■ Results and Discussion
Six culture media from three subtypes (T cell media, IVF media, and
DMEM) were analyzed using the MS scan method. Three embryo cul-
ture media from the same vendor, two serum-free and xeno-free T
cell media from two different vendors, and one DMEM medium were
investigated. Cell culture media PCA showed clear delineation be-
tween the different media subtypes (Fig. 2). Even though the two T
cell culture media were sourced from two different vendors, they
are still closely grouped on the PCA score plot and can be distin-
guished from the other media subtypes.
The two T cell culture media were also analyzed using the cell culture
media profiling MRM method targeting one hundred and twentyfive
cell culture related compounds. Sugar (glucose), nucleic acid metab-
olite (uridine monophosphate), amino acids and metabolites
(cystine, asparagine, aspartic acid, serine, glycine, glutamine,
threonine, glutamic acid, alanine, citrulline, ornithine, proline, lysine,
2-aminobutyric acid, histidine, argininosuccinic acid,
arginine, 5-oxoproline, valine, methionine, pipecolic acid, tyrosine,
isoleucine, leucine, and phenylalanine), and vitamins (niacinamide,
pantothenic acid, folic acid, and riboflavin) were detected in both T
cell media (Fig. 3). In addition to the PCA plot, similarities of the
two T cell culture media were further revealed by the abundance of
the identical components observed in both media.
Significant differences between the two T cell culture media were
also identified (Fig. 4). The discrepancies are especially apparent in
glutamine (m/z 145.20 > 127.15; t
R
: 1.2 min), alanyl-glutamine (m/z
216.00 > 154.05; t
R
: 1.5 min), pipecolic acid (m/z 130.10 > 84.05; t
R
:
2.3 min), and biotin (m/z 245.10 > 226.95; t
R
: 5.1 min).
Dispense Mix Centrifuge Inject supernatant
*1 mM 2-Isopropylmalic acid aq.
Sample media: 50 μL
Water: 130 μL
ISTD solution*: 20 μL
Acetonitrile : 400 μL
Fig. 1 Sample preparation procedure for cell culture profiling analysis.
LC Conditions [LC-40 series]
Mobile Phase A : Water containing 0.1% formic acid
Mobile Phase B : Acetonitrile containing 0.1% formic acid
Chromatography Gradient : 0% B -> 95% B
Total Run Time
Column
: 17 minutes
: Cell Culture Profiling Column (150mm x 2.1mm, 3 μm)
(P/N 220-91581-10)
Flow Rate : 0.35 mL/min
Column Oven Temperature : 40
Injection Volume : 10 μL
MS Conditions [LCMS-8050]
Ionization Mode : ESI +/-
Acquisition Mode : MRM and Scan
Scan Mass Range : m/z 200-1000
Nebulizing Gas Flow : 3.0 L/min
Heating Gas Flow : 10.0 L/min
Interface Temperature : 300
º
C
DL Temperature : 250
º
C
Heat Block Temperature : 400
º
C
Drying Gas Flow : 10.0 L/min
Table 2 Analytical conditions