No. SSI -LCMS-150 This application news summarizes the performance of the Shimadzu LCMS-8060NX Liquid Chromatography Mass Spec- trometer (LC/MS/MS) (Fig. 1) for all analytes listed in ASTM D8421. Results meet or exceed the requirements outlined in the method. The reporting range and the target analytes are listed in Ta- ble 1. The reporting limit (RL) for the test method is defined as an integer value that is equal to the concentration of the lowest calibration standard. Fig. 1 Shimadzu LCMS-8060NX Introduction and Background ASTM International published ASTM D8421 1 for the analysis of 44 per- and polyfluorinated alkyl substances and 24 la- beled isotopes in non-potable water samples. This method extracts the substances in a 1+1 ratio of sample and methanol, filters and then measures the targeted com- pounds using external standard calibration liquid chro- matography-tandem mass spectrometry (LC/MS/MS). The minimum reporting limit is 10 ng/L with an analytical range of 10 - 400 ng/L for most compounds. The method requires standard solutions be prepared by the laboratory from neat compounds. To save a laboratory’s time and effort from preparing stock standards individually, we optimized the method using com- mercially available native and labeled calibration standard mixes. Additionally, we optimized chromatography, achiev- ing better peak shape for early-eluting compounds, such as PFBA and PFPrA. Table 1 Analyte List with D8421 Reporting Range CAS Number Acronym Analyte Name Range (ng/L) Perfluorotetradecanoic acid PFTreA 376-06-7 10-400 Perfluorotridecanoic acid PFTriA 72629-94-8 10-400 Perfluorododecanoic acid PFDoA 307-55-1 10-400 Perfluoroundecanoic acid PFUnA 2058-94-8 10-400 Perfluorodecanoic acid PFDA 335-76-2 10-400 Perfluorononanoic acid PFNA 375-95-1 10-400 Perfluorooctanoic acid PFOA 335-67-1 10-400 Perfluoroheptanoic acid PFHpA 375-85-9 10-400 Perfluorohexanoic acid PFHxA 307-24-4 10-400 Perfluoropentanoic acid PFPeA 2706-90-3 50-1000 Perfluorobutanoic acid PFBA 375-22-4 50-1000 Perfluorodecanesulfonic acid PFDS 335-77-3 10-400 User Benefits ◆ The LCMS-8060NX easily meets and exceeds method performance criteria of ASTM D8421 for 44 PFAS and 24 surrogates. ◆ Optimized chromatography and MS conditions for excellent peak shape for improved precision and accuracy. ◆ ASTM D8421 is a simple extraction procedure validated by ASTM for the analysis of PFAS in wastewater samples. Liquid Chromatography Mass Spectrometry ASTM D8421-22 Standard Test Method for Determination of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous Matrices by Co-solvation Followed by Analysis Using the Shimadzu LCMS-8060NX Nami Iwasa 1 , Landon Wiest 2 , William Lipps 2 1 Shimadzu Corporation, 2 Shimadzu Scientific Instruments, Inc.
Perfluorononanes`ulfonic acid PFNS 68259-12-1 10-400 Perfluorooctanesulfonic acid PFOS 1763-23-1 10-400 Perfluoroheptanesulfonic acid PFHpS 375-92-8 10-400 Perfluorohexanesulfonic acid PFHxS 355-46-4 10-400 Perfluoropentanesulfonic acid PFPeS 2706-91-4 10-400 Perfluorobutanesulfonic acid PFBS 375-73-5 10-400 Perfluorooctanesulfonamide PFOSA 754-91-6 10-400 8:2 Fluorotelomer sulfonic acid 8:2 FTS 39108-34-4 10-400 6:2 Fluorotelomer sulfonic acid 6:2 FTS 27619-97-2 10-400 4:2 Fluorotelomer sulfonic acid 4:2 FTS 757124-72-4 10-400 N-Ethylperfluorooctanesulfonamidoacetic acid NEtFOSAA 2991-50-6 10-400 N-Methylperfluorooctanesulfonamidoacetic acid NMeFOSAA 2355-31-9 10-400 Perfluorododecanesulfonic acid PFDoS 79780-39-5 10-400 