User Benefits Application News Kneading machine treatment Heating conditions Sample A Without kneading Kneaded in molding machine, 260 ˚C, 250 s (approx.) Sample B With kneading Kneading machine treatment, 260 ˚C, 120 s + Kneaded in molding machine, 260 ˚C, 250 s (approx.) Instruments : IRTracer-100, QATR ™ 10 (diamond prism) Wavelength range : 400 - 4000 cm -1 Resolution : 4 cm -1 Accumulation : 40 times Apodization function : Happ-Genzel Detector : DLATGS Sample A Sample B 500 1000 1500 2000 3000 4000 cm-1 Abs With kneading_1 With kneading_2 With kneading_3 Without kneading_1 Without kneading_2 Without kneading_3 Multifaceted Evaluation of Plastics: Differences due to Kneading Conditions Multifaceted evaluation by instrumental analysis can provide a more detailed understanding of changes in properties due to differences in molding conditions such resin material kneading conditions, and the data obtained are effective for studying the optimum molding conditions. Use of the MALDI-TOF MS makes it possible to analyze changes in the molecular structure of resin materials during molding. Kazumi Kawahara, Fumiaki Yano, Tsukasa Nishimura, Yuzo Yamazaki, and Mitsuru Ohta IRTracer™-100 Fourier Transform Infrared Spectrophotometer, AGX™-V Autograph Precision Universal Testing Machine, DUH™-210 Dynamic Ultra Micro Hardness Tester, DSC-60 Plus Differential Scanning Calorimeter, UV-2600i UV-VIS Spectrophotometer, MALDI-8020 Matrix Assisted Laser Desorption/Ionization Time of Flight Mass Spectrometer Fig. 1 Photograph of Samples Differences in the compositional homogeneity of the blended resins in Sample A and Sample B were confirmed by using a Fourier transform infrared spectrophotometer (FTIR). The single reflection ATR method was used in the measurements, which were performed at three locations (center and near the two ends) of the dumbbell-shaped specimens (see Fig. 1). Confirmation of Compositional Homogeneity Fig. 5 DSC Curves It is necessary to study the optimum conditions for injection molding of resin because various molding conditions, such as temperature, pressure, and time, affect the properties of molded products. In particular, when using blended resins, changes in property due to differences in kneading during molding are a concern. In this experiment, multifaceted measurements of the properties of samples prepared under different kneading conditions (differences with/without kneading treatment by a kneading machine) were carried out using a polycarbonate (PC)/acrylonitrile-butadiene-styrene (ABS) blended resin to confirm the effects of kneading conditions. Introduction Here, the material properties of two types of specimens were compared, Sample A with PC : ABS = 50 : 50, which was prepared by injection molding with kneading only in the molding machine without kneading treatment by the kneading machine, and Sample B with the same PC : ABS compounding ratio, which was injection molded after kneading treatment with the kneading machine. Table 1 shows the molding conditions, and Fig. 1 shows a photograph of the samples. Test Specimens Table 1 Molding Temperature and Kneading Time Table 2 Measurement Conditions Fig. 2 IRTracer™-100 Fig. 3 Infrared Spectra From Fig. 3, the spectral shapes at the total of six locations on the two samples were in good agreement. Based on these infrared spectra, there are considered to be no compositional differences between the samples due to differences in the kneading conditions of this experiment. Measurements were carried out by differential scanning calorimeter (DSC). Fig. 5 shows the DSC curves. The glass transitions of the ABS and PC resins were measured at around 110 ˚C and 140 ˚ C, respectively. The glass transition temperature is an index of heat resistance. No significant differences were seen between Sample A and Sample B. Thermal Analysis Fig. 4 DSC-60 Plus Table 3 Measurement Conditions Instrument : DSC-60Plus Heating rate : 20 ˚C/min Specimen weight : 8 mg Atmosphere : Nitrogen 106.78 ˚C 139.14 ˚C 106.34 ˚C 138.99 ˚C Sample B With kneading Sample A Without kneading Temp 100.00 150.00 [˚C] -2.00 -1.00 0.00 mW DSC
Application News Tensile Test The following shows the results of a tensile test conducted with a precision universal testing machine. Yellowness Index Measurement Comparing the appearances of the two molded specimens, the colors of Sample A and Sample B were different. Therefore, yellowness index was evaluated with an ultraviolet-visible spectrophotometer (UV). Yellowness index was obtained by calculation from reflectance, and was higher in Sample B, which received kneading treatment by the kneading machine. Fig. 6 UV-2600i Table 4 Measurement Conditions Instruments : UV-2600i, ISR-2600Plus Measured wavelength range : 380 - 780 nm Data interval : 1.0 nm Scan speed : Medium Slit width : 5.0 nm Fig. 7 Reflection Spectra of Specimens Table 5 Results of Yellowness Index Measurements Kneading machine treatment Yellowness Sample A Without kneading 3.44 Sample B With kneading 12.99 Wavelength (nm) Reflectance (%) ー Without kneading by kneading machine ー With kneading by kneading machine Table 6 Instrument Configuration and Test Conditions Precision universal testing machine : AGX-V Load cell : 5 kN Grip : Pneumatic flat grip Extensometer : TRViewX240S Software : TRAPEZIUM™X-V Test speed : 1 mm/min 50 mm/min (switched at 0.3 % displacement) Gauge length : 75 