User Benefits Application News Online Dissolution Testing of Roxatidine Acetate Hydrochloride Extended-Release Capsules Dissolution testing can be automated. The automatic sampling function enables testing with good reproducibility. The automatic dilution function improves work efficiency. K. Matsuoka, A. Tanabe Ultra High Performance Liquid Chromatograph Nexera™ FV Introduction Dissolution testing is conducted for formulation development, quality control, bioequivalence tests of generic drugs, etc. In dissolution testing, the dissolution properties of a drug product are checked under specific conditions for certain periods. Dissolution testing takes a lot of labor and time, since dissolution media must be sampled from multiple vessels every sampling time and analyzed. Nexera FV is an HPLC system for online dissolution testing. It can automate the processes from sampling dissolution media, HPLC analysis, up to report output. By automating tasks that used to be conducted manually by operators, the system realizes labor saving and throughput improvement. In addition, automating tasks prevent human errors from whole processes. This article introduces the workflow of online dissolution testing of roxatidine acetate hydrochloride extended-release capsules, using Nexera FV. Nexera FV has two modes of analysis: direct injection mode to inject the dissolution media delivered from the dissolution tester directly to the HPLC, and fraction analysis mode to fractionate the dissolution media into vials and analyze them. The former is effective when an analysis is enough short to complete by the next sampling time. The latter is used in tests with short sampling intervals, and in cases where dilution or addition of internal standards is required. Furthermore, the fraction analysis mode also makes it possible to conduct HPLC analyses between sampling intervals. As a result, the total time could be reduced by 35%. Fig. 1 Comparison of Workflow of Online Dissolution Testing *2 Online Dissolution Testing with Nexera FV Fig. 1 shows a comparative example of the workflow of conventional method and online dissolution testing with Nexera FV. The Nexera FV system can automate sampling of dissolution medium at designated times, filtration, dilution, addition of internal standards, HPLC analysis, and report generation processes which had been conducted manually. Furthermore, the HPLC analytical conditions can be set easily by the dedicated software DT-Solution (Fig. 2), and a report summarizing multiple data such as the dissolution rate is prepared simultaneously with completion of the analysis (Fig. 3). Fig. 2 DT-Solution Setting Window Fig. 3 Multi-Data Report *1 Window *1 Multi-Data Report is an optional function of LabSolutions™DB/CS to generate reports automatically. *2 The time above is an example of roxatidine acetate hydrochloride extended-release capsules with a dissolution test of 8 hours and an HPLC analysis of 5.5 hours (42 analyses of 8 minutes each). Preparation of dissolution tester 2 hours Eluate sampling 8 hours Filtration+dilution+HPLC analysis 7 hours Report creation 1 hour Cleanup 0.5 hours Nexera FV (Fraction analysis mode) 9.5 hours Cleanup 0.5 hours Conventional method Total 18.5 hours 35% reduction Nexera FV Total 12 hours Preparation of dissolution tester 2 hours
Application News System Suitability Test Table 1 shows the analytical conditions, and Fig. 4 shows the chromatogram of the standard roxatidine acetate hydrochloride. Table 2 then shows the results of the system suitability test. The analytical conditions were determined by referring to the test conditions listed in the Japanese Pharmacopoeia, 18th edition, "Roxatidine Acetate Hydrochloride Extended-Release Capsules". For reference, Table 3 shows the suitability requirements described in the Japanese Pharmacopoeia, 18th edition. The system suitability was confirmed by the repeated analyses at 42 mg/L standard solution (n=6) under the conditions shown in Table 1. Both the system performance and reproducibility meet requirements described in the 18th edition of the Japanese Pharmacopoeia. Column : Shim-pack™ VP-ODS *1 (150 mm × 4.6 mm I.D., 5 µm) Mobile phase : Water / Acetonitrile / Triethylamine / Acetic Acid = 340 : 60 : 2 : 1 Flow rate : 0.8 mL/min Column temp. : 40 ˚C Injection vol. : 20 µL Vial : Shimadzu Vial, LC, 1.1 mL, Glass *2 Detection : UV 274 nm Table 1 HPLC Conditions 15.0 mAU Roxatidine Acetate Hydrochloride 12.5 10.0 7.5 5.0 2.5 0.0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 min Fig. 4 Chromatogram of Standard Roxatidine Acetate Hydrochloride (42 mg/L) *1 P/N: 228-34937-91 *2 P/N: 228-21283-91 Table 3 System Suitability Test Requirements (Japanese Pharmacopoeia, 18th Edition) System Repeatability Area %RSD ≤ 1.0% *3 System Performance Theoretical Plate Number ≥ 3000 *4 and Symmetry Factor ≤ 2.0 *4 *3 Condition: analyzing 100 μL of 42 mg/L standard solution repeatedly (n=6) *4 Condition: analyzing 100 μL of 42 mg/L standard solution Table 2 Results of System Suitability Test (42 mg/L) Retention Time (min) System Reproducibility System Performance Area Theoretical Plate Number Symmetry Factor 1 st 5.077 149282 7362 1.15 2 nd 5.078 149485 7398 1.15 3 rd 5.079 149358 7353 1.15 4 th 5.082 249368 7372 1.15 5 th 5.081 149885 7371 1.15 6 th 5.083 149107 7372 1.15 Averages 5.08 149248 - - %RSD 0.05 0.15 - - Judgement PASSED PASSED PASSED
Application News www.shimadzu.com/an/ Shimadzu Corporation Analytical & Measuring Instruments Division Global Application Development Center © Shimadzu Corporation, 2021 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. First Edition: Oct. 2021 01-00029-EN Fig. 5 Chromatogram of Roxatidine Acetate Hydrochloride Extended-Release Capsules (Dissolution Time of 8 Hours) Table 4 Dissolution Conditions Conclusion Online dissolution test of roxatidine acetate hydrochloride extended-release capsules was carried out. The dissolution test was able to be carried out more easily than conventional methods. In this case, about 35% of the total time required for the entire processes was able to be reduced. Fig. 6 Dissolution Curve Nexera, LabSolutions and Shim-pack are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries. Dissolution Test Fig. 5 shows the chromatogram of the dissolution medium from commercially available roxatidine acetate hydrochloride extended-release capsules (labeled amount: 75 mg per capsule, dissolution time: 8 hours). Table 4 shows the dissolution conditions. HPLC analytical conditions are the same as those in Table 1. The conditions for dissolution and HPLC analysis were determined by referring to the test conditions listed in the Japanese Pharmacopoeia, 18th edition, "Roxatidine Acetate Hydrochloride Extended-Release Capsules". Fig. 6 shows the dissolution curve, Table 5 shows the results of the dissolution test, and Table 6 shows the tolerances for the dissolution test of roxatidine acetate hydrochloride extended-release capsules listed in the Japanese Pharmacopoeia, 18th edition, as reference values. In this test, we used sinkers for dissolution. The dissolution medium was automatically sampled and filtered at specified timing, automatically diluted 2 times with water using Nexera FV, and then analyzed. mAU Roxatidine Acetate Hydrochloride 12.5 10.0 7.5 5.0 2.5 0.0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 min System : NTR-6600AST (TOYAMA SANGYO CO., LTD.) Dissolution method : Paddle Dissolution media : Pure water Media volume : 900 mL Rotation speed : 50 rpm Bath temperature : 37 ˚C Total time : 480 min Sampling time : 15, 30, 60, 90, 120, 300 and 480 min 100 80 60 40 20 0 0 100 400 500 Dissolution rate* (%) 1 Dissolution time (min) 4 6 * Dissolution rate (%) = Concentration (mg/L) × Media volume 0.9 (L) / Labeled amount 75 (mg) × 100 Vessel No. 200 300 5 3 2 Table 6 Tolerances of the Dissolution Test (Japanese Pharmacopoeia, 18th Edition) Time (min) Results * 60 20 to 50% 90 35 to 65% 480 ≥ 70% Table 5 Dissolution Rate of 75 mg Capsule (%) Time (min) 15 30 60 90 120 300 480 Vessel No. 1 4.74 14.46 35.98 48.40 55.33 73.81 81.30 2 5.08 15.12 34.92 45.61 54.09 72.33 79.39 3 4.76 14.85 38.09 50.43 57.30 75.84 83.63 4 4.57 13.76 35.62 47.59 54.17 72.97 80.06 5 5.46 16.70 37.14 48.52 55.72 72.93 79.42 6 4.38 13.40 35.84 48.03 55.71 73.60 81.83 Judgement PASSED PASSED PASSED * Condition: analyzing 100 μL of dissolution medium
Related Products Some products may be updated to newer models. Nexera FV Ultra High Performance Liquid Chromatograph for On… Shim-pack VP Series HPLC Column Related Solutions Small Molecule Pharmaceutical Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)