N-Methylperfluorooctanesulfonamide NMeFOSA 31506-32-8 10-400 N-Ethylperfluorooctanesulfonamide NEtFOSA 4151-50-2 10-400 N-Methylperfluorooctanesulfonamidoethanol NMeFOSE 24448-09-7 10-400 N-Ethylperfluorooctanesulfonamidoethanol NEtFOSE 1691-99-2 10-400 Hexafluoropropylene oxide dimer acid HFPO-DA 13252-13-6 10-400 4,8-dioxa-3H-perfluorononanoic acid ADONA 919005-14-4 10-400 9-chlorohexadecafluoro-3-oxanonane-1-sulfonic acid 9Cl-PF3ONS 756426-58-1 10-400 11-chloroeicosafluoro-3-oxaundecane-1-sulfonic acid 11Cl-PF3OUdS 763051-92-9 10-400 Pentafluorpropanoic acid PFPrA 422-64-0 50-1000 Perfluoro-3,6-dioxaheptanoic acid NFDHA 151772-58-6 10-400 Perfluoro(2-ethoxyethane) sulfonic acid PFEESA 113507-82-7 10-400 Perfluoro-3-methoxypropanoic acid PFMPA 377-73-1 10-400 Perfluoro-4-methoxybutanoic acid PFMBA 863090-89-5 10-400 2H,2H,3H,3H-Perfluorohexanoic Acid 3:3 FTCA 356-02-05 10-400 2H,2H,3H,3H-Perfluorooctanoic Acid 5:3 FTCA 914637-49-3 10-400 2H,2H,3H,3H-Perfluorodecanoic acid 7:3 FTCA 812-70-4 10-400 2H-perfluoro-2-octenoic acid FHUEA 70887-88-6 10-400 2H-perfluoro-2-decenoic acid FOUEA 70887-84-2 10-400 Lithium Bis(trifluoromethane)sulfonimide * HQ-115 90076-65-6 10-400 Surrogates Perfluoro-n-[ 13 C4] butanoic acid MPFBA NA 10-400 Perfluor0-n-[ 13 C5] pentanoic acid M5PFPeA NA 10-400 Perfluoro-n-[1,2,3,4,6- 13 C5] hexanoic acid M5PFHxA NA 10-400 Perfluoro-n-[1,2,3,4- 13 C4] heptanoic acid M4PFHpA NA 10-400 Perfluoro-n-[ 13 C8] octanoic acid M8PFOA NA 10-400 Perfluoro-n-[ 13 C9] nonanoic acid M9PFNA NA 10-400 Perfluoro-n-[1,2,3,4,5,6- 13 C6] decanoic acid M6PFDA NA 10-400 Perfluoro-n-[1,2,3,4,5,6,7- 13 C7] undecanoic acid M7PFUnA NA 10-400 Perfluoro-n-[1,2- 13 C2] dodecanoic acid MPFDoA NA 10-400 Perfluoro-n-[1,2- 13 C2] tetradecanoic acid M2PFTreA NA 10-400 Perfluoro-1-[ 13 C8] octanesulfonamide M8FOSA NA 10-400 N-methyl-d3-perfluoro-1-octanesulfonamidoacetic acid D3-N-MeFOSAA NA 10-400 N-ethyl-d5-perfluoro-1-octanesulfonamidoacetic acid D5-N-EtFOSAA NA 10-400 N-methyl-d3-perfluoro-1-octanesulfanamide d-N-MeFOSA NA 10-400 N-ethyl-d5-perfluoro-1-octanesulfanamide d-N-EtFOSA NA 10-400 2-(N-methyl-d3-perfluoro-1-octanesulfonamido) ethan-d4-ol d7-N-MeFOSE NA 10-400 2-(N-ethyl-d5-perfluoro-1-octanesulfonamido) ethan-d4-ol D9-N-EtFOSE NA 10-400 2,3,3,3-Tetrafluoro-2-(1,1,2,2,3,3,3-heptafluoropropoxy- 13 C3-propanoic acid MHFPO-DA NA 10-400 1H,1H,2H,2H-perfluoro-1-[1,2- 13 C2] hexane sulfonate M4:2FTS NA 10-400 1H,1H,2H,2H-perfluoro-1-[1,2- 13 C2]-octane sulfonate M6:2FTS NA 10-400 1H,1H,2H,2H-perfluoro-1-[1,2- 13 C2]-decane sulfonate M8:2FTS NA 10-400 Perfluoro-1-[ 13 C8] octanesulfonate M8PFOS NA 10-400 Perfluoro-1-[2,3,4- 13 C3] butanesulfonate MPFBS NA 10-400 Perfluoro-1-[1,2,3- 13 C3] hexanesulfonate M3PFHxS NA 10-400 Compounds in red are not in the Method 1633 standards and need to be added separately
The individual standard solution was prepared in 50:50 (vol: vol) methanol/water with 0.1% acetic acid to obtain final concentrations shown in Table 2. Materials and Methods Stock standard solutions containing native analytes and labeled isotopes (surrogates) were diluted from commercially available mixed stock standards (Wellington Method 1633 standard mixes) to be within the calibration range per analyte as shown in Table 1. Table 2 Concentrations of each Calibration Standard (CS) in ng/L These standards were not filtered. Calibration is performed using a 6 to 10-point curve, depending on the analyte. To obtain the calibration levels from the commercial stock solutions, 