User Benefits Application News USP-Compliant Online Dissolution Testing of Antipyretic Analgesic: Automatic Addition of Internal Standard Dissolution testing can be automated. The automatic sampling function enables testing with good reproducibility. The automatic dilution and the automatic addition of internal standard improve work efficiency. K. Matsuoka, A. Tanabe Ultra High Performance Liquid Chromatograph Nexera™ FV Introduction Dissolution testing is conducted for formulation development, quality control, bioequivalence tests of generic drugs, etc.. In dissolution testing, the dissolution properties of a drug product are checked under specific conditions for certain periods. Dissolution testing takes a lot of labor and time, since dissolution media must be sampled from multiple vessels every sampling time and analyzed. Nexera FV is an HPLC system for online dissolution testing. It can automate the processes from sampling dissolution media, HPLC analysis, up to report output. By automating tasks that used to be conducted manually by operators, the system realizes labor saving and throughput improvement. In addition, automating tasks prevent human errors from whole processes. This article introduces an example of automatic addition of the internal standard (ISTD) to the dissolution medium and a USP- compliant online dissolution testing of an antipyretic analgesic tablet using the Nexera FV. For additional information on online dissolution testing, please refer to Application News 01-00029 and 01-00031. Fig. 1 Comparison of Workflow of Online Dissolution Testing *2 Nexera FV has two modes of analysis: direct injection mode to inject the dissolution media delivered from the dissolution tester directly to the HPLC, and fraction analysis mode to fractionate the dissolution media into vials and analyze them. The former is effective when an analysis is enough short to complete by the next sampling time. The latter is used in tests with short sampling intervals, and in cases where dilution or addition of internal standards is required. As described in this article, use of the Nexera FV fraction analysis mode to automatically add the ISTD made it possible to achieve high efficiency in work that had been done manually in the conventional process. Online Dissolution Testing with Nexera FV Fig. 1 shows a comparative example of the workflow of conventional method and online dissolution testing with Nexera FV. The Nexera FV system can automate sampling of dissolution medium at designated times, filtration, dilution, addition of internal standards, HPLC analysis, and report generation processes which had been conducted manually. Furthermore, the HPLC analytical conditions can be set easily by the dedicated software DT-Solution (Fig. 2), and a report summarizing multiple data such as the dissolution rate is prepared simultaneously with completion of the analysis (Fig. 3). Preparation of dissolution tester 2 hours Sampling of dissolution medium 1 hour Filtration+dilution+ISTD addition+HPLC analysis 1.5 hours Preparation of report 1 hour Cleanup 0.5 hours Preparation of dissolution tester 2 hours Nexera FV (Fraction analysis mode) 2 hours Cleanup 0.5 hours Nexera FV Total 4.5 hours Conventional method Total 6.0 hours 25 % reduction Fig. 3 Multi-Data Report *1 Windows Fig. 2 DT-Solution Setting Window *1 Multi-Data Report is an optional function of LabSolutions™DB/CS to generate reports automatically. *2 The time above is an example of an antipyretic analgesic with a dissolution test of 1 hour and an HPLC analysis of 0.5 hours (6 analyses of 6 minutes each).
