User Benefits
Application
News
Quantitation of NDMA and NDEA in Metformin and 5
Sartan APIs as per the EDQM method Procedure B
This application note demonstrates applicability of EDQM method (Procedure B) for Metformin in addition to Sartan APIs.
The GCMS-QP2020 NX with AOC-30i system achieves LOQ lower than the EDQM method (Procedure B) meeting the defined criteria of
S/N and recovery.
Introduction
Overview : The Drug Regulatory Authorities first noticed the
presence of the N-Nitrosamine impurity (NSA), N-
Nitrosodimethylamine (NDMA) in products containing valsartan
in July 2018. Valsartan is an Angiotensin II Receptor Blocker
(ARB) and belongs to a family of analogue compounds
commonly referred to as the Sartans. Similarly, NSA has also
been detected in other drug products such as Metformin.
Metformin is a prescription drug used to control high blood
sugar in patients with Type 2 diabetes. Considering the
significance of these drugs, it is crucial to make Sartans and
Metformin available with safe levels of NSA.
Like USFDA, European Directorate for the Quality of Medicines
and Healthcare (EDQM) ensure access to good quality medicines
in Europe. EDQM has been working actively at various levels to
address the presence of Nitrosamines in active substances and
medicines. EDQM has been regularly informing all stakeholders,
from national authorities to manufacturers, on the state of the
works and on initiatives taken.
EDQM procedure for NSA: EDQM enlists 3 procedures for
determination of NSA viz procedure A, B and C for LC-MS/MS,
GC-MS and GC-MS/MS, respectively. Procedures A and B have
been validated as limit tests (30 ppb) with recovery
demonstrated at limit level. The procedure C has been validated
as a quantitative test wherein a three-point calibration is
plotted with recovery performed at limit level (30 ppb). This
application news describes analytical procedures for the
detection of 2 commonly found N-Nitrosamines namely
N-Nitrosodimethylamine (NDMA) and N-Nitrosodiethylamine
(NDEA) in Sartan and Metformin APIs by procedure B. The
procedure B further describes 2 sample preparations
procedures namely sample preparation 1 and 2. The sample
preparation 1 is used for Valsartan, Losartan and Olmesartan
additionally Metformin API is also analysed with this method.
The sample preparation 2 is used for Candesartan and
Irbesartan. When a procedure is applied to substances outside
of the scope covered by the initial validation or to medicinal
products or if procedure A or B is used quantitatively, then it
must be validated.
Experimental
A mixture of NDMA, NEMA and NDEA standards was analyzed
using scan mode for identification. By referring to the
parameters mentioned in procedure B, a GC-MS quantitation
method was created (Table 1). EDQM method B uses a two
single point calibration methods each for sample preparation -1
and 2. However, in this application news two different
calibrations (three-point) each for sample preparation 1 and 2
were plotted.
Gas Chromatograph Mass Spectrometer GCMS-QP
TM
2020 NX, AOC-30i
Nitish Suryawanshi, Dheeraj Handique, Prashant Hase, Sanket Chiplunkar, Durvesh Sawant, Aseem
Wagle, Rahul Dwivedi, Jitendra Kelkar and Pratap Rasam
Shimadzu Analytical (India) Pvt. Ltd.
User Benefits
Figure 1: GCMS-QP2020 NX system with AOC-30i autosampler
Method
Table 1: Instrument configuration and analytical conditions
The calibration for sample preparation 1 is ranging from 3.75
ppb to 7.5 ppb whereas, for sample preparation 2 is ranging
from 1.5 ppb to 6.0 ppb. Quantitation of NDMA and NDEA was
performed using Shimadzu GCMS-QP2020 NX system with
AOC-30i autosampler (Figure 1).
Note = All above concentrations are as such.
GCMS System : GCMS-QP2020 NX with AOC-30i
Column
: SH-I-624Sil MS 30 m, 0.25 mm I.D., 1.4
µm df (P/N: 221-75962-30)
Injection Mode : Splitless (Sampling time 0.5 min)
Flow Control Mode : Column Flow
Injector Port Temp. : 250 ºC
Carrier Gas : Helium
Column Flow : 1.0 mL/min
Injection Volume : 2.0 µL*
Temp. Program
Ionization Mode : Electron Ionization (EI)
Interface Temp. : 240 ºC
Ion Source Temp. : 230 ºC
Acquisition Mode : SIM
Ions : 74, 88 and 102 amu
Ramp Rate
(ºC/min)
Temp.
(ºC)
Hold Time
(min)
- 40 0.5
58.8 140 2
20 180 0.5
30 240 1.8
40 280 2.5
*: EDQM mentions injection volume of 2.5µL