15 individual calibration standards were prepared and analyzed. Only the calibration points within the method- specified range were used. The stock solutions were prepared and stored in PFAS-free polypropylene (PP) containers. Prior to the analysis, the solutions were shaken thoroughly, then transferred to a 2 mL PP LC vial and analyzed within 24 hours. If samples or standards are allowed to sit in the LC vials, some PFAS compounds may settle, rise, precipitate, or adsorb on the surface. To ensure a homogenous solution and optimum results, the solutions were vortexed prior to injection. 2.1 Sample Preparation The surrogate spiking mix is added to 5 mL of sample contained in a 15 mL polypropylene vial. Add 5 mL of methanol and mix by vortex for ~2 minutes. After mixing, add acetic acid and adjust the pH ~4. Transfer an aliquot to a LC vial and cap with a Shimadzu GLC PP vial with septum confirmed to not contain PFAS. Analyze per the conditions shown in Table 3. Concentrations obtained from the curve are multiplied by two to obtain the final concentration in the samples. Compounds CS1 CS2 CS3 CS4 CS5 CS6 CS7 CS8 CS9 CS10 CS11 CS12 CS13 CS14 CS15 Analyte All analytes unless otherwise noted 1 2.5 5 10 25 40 60 80 100 150 200 250 375 500 800 PFPeA 2 5 10 20 50 80 120 160 200 300 400 500 750 1000 1600 PFBA, 4:2-FTS, 6:2- FTS, 8:2-FTS 4 10 20 40 100 160 240 320 400 600 800 1000 1500 2000 3200 PFPrA, 5:3 FTCA, 7:3 FTCA 5 12.5 25 50 125 200 300 400 500 750 1000 1250 1875 2500 4000 NMeFOSE, NEtFOSE 10 25 50 100 250 400 600 800 1000 1500 2000 2500 3750 5000 8000 Surrogate 13C9-PFNA, 13C6- PFDA, 13C7-PFUnA, 13C2-PFDoA, 13C2- PFTreA 0.25 0.625 1.25 2.5 6.25 10 15 20 25 37.5 50 62.5 93.75 125 200 13C5-PFHxA, 13C4- PFHpA, 13C8-PFOA, 13C8-PFOSA, D3- NMeFOSA, D5- NEtFOSA, 13C8- PFOS, 13C3-PFBS, 13C3-PFHxS 0.5 1.25 2.5 5 12.5 20 30 40 50 75 100 125 187.5 250 400 13C5-PFPeA, 13C2- 4:2FTS,13C2-6:2FTS, 13C2-8:2FTS, D3- NMeFOSAA, D5- NEtFOSAA 1 2.5 5 10 25 40 60 80 100 150 200 250 375 500 800 13C4-PFBA, 13C3- HFPO-DA 2 5 10 20 50 80 120 160 200 300 400 500 750 1000 1600 D7-NMeFOSE, D9-NEtFOSE 5 12.5 25 50 125 200 300 400 500 750 1000 1250 1875 2500 4000
2.2 Analytical Conditions Table 3 Instrument Configuration and Analytical Conditions for ASTM D8421 PFAS using the Shimadzu LCMS-8060NX. Mobile Phase A: 2 mmol/L Ammonium Acetate in H2O/ Acetonitrile = 95/5 B: Acetonitrile Delay Column Shim-pack Scepter TM C18-120 2.1 mm x 100 mm, 3 µm (P/N: 227-31014-05) Analytical Column Shim-pack TM GIST-HP C18 3.0 mm x 100 mm, 3 um (P/N: 227-30040-04) Gradient (%B) 10% (0 min) 22% (2.3-3.0 min) 45% (6.0 min) 80% (13.0 min) 95% (14.0-16 min ) 10% (16.01-20.0 min ) Interface IonFocus ESI Column Oven Temp. 40 o C Flow rate 0.6 mL/min Injection volume Multiple draw injection program Co-injection 25 µL Sample 25 µL 0.1% Acetic acid in H2O Interface Temp. 170 Probe position +3 mm Neblizer gas flow 3 L/min Heating gas flow 15 L/min Interface Voltage -0.5 kV (same value for all compounds DL Temp. 200 Heatblock Temp. 300 Drying gas flow 5 L/min Focus bias -2 kV (same value for all compounds Results and Discussion A single laboratory validation of this method for specificity, linearity, recovery, and precision in nine wastewater matrices according to ASTM D8272 2 was previously described. 