Application News Addition of ISTD Using Sample Preparation Function In this article, the acetaminophen, aspirin, and caffeine in an antipyretic analgesic tablet were determined using the internal standard method. The sample preparation function of a Shimadzu SIL-30ACFV autosampler was used to add the ISTD to the standard sample and the fractions of dissolution medium. Fig. 5 shows the flow of ISTD addition using the autosampler, and Table 1 shows the sample preparation program. Fig. 4 shows an example of the vial arrangement in the autosampler when the sample preparation program in Table 1 is carried out. The ISTD vial (❶) is set in the control vial rack, and the flow vials which receive the dissolution media from the dissolution tester are set in the flow vial rack before starting the test. As dissolution medium fractionation vials (❷) and dilution/ISTD addition vials (❸), empty vials are set at the positions shown in Fig. 4. When the test is started and the specified time is reached, (1) the dissolution media from the dissolution tester are pumped to the flow vials, and (2) the dissolution media are fractionated to the dissolution medium fractionation vials (❷) by the autosampler. Following this, (3) the specified amounts of the ISTD in the ISTD vial (❶) and the fractionated dissolution medium in the fractionation vials (❷) are suctioned with a needle, and are then discharged together with the diluent into the dilution/ISTD addition vials (❸) and mixed. Finally, (4) 10 μL of the dissolution medium with addition of the prepared diluted ISTD by (3) is injected from the dilution/ISTD addition vials (❸) into the injection port (❹). These procedures of (❸),(❹) are carried out for all fractions. The ISTD used here was a methanol solution (before addition 3600 mg/L) of benzoic acid, and the diluent was methanol/glacial acetic acid = 95 : 5. System Suitability Test Table 2 shows the analytical conditions, and Fig. 6 shows the chromatogram of the mixed standard sample of acetaminophen, acetylsalicylic acid, and caffeine. It is thought that the sample also contained a small amount of salicylic acid which was formed by hydrolysis of the acetylsalicylic acid, and this appeared as a peak. Table 3 shows the result of the system suitability test. The test conditions are described in the United States Pharmacopeia USP43-NF38 Acetaminophen, Aspirin, and Caffeine Tablets DISSOLUTION (1) . Table 4 shows the requirements of the system suitability test in USP43-NF38. System suitability was confirmed by the repeated analyses (n=6) of a 100 mg/L solution under the conditions shown in Table 2. Both the performance and the reproducibility of the system satisfied the requirements of USP43-NF38. Flow vial rack Control vial rack ❶ ISTD vial ❷ Dissolution medium fractionation vials ∙ ∙ ∙ ❸ Dilution/ISTD addition vials ❹ Injection port (3) (3) (4) (1) (2) 1 n.drain 2 disp 600,rs 3 vial.n 0,9 4 n.strk 52 5 aspir 10,ss 6 air.a 0.1,ss 7 d.rinse 8 a0=sn+6 9 vial.n 1,a0 10 n.strk ns 11 aspir 20,ss 12 vial.n rn,sn 13 n.strk ns 14 disp 100,ss 15 mix 3,5,45,2,10 16 n.drain 17 disp 600,rs 18 d.rinse 19 inj.p 20 v.inj 21 wait 2.0 22 goto f0 23 end From dissolution tester Table 1 Sample Preparation Program Table 2 HPLC Conditions Column Shim-pack™ Scepter C18-120 *1 (100 mm × 4.6 mm I.D., 5 µm) Mobile phase Methanol / Glacial Acetic Acid / Water = 28 : 3 : 69 Flow rate 2 mL/min Column temp. 45 ˚C Injection vol. 10 µL Vial Shimadzu Vial, LC, 1.1 mL, Glass *2 Shimadzu Vial, LC *3 Detection UV 275 nm *1 P/N : 227-31020-04 *2 P/N : 228-21283-91 *3 P/N : 228-31600-91 0.0 1.0 2.0 3.0 4.0 5.0 min 0 50 100 150 200 250 300 mAU Acetaminophen Caffeine Acetylsalicylic acid Benzoic acid Salicylic acid 4.0 5.0 min -1.00 0.00 1.00 2.00 Benzoic acid Salicylic acid Enlarged View ❸ Dilution/ISTD addition vial ❶ ISTD vial Suction of 10 μL of ISTD ❷ Dissolution medium fractionation vial Suction of 20 μL of dissolution medium Suction of 0.1 μL of air Injection of 10 μL ❹ Injection port Discharge of ISTD + dissolution medium + diluent (total:100 μL) Mixing Fig. 4 Example of Vial Arrangements in Autosampler Fig. 5 Flow of Internal Standard (ISTD) Addition by Autosampler Fig. 6 Chromatograms of Mixed Standard (each 100 mg/L) of Acetaminophen, Acetylsalicylic Acid and Caffeine
