User Benefits
Application
News
Analysis of Elemental Impurities in Oral Drug
Products Using ICPMS-2040/2050 ―ICH Q3D―
Robust quantitative analysis of control thresholds even when Option 1 (ICH Q3D) is used to convert PDEs to concentration limits
Easily meets acceptance criteria for accuracy and precision for the analytical procedure used to quantify elemental impurities
Preset methods eliminate the labor needed to establish analytical conditions and allow anyone to perform analysis with ease
Kana Matsuno and Kisho Hori
Inductively Coupled Plasma Mass Spectrometer ICPMS-2040/2050
Sample Preparation
Test samples were prepared for analysis by the procedure
shown in Fig. 2.
Approximately. 0.2 g of the test sample was added 4 mL of pure
water, 4 mL of nitric acid, and 0.5 mL of hydrochloric acid, and
then digested in a microwave digestion system (200 °C for
approximately 60 minutes). Hydrochloric acid was added to
improve the stability of Hg and other elements in the solution.
The digestion vessel was cooled to room temperature to avoid
vaporization of elements with a low boiling point, and then the
sample solution was made up to 50 mL for analysis (250-fold
dilution).
A preparation blank solution containing no test sample was also
prepared by the same procedure to verify the amounts of each
element introduced as contaminants during the digestion
process.
Target Limits and Test Solution Concentrations
When evaluating the significance of elemental impurity levels,
target limits can be set to 30 % of the PDE (control threshold)
divided by the maximum daily dose (Equation 1). PDEs were
converted into concentration limits using Option 1, which
assumes a maximum total daily dose of 10 g of the drug
products.
Target Limit [µg/g] =
PDE [µg/day] × 30 [%]
10 [g/day]
...[1]
Target limits were converted to target concentrations in the
sample solution using Equation 2.
Target concentration [µg/L]
=
Target Limit [µg/g]×Sample Digestion Amount [g]
Volume [L]
…[2]
Standard Solution
Standard Solutions
Standard solutions containing Class 1 and Class 2A elements
at 50 %, 100 %, and 150 % of the target concentrations were
prepared along with a blank solution. The same volumes of
nitric acid and hydrochloric acid as the sample solutions were
added to the standard solutions.
Internal Standard Solution
Commercially available single-element standard solutions
were diluted and mixed to prepare an internal standard
solution containing Sc at 2000 µg/L, Ga at 1000 µg/L, and In
and Bi at 100 µg/L. The same volumes of nitric acid and
hydrochloric acid as the internal standard solutions were
added to the sample solutions. To reduce the labor involved
in sample preparations, an online internal standard kit was
used to automatically add the internal standard solution to a
sample at the internal standard ratio of approximately 9:1.
Fig. 2 Sample Preparation Procedure
(1) Place the acid and
test sample in the
digestion vessel
(2) Digest in a microwave
digestion system
(3) Make up to a
fixed volume
Cool
Introduction
ICH Q3D (R2) Guideline for Elemental Impurities
1)
establishes
permitted daily exposure (PDE) levels for 24 elements of
toxicological concern. The guideline requires that levels of these
elements be controlled by appropriate analytical methods.
In response to this guideline, section “2.66 Elemental Impurities”
2)
was included in the Japanese Pharmacopoeia (JP), sections
“<232> Elemental Impurities―Limits” and “<233> Elemental
Impurities―Procedures”
3)
were included in the United States
Pharmacopeia (USP), sections “5.20 Elemental Impurities” and
“2.4.20 Determination of elemental impurities”
4)
were included
in the European Pharmacopoeia (EP), and inductively coupled
plasma mass spectrometry (ICP-MS) was added as an analytical
method.
This Application News describes using the ICPMS-2050 (Fig. 1)
to analyze commercially available oral drug products (a
gastrointestinal drug). Assuming that elemental impurities are
measured at concentrations lower than 30 % of the PDE (control
threshold), accuracy, precision, and quantitative limits were
confirmed in an example analysis performed according to the
quantitative procedures described in the JP, USP, and EP.
Sample
Test Sample: Oral drug products (gastrointestinal drug,
orally disintegrating tablet)
Standard Solutions: Multi-element mixed standard solution for
ICH Q3D oral agents (Cd, Pb, As, Co, V, and Ni)
Mercury standard solution for ICH Q3D
Single-element standard solutions of Sc, Ga,
In, and Bi
Fig. 1 ICPMS-2050 and AS-20
Target Elements
The elements targeted in a risk assessment differ depending on
the route of administration of the drug products. Quantities of
Class 1 and 2A elements were determined as required for a risk
assessment of oral drug products.