Application
News
Addition of ISTD Using Sample Preparation
Function
In this article, the acetaminophen, aspirin, and caffeine in an
antipyretic analgesic tablet were determined using the internal
standard method. The sample preparation function of a
Shimadzu SIL-30ACFV autosampler was used to add the ISTD to
the standard sample and the fractions of dissolution medium.
Fig. 5 shows the flow of ISTD addition using the autosampler,
and Table 1 shows the sample preparation program. Fig. 4
shows an example of the vial arrangement in the autosampler
when the sample preparation program in Table 1 is carried out.
The ISTD vial (❶) is set in the control vial rack, and the flow vials
which receive the dissolution media from the dissolution tester
are set in the flow vial rack before starting the test. As
dissolution medium fractionation vials (❷) and dilution/ISTD
addition vials (❸), empty vials are set at the positions shown in
Fig. 4. When the test is started and the specified time is reached,
(1) the dissolution media from the dissolution tester are
pumped to the flow vials, and (2) the dissolution media are
fractionated to the dissolution medium fractionation vials (❷)
by the autosampler. Following this, (3) the specified amounts of
the ISTD in the ISTD vial (❶) and the fractionated dissolution
medium in the fractionation vials (❷) are suctioned with a
needle, and are then discharged together with the diluent into
the dilution/ISTD addition vials (❸) and mixed. Finally, (4) 10 μL
of the dissolution medium with addition of the prepared diluted
ISTD by (3) is injected from the dilution/ISTD addition vials (❸)
into the injection port (❹). These procedures of (❸),(❹) are
carried out for all fractions. The ISTD used here was a methanol
solution (before addition 3600 mg/L) of benzoic acid, and the
diluent was methanol/glacial acetic acid = 95 : 5.
System Suitability Test
Table 2 shows the analytical conditions, and Fig. 6 shows the
chromatogram of the mixed standard sample of
acetaminophen, acetylsalicylic acid, and caffeine. It is thought
that the sample also contained a small amount of salicylic acid
which was formed by hydrolysis of the acetylsalicylic acid, and
this appeared as a peak.
Table 3 shows the result of the system suitability test. The test
conditions are described in the United States Pharmacopeia
USP43-NF38 Acetaminophen, Aspirin, and Caffeine Tablets
DISSOLUTION
(1)
. Table 4 shows the requirements of the system
suitability test in USP43-NF38. System suitability was confirmed
by the repeated analyses (n=6) of a 100 mg/L solution under the
conditions shown in Table 2. Both the performance and the
reproducibility of the system satisfied the requirements of
USP43-NF38.
Flow vial rack
Control vial rack
❶
ISTD vial
❷ Dissolution medium
fractionation vials
∙ ∙ ∙
❸ Dilution/ISTD
addition vials
❹
Injection
port
(3)
(3)
(4)
(1)
(2)
1 n.drain
2 disp 600,rs
3 vial.n 0,9
4 n.strk 52
5 aspir 10,ss
6 air.a 0.1,ss
7 d.rinse
8 a0=sn+6
9 vial.n 1,a0
10 n.strk ns
11 aspir 20,ss
12 vial.n rn,sn
13 n.strk ns
14 disp 100,ss
15 mix 3,5,45,2,10
16 n.drain
17 disp 600,rs
18 d.rinse
19 inj.p
20 v.inj
21 wait 2.0
22 goto f0
23 end
From dissolution tester
Table 1 Sample Preparation Program
Table 2 HPLC Conditions
Column Shim-pack™ Scepter C18-120
*1
(100 mm × 4.6 mm I.D., 5 µm)
Mobile phase Methanol / Glacial Acetic Acid / Water
= 28 : 3 : 69
Flow rate 2 mL/min
Column temp. 45 ˚C
Injection vol. 10 µL
Vial Shimadzu Vial, LC, 1.1 mL, Glass
*2
Shimadzu Vial, LC
*3
Detection UV 275 nm
*1 P/N : 227-31020-04 *2 P/N : 228-21283-91
*3 P/N : 228-31600-91
0.0 1.0 2.0 3.0 4.0 5.0 min
0
50
100
150
200
250
300
mAU
Acetaminophen
Caffeine
Acetylsalicylic acid
Benzoic acid
Salicylic acid
4.0 5.0 min
-1.00
0.00
1.00
2.00
Benzoic acid
Salicylic acid
Enlarged View
❸
Dilution/ISTD
addition vial
❶
ISTD vial
Suction of
10 μL of
ISTD
❷
Dissolution medium
fractionation vial
Suction of
20 μL of
dissolution
medium
Suction
of 0.1 μL
of air
Injection of
10 μL
❹
Injection port
Discharge of
ISTD +
dissolution
medium +
diluent
(total:100 μL)
Mixing
Fig. 4 Example of Vial Arrangements in Autosampler
Fig. 5 Flow of Internal Standard (ISTD) Addition by Autosampler
Fig. 6 Chromatograms of Mixed Standard (each 100 mg/L) of Acetaminophen,
Acetylsalicylic Acid and Caffeine