Application News www.shimadzu.com/an/ Shimadzu Corporation Analytical & Measuring Instruments Division Global Application Development Center © Shimadzu Corporation, 2022 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. First Edition: Mar. 2022 01-00030-EN Table 5 Dissolution Conditions System NTR-6600AST (TOYAMA SANGYO CO., LTD.) Dissolution method Paddle Dissolution media Pure water Media volume 900 mL Rotation speed 100 rpm Bath temperature 37 ˚C Total time 60 min Sampling time 60 min Dissolution Test Fig. 7 shows the chromatogram of the dissolution medium from the commercially-available antipyretic analgesic tablet. Table 5 shows the dissolution conditions. The HPLC analytical conditions are the same as that shown in Table 2. The dissolution media were sampled automatically and filtered at the specified time, the ISTD was added automatically by using the Nexera FV, and then analyzed. Table 6 shows the dissolution rates of each component at a dissolution time of 60 min. USP43-NF38 Acetaminophen, Aspirin, and Caffeine Tablets DISSOLUTION requires dissolution of 75 % of the label contents of each of the three components acetaminophen, acetylsalicylic acid, and caffeine within the dissolution time of 60 min. In this test, the results were within the tolerance required in the USP, as dissolution rates of 75 % or more were achieved in the dissolution time of 60 min with all components and all vessels. Conclusion A USP-compliant online dissolution test was carried out using a commercially-available antipyretic analgesic tablet containing acetaminophen, acetylsalicylic acid, and caffeine. In comparison with the conventional method, a large reduction in work time and increase in work efficiency were achieved by automatic dilution and addition of the internal standard using the Nexera FV. As a result, it was found that the time required for the total process can be reduced by approximately 25 %. Fig. 7 Chromatograms of Antipyretic Analgesic (Dissolution Time : 60 min) Suitability requirements Target Compound Internal Standard Judgement Acetaminophen Caffeine Acetylsalicylic acid Benzoic Acid Resolution (n=6) ≥1.4 19.32 15.73 4.66 PASSED Tailing factor (n=6) ≤1.2 0.87 0.98 0.94 0.97 PASSED Retention Time (%RSD) ≤2.0 0.08 0.07 0.28 0.17 PASSED Area (%RSD) ≤2.0 0.36 0.38 0.35 0.44 PASSED Table 4 Requirements of System Suitability Test (USP43-NF38) Resolution ≥1.4 between any of the analyte and ISTD Tailing factor ≤1.2 for each analyte peak Relative standard deviation ≤2.0 % Table 3 Results of System Suitability Test (each 100 mg/L) 0.0 1.0 2.0 3.0 4.0 5.0 min 0 20 40 60 80 mAU Acetaminophen Caffeine Acetylsalicylic acid Benzoic acid Salicylic acid 4.00 5.00 min -1.00 0.00 1.00 2.00 Benzoic acid Salicylic acid Enlarged View Acetaminophen Caffeine Acetylsalicylic acid 1 109.3 123.1 104.6 2 106.4 121.5 109.7 3 107.7 122.9 105.6 4 108.9 119.3 103.7 5 113.2 131.8 102.8 6 110.7 121.2 103.2 Judgement PASSED PASSED PASSED Compound Vessel No. * Dissolution rate (%) = Concentration (mg/L) × Media volume 0.9 (L) / Labeled amount (mg) × 100 Table 6 Dissolution Rates* in Dissolution Test of Antipyretic Analgesic Tablets (Dissolution Time : 60 min, %) Nexera, LabSolutions, and Shim-pack Scepter are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries. <Reference> (1) USP43-NF38-49 “Acetaminophen, Aspirin, and Caffeine Tablets DISSOLUTION”
Related Products Some products may be updated to newer models. Nexera FV Ultra High Performance Liquid Chromatograph for On… Shim-pack Scepter LC Columns HPLC Column Related Solutions Development (Formulation, Scale- up, Method develop Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)