User Benefits Application News TOC Evaluation of Ammonia Solution The total organic carbon (TOC) in 2 % ammonia solution can be determined directly without dilution, which is optimal for organic contamination management of ammonia solution. By measuring inorganic carbon (IC), the contained carbon dioxide (CO 2 ) concentration can also be quantified. Automatic analysis of multiple samples is possible, with the use of the auto-sampler ASI-L. Jiajie Du Total Organic Carbon Analyzer TOC-L Ammonia solution is widely used for various purposes in modern society, such as use as a laundry detergent, in cosmetics and pharmaceuticals, as well as for the treatment of nitrogen oxides (NOx) generated in thermal power plants and waste treatment plants. Furthermore, ammonia fuel has gained attention in recent years as a zero-emission fuel for realization of a decarbonized society. Various types of research on ammonia fuel are underway due to its lack of CO 2 emissions. To use ammonia solution in these applications, there is a requirement for high purity, so control of organic impurities is important. Using the Shimadzu total organic carbon (TOC) analyzer enables rapid and simple control of organic impurities in ammonia solution. This article introduces an example of TOC measurement of 2 % ammonia solution, using the Shimadzu TOC-L analyzer. In this experiment, 25 % ammonia solution (special reagent grade) was diluted with pure water to prepare 2 % ammonia solution. Then, potassium hydrogen phthalate was added to the ammonia solution to achieve TOC concentrations of 1 mgC/L, 5 mgC/L, and 10 mgC/L, respectively. Table 1 summarizes the sample preparation. Analysis Method Table 2 presents the measurement conditions used for TOC analysis of ammonia solution using the total carbon (TC) and inorganic carbon (IC) subtraction method. This method involves measuring the TC and IC in the sample, and then subtracting IC from TC to obtain the TOC concentration. In addition, when it comes to TOC analysis, IC typically refers to dissolved CO 2 and carbonates (carbonic acid H 2 CO 3 , bicarbonate ion HCO 3 - , carbonate ion CO 3 2- ) in the solution. The results were calculated using the mean value of three tests, up to a maximum of five tests. If a measurement result exceeded the set limit of maximum standard deviation (SD) and maximum coefficient of variation (CV), additional tests were performed up to the maximum of five tests. Calibration of the instrument was performed using 0 and 20 mgC/L potassium hydrogen phthalate aqueous solutions for TC measurements, and 0 and 20 mgC/L sodium carbonate and sodium hydrogen carbonate aqueous solutions for IC measurements. The calibration curves were corrected by performing an origin shift to account for the TOC contained in the pure water used in preparing the standard solution. Introduction Fig. 1 Applications of Ammonia Solution Daily supplies Pharmaceuticals Industries Table 1 Sample Preparation Samples Ammonia solution [%] TOC [mgC/L] 2 % ammonia solution 2 0 2 % ammonia solution + TOC 1 mgC/L 1 2 % ammonia solution + TOC 5 mgC/L 5 2 % ammonia solution + TOC 10 mgC/L 10 Fig. 2 Shimadzu TOC-L Analyzer (Right) and Auto-Sampler ASI-L (Left) Instrument TOC-LCPH total organic carbon analyzer Catalyst Standard catalyst Measurement Item TOC (based on TC-IC) No. of Injections 3 / Max. 5 Calibration Curves TC: 2-point calibration curve with 0 and 20 mgC/L of potassium hydrogen phthalate aqueous solution IC: 2-point calibration curve with 0 and 20 mgC/L of sodium carbonate and sodium hydrogen carbonate aqueous solution Injection Volume 50 μL Samples Special reagent grade, 25 % ammonia solution Potassium hydrogen phthalate Table 2 Analytical Conditions
Application News www.shimadzu.com/an/ Shimadzu Corporation © Shimadzu Corporation, 2023 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. Fig. 5 Correlation between the TOC Added Concentration and the Measured Concentration First Edition: Mar. 2023 01-00502-EN Measurement Results Conclusion The results of ammonia solution measurement are shown in Table 3. The TC and IC concentrations were obtained from the mean value of up to 5 tests, and the TOC results were calculated by subtracting IC from TC. Examples of detailed peak data from the TOC-L analyzer are presented in Figs. 3 and 4. The correlation between the TOC added concentration and the measured concentration is shown in Fig. 5, where the correlation coefficient was 0.9999, indicating good correlation and confirming the accurate measurement of added TOC. From these results, it is clear that the TOC concentration of 2 % ammonia solution can be measured up to 1 mgC/L, by using the Shimadzu TOC-L analyzer. Samples TC [mgC/L] IC [mgC/L] TOC [mgC/L] 2 % ammonia solution 0.950 0.888 0.0624 2 % ammonia solution + TOC 1 mgC/L 1.87 0.754 1.11 2 % ammonia solution + TOC 5 mgC/L 6.52 1.45 5.07 2 % ammonia solution + TOC 10 mgC/L 10.3 0.481 9.83 Table 3 Measurement Results of 2 % Ammonia Solution Fig. 4 Results of 2 % Ammonia Solution + TOC 5 mgC/L (a) TC Measurement (b) IC Measurement (a) (b) This article confirmed that the Shimadzu TOC-L analyzer is applicable in TOC analysis of 2 % ammonia solution with a precision of up to 1 mgC/L. In the case of higher concentration ammonia solution, appropriate dilution prior to analysis can be utilized for TOC measurement. Moreover, the amount of CO 2 in ammonia solution can be quantified through IC measurement. Ammonia has attracted attention for realizing a decarbonized society, and it is expected that the TOC-L analyzer will play a vital role in managing organic impurities in ammonia solution. (a) Fig. 3 Results of 2 % Ammonia Solution + TOC 1 mgC/L (a) TC Measurement (b) IC Measurement (b)
Please fill out the survey Related Products Some products may be updated to newer models. TOC-L Series Total Organic Carbon Analyzer Related Solutions Carbon Neutrality Manufacturing, QA/QC Hydrogen, Ammonia Clean Energy Fuel Cells Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)