Japanese Pharmacopoeia, 18th Edition)
Time (min) Results
*
60 20 to 50%
90 35 to 65%
480 ≥ 70%
Table 5 Dissolution Rate of 75 mg Capsule (%)
Time (min)
15 30 60 90 120 300 480
Vessel No.
1 4.74 14.46 35.98 48.40 55.33 73.81 81.30
2 5.08 15.12 34.92 45.61 54.09 72.33 79.39
3 4.76 14.85 38.09 50.43 57.30 75.84 83.63
4 4.57 13.76 35.62 47.59 54.17 72.97 80.06
5 5.46 16.70 37.14 48.52 55.72 72.93 79.42
6 4.38 13.40 35.84 48.03 55.71 73.60 81.83
Judgement PASSED PASSED PASSED
* Condition: analyzing 100 μL of dissolution medium