mm Number of tests : n = 5 Specimen width : 10 mm Specimen thickness : 4 mm Grip distance : 115 mm Fig. 8 AGX™-V Fig. 9 Condition of Measurement ー Without kneading by kneading machine ー With kneading by kneading machine Fig. 10 Test Curves Table 7 Test Results Kneading machine treatment Tensile strength (MPa) Elastic modulus (MPa) Breaking elongation (%) Sample A Without kneading 54.23 2459.13 120.92 Sample B With kneading 53.90 2492.45 71.75 Stress (N/mm 2 ) Displacement (strain) (%) No large differences were found in the tensile strength and elastic modulus of the two specimens. However, the breaking elongation of Sample B was smaller than that of Sample A. The hardness test was conducted using a dynamic ultra micro hardness tester (DUH). Hardness Test Fig. 11 DUH™-210 Table 9 Test Results Fig. 12 Test Curves Force-indentation depth graph Indentation depth [μm] Force [mN] With kneading Without kneading Kneading machine treatment H IT (MPa) Sample A Without kneading 166.46 Sample B With kneading 160.36 Table 8 Measurement Conditions (Conforming to ISO/TS 19278) Instrument : DUH-210 Indenter : Berkovich indenter Test mode : Load-unload test Force : 500 mN Loading time/ unloading time : 30 s Load holding time : 40 s Number of tests : 5 (central 3 tests were extracted) Ambient temperature : 23±2 ˚C Humidity : 50±10 % No significant difference in the indentation hardness (H IT ) of the two specimens was seen.
Application News www.shimadzu.com/an/ Shimadzu Corporation © Shimadzu Corporation, 2023 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. First Edition: Oct. 2023 01-00590-EN Measurement of Changes in Molecular Structure Conclusion AGX, TRAPEZIUM, DUH, IRTracer, and QATR are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries. In the results in the previous sections, differences in yellowness and breaking elongation were seen in Sample A and Sample B. These differences were assumed to occur due to changes in the chemical structure of the resin caused by the difference in the kneading condition. To confirm this assumption, the differences in the molecular structures of the two specimens were measured with a matrix assisted laser desorption/ionization time of flight mass spectrometer (MALDI-TOF MS). The results are shown in Fig. 14. In this evaluation, there were no differences in the compositional homogeneity, tensile strength, elastic modulus, hardness, or glass transition temperature of the specimens prepared under the different kneading conditions. Thus, no effect of kneading on these properties was observed. On the other hand, when kneading treatment was conducted, deterioration of the yellowness index and breaking elongation properties could be seen, and progress of changes in the molecular structure due to kneading treatment was also confirmed. Multifaceted evaluation like that introduced here can provide a detailed understanding of the changes in material properties that occur as a result of differences in overall molding conditions, not limited simply to kneading. Therefore, this is considered to be an effective approach for study of the optimum molding conditions. Fig. 13 MALDI-8020 Fig. 14 Comparison by MALDI-TOF MS A) Without kneading, B) With kneading, C) PC : New signal due to kneading : Characteristic signal due to blending (molding) Cyclic; (n) Both-end capped; –(n)– One-end hydroxyl group; HO(n)– Fig. 15 End Terminal Structures of Polycarbonate (PC) Table 10 Measurement Conditions Instrument : MALDI-8020 Laser source : Solid-state laser (λ = 355 nm) Matrix : DCTB, Nal (cationization agent) Mass range : m/z 1-5000 Detection ion : Positive ion mode Flight mode : Linear mode Delayed extraction condition : m/z 2500 Changes in the signal originating from PC in the blended product were observed, regardless of whether the material was treated with the kneading machine or not. In comparison with virgin PC, there was a relative decrease in both-end capped molecules in both Samples A and B, and the number of molecules having one end capped with a hydroxyl group increased. Furthermore, the number having a hydroxyl group at one end decreased in the specimen treated with the kneading machine. Although the chemical structure could not be estimated, new characteristic signals appeared, as indicated by the yellow arrows in Fig. 14. The possibility that the origin of these signals may be some type of reaction with the one-end hydroxyl groups due to the extended kneading time is conceivable. A precise analysis extending to the cause of the changes in yellowness and breaking elongation was not possible. However, it was at least possible to confirm that changes occur in the end groups of PC as a result of a chemical reaction during kneading. Thus, in addition to analytical methods to understand the physical properties of the material, such as strength, analysis of chemical structure at the molecular level is considered to be effective for ensuring proper molding.
Please fill out the survey Related Products Some products may be updated to newer models. AGX-V Series AUTOGRAPH Precision Universal Tester IRTracer-100 Fourier Transform Infrared Spectrophotometer DSC-60 Plus Series Differential Scanning Calorimeter UV-2600i, UV-2700i UV-VIS-NIR Spectrophotometer MALDI-8020 Matrix-Assisted Laser Desorption/Ionization Time-o… Related Solutions Automotive Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)