Application
News
www.shimadzu.com/an/
Shimadzu Corporation
© Shimadzu Corporation, 2023
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
See http://www.shimadzu.com/about/trademarks/index.html for details.
Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not
they are used with trademark symbol “TM” or “”.
Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
to change without notice.
Before reaction
20 minutes after
reaction started
1.052
0.996
477
Lightway and Nexis are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries.
Conclusion
This article describes using a Lightway photoreaction evaluation
system and Nexis GC-2030 gas chromatograph to evaluate the
efficiency of a visible light-responsive photocatalytic hydrogen
generation system, which has attracted attention within the
artificial photosynthesis research field for achieving carbon
neutrality. The results confirmed excellent quantum yield
performance compared with preceding studies. The method can
also be used to estimate the decrease in quantity of
photosensitizer based on the change in the absorption spectra
before and after reactions.
Confirmation of Decrease in Photosensitizer
Quantity
The decrease in the iridium complex photosensitizer quantity
was confirmed based on the change in the absorption spectrum
before and after the reaction. Absorption spectra obtained
using the Lightway system before and after the reaction are
shown in Fig. 6.
The absorbance at 477 nm, which is the peak from the iridium
complex, was 1.052 before the reaction and 0.996 twenty
minutes after the reaction. This indicates that absorbance
decreased about 5 % during the reaction. Assuming that the
decrease in absorbance is proportional to the decrease in
iridium complex quantity, it can be inferred that the iridium
complex quantity also decreased about 5 %.
Φ
1
= Number of Hydrogen Generated
/ Number of Photons Absorbed … (1)
Φ
2
= Number of Electrons Used in Reaction
/ Number of Photons Absorbed …(2)
References
1) S. Takizawa et. al., Inorg. Chem. 2016, 55, 8723−8735.
2) S. De. Kreijger et. al., Inorg. Chem. 2022, 61, 5245−5254
Fig. 5 Relationship between Number of Absorbed Photons and Number of
Generated Hydrogen Molecules.
0.0 0.5 1.0 1.5 2.0
min
0.00 0.25 0.50 0.75
Concentration
0
100000
200000
300000
Area
1.00
400000
Fig. 4 Calibration Curve of Hydrogen (0.1 %, 0.5 %, 1 %)
R
2
= 0.9996
Calculation of Photoreaction Quantum Yield
Accurate quantum yield can be calculated from the number of
absorbed photons measured by Lightway and the number of
generated hydrogen molecules measured by Nexis GC-2030.
The relationship between the number of absorbed photons and
the number of generated hydrogen molecules is shown in Fig. 5.
This confirms excellent linearity between the number of absorbed
photons and the number of generated hydrogen molecules.
In this case, the following two formulas were used to calculate
the quantum yield.
The hydrogen concentration was determined by the calibration
curve in Fig. 4. The volume of hydrogen generated was calculated
based on the hydrogen concentration and the cell headspace
volume. In addition, the hydrogen volume was converted to
hydrogen mass assuming one mole of hydrogen per 22.4 L.
Hydrogen Quantitation by Gas Chromatograph
TCD was used as the Nexis GC-2030 detector and argon as the
carrier gas. TCD detects target compounds based on their
difference in thermal conductivity from the carrier gas.
Therefore, argon is commonly used as a carrier gas for hydrogen
analysis due to its large difference in thermal conductivity
compared with hydrogen. An example of a chromatogram
obtained from this measurement is shown in Fig. 3. The peak of
hydrogen was detected at a retention time of 0.5 minutes.
The calibration curve obtained by the analysis of the standard
hydrogen gas samples is shown in Fig. 4.
Fig. 6 Absorption Spectra of Iridium Complex Before and After Photocatalytic
Hydrogen Generation Experiment
Fig. 3 Chromatogram of Sample Analysis Obtained with Gas Chromatograph
Hydrogen
(Quantitative
concentration: 0.61 %)
First Edition: Aug. 2023
Acknowledgments
We would like to express our sincere gratitude to Dr. Shinya Takizawa,
Assistant Professor, Department of Basic Science, Graduate School of Arts and
Sciences, The University of Tokyo, for providing the iridium and cobalt
complexes used for measurements.
Equation (1) has a value equivalent to the slope in Fig. 5, which
results in a calculation of Φ
1
= 1.44 %. Assuming one photon is
transferred per electron, two electrons are required to generate
a hydrogen, resulting in a maximum quantum yield value of
50 % based on the definition in equation (1).
With equation (2), given that two electrons are used per
molecule of hydrogen, the slope in Fig. 5 is doubled to result in
a calculation of Φ
2
= 2.88 %. That method results in a maximum
quantum yield of 100 % for all photoreactions premised on one
photon transfer per electron. When comparing quantum yield
with previous researches, it is necessary to confirm which
definition was used as the basis for calculating the yield.
In reference 2), the quantum yield of photocatalytically generated
hydrogen was measured based on an iridium complex with a
ligand bond other than the coumarin ligand. In that case, the
calculation is reported as Φ
2
= 0.13 % (446 nm), based on
equation (2). A comparison showed that the method in this article
produced far superior quantum yield, with about 22 times higher
quantum yield than the method in reference 2).