Application
News
www.shimadzu.com/an/
Shimadzu Corporation
© Shimadzu Corporation, 2023
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
See http://www.shimadzu.com/about/trademarks/index.html for details.
Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not
they are used with trademark symbol “TM” or “”.
Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
to change without notice.
First Edition: May 2023 01-00505-EN
Fig. 2 Calibration Curves
Calibration Curves
Fig. 2 shows the calibration curves of the three target
compounds for concentrations from 0.1 to 10 mg/L each.
Excellent linearities were obtained for all compounds, with
coefficients of determination (r
2
) of 0.999 or greater. In addition,
high sensitivity analysis was achieved by using a phosphate
buffer as a mobile phase with low UV absorption. Table 3 shows
the limit of quantitation calculated from the lowest calibration
curve concentrations for the three target compounds
(calculation method: ASTM).
Conc.
Conc.
Table 3 Limit of Quantitation
Fig. 3 Chromatograms of Mouthwash
(Solid line: Sample, Broken line: Standard solution)
Table 2 Repeatability (%RSD) in Six Repeated Analyses
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 min
0.0
2.5
5.0
7.5
10.0
12.5
mAU
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 min
0.0
2.5
5.0
7.5
10.0
12.5
15.0
mAU
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 min
0
1
2
3
4
5
6
7
mAU
1
3
2
Peak
3. IPMP
Peak
2. CPC
Peak
1. GK2
250 nm
258 nm
280 nm
Compound Concentration (mg/L) %RSD
GK2 1.65 0.27
CPC 4.25 0.24
IPMP 2.92 0.15
(mg/L)
CPC
GK2
0.0 5.0
0.0
0.5
1.0
1.5
2.0
2.5
3.0
(×10,000)
0.0 5.0
0.0
1.0
2.0
3.0
4.0
(×10,000)
0.0 5.0
0.0
1.0
2.0
3.0
4.0
(×10,000)
(mg/L)
(mg/L)
IPMP
Compound Retention time Peak area
GK2 0.11 1.01
CPC 0.12 0.48
IPMP 0.12 0.64
Compound LOQ (mg/L)
GK2 0.079
CPC 0.076
IPMP 0.053
Peak area Peak area
Peak area
Conc.
Conclusion
A high-speed and high sensitivity method for analyzing CPC,
GK2, and IPMP, which are active ingredients in mouthwash, was
developed. In combination with a Shim-pack Arata C18 column
and a phosphate buffer as the mobile phase, tailing of the peaks
of basic compounds was suppressed, enabling the
quantification of trace compounds in a short period of time.
Analysis of Mouthwash
A commercial mouthwash was used as the sample. Analytes
were prepared by diluting the mouthwash 100-fold with 50 %
methanol solution and filtration with a 0.2 µm membrane filter.
Fig. 3 shows the chromatograms of the mouthwash, and Table 4
shows the concentrations of the three compounds in the
mouthwash. Note that these concentrations shown in Fig. 4 are
average concentrations obtained from six repeated analyses of
samples after pretreatment.
Table 4 Analysis Results (n=6)
Nexera, Shim-pack Arata and SHIMADZU LabTotal are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries.
References
1)U. S. Pharmacopeia 41-NF 36, 2017, Official Monographs
“Cetylpyridinium Chloride”
Related Applications
1. High-Speed and Simultaneous Analysis of Active Ingredients CPC
and GK2 in Mouthwash, Application News No. 01-00116A