3 For this application news, a study was made to improve peak shape, particularly of early-eluting compounds, such as PFPrA and PFBA. This included evaluation of injection technique, columns, and flow rate. Co-injection of 25 µL sample with 25 µL 0.1% acetic acid in reagent water significantly improved the peak shapes of PFPrA, PFBA, and PFMPA (Fig. 2). A large diameter column with a long column length and large particle size, combined with a high flow rate, allowed greater axial diffusion, improving peak shape (Fig. 3). Finally, to better separate impurities from the mobile phase, a new delay column was chosen, and the gradient program was modified (Fig. 4). Upon optimization of chromatography and mass spectrometer conditions, calibration mixtures (Table 2) were prepared and used for subsequent analysis. Compound parameters, including quantitation ion, confirmation ion and collision energies, were optimized using LabSolutions TM software. At least two MRM transi- tions, if available, were used. Fig. 2 Optimization of injection technique to improve peak shape
Fig. 3 Example chromatograms for final column and flow rate with co-injection applied These compounds were chosen to illustrate because of their likelihood for regulation in wastewater. Additionally, calibration curves and midpoint chromatograms of PFPrA and NEtFOSE are shown in Figures 7 and 8. These compounds were chosen because they are the earliest and latest eluting compounds respectively. Linearity Study Calibration curves for each analyte were found by Shimadzu Lab Solutions Insight data processing software to have a % RSD of less than 30%, as required by ASTM D8421. Calibration data, MRM transitions for the quantitation and confirmation ions (when available), and retention times are shown in Table 4. Calibration curves along with a midpoint standard chromatogram of PFOA and PFOS are shown in Figures 5 and 6. Table 4 Summary of calibration data. Compound Quantitation Ion Confirmation Ion Retention Time (min) r 2 PFTreA 712.95>668.95 712.95>169.00 11.756 0.9962 PFTriA 662.95>618.95 662.95>169.00 11.026 0.9978 PFDoA 612.95>568.95 612.95>319.00 10.286 0.9987 PFUnA 562.95>518.95 562.95>269.00 9.543 0.9991 PFDA 512.95>468.95 512.95>219.00 8.813 0.9967 PFNA 462.95>418.95 462.95>219.00 8.11 0.9921 PFOA 412.95>369.00 412.95>169.00 7.451 0.9929 PFHpA 362.95>319.00 362.95>169.00 6.807 0.9989 PFHxA 312.95>269.00 312.95>119.00 6.028 0.9976 PFPeA 263.00>219.00 263.00>69.00 4.728 0.9996 PFBA 213.00>169.00 ---- 3.026 0.9985 PFDS 598.90>79.95 598.90>98.95 10.785 0.9971 PFNS 548.95>79.95 548.95>98.95 10.033 0.9975 PFOS 498.95>79.95 498.95>98.95 9.275 0.9951 PFHpS 448.95>79.95 448.95>98.95 8.522 0.9951 PFHxS 398.95>79.95 398.95>98.95 7.783 0.9917 PFPeS 348.95>79.95 348.95>98.95 7.059 0.9917 PFBS 298.95>79.95 298.95>98.95 6.17 0.9990 PFOSA 497.95>77.95 497.95>477.95 11.075 0.9979 8:2FTS 526.95>506.95 526.95>80.90 8.426 0.9976 6:2FTS 426.95>406.95 426.95>80.90 7.148 0.9960 4:2FTS 326.95>306.95 326.95>80.90 5.678 0.9970 Fig. 4 Final gradient with chromatogram of PFPrA, the earliest eluting peak
NEtFOSAA 584.00>418.95 584.00>526.00 9.01 0.9950 NMeFOSAA 569.95>418.95 569.95>482.95 8.703 0.9929 PFDoS 698.90>79.95 698.90>98.95 12.228 0.9959 NMeFOSA 511.95>219.00 511.95>169.00 13.556 0.9956 NEtFOSA 526.00>219.00 526.00>169.00 14.149 0.9988 NMeFOSE 616.00>59.00 ---- 13.246 0.9996 NEtFOSE 630.00>59.00 ---- 13.853 0.9998 HFPO-DA 285.00>169.00 285.00>185.00 6.365 0.9971 ADONA 376.95>251.00 376.95>85.00 7.064 0.9980 9Cl-PF3ONS 530.90>350.95 532.90>352.95 9.809 0.9994 11Cl-PF3OUdS 630.90>450.95 632.90>452.95 11.308 0.9994 PFPrA 163.00>119.00 ---- 1.589 0.9996 NFDHA 294.95>201.00 294.95>85.00 5.937 0.9953 PFEESA 314.95>135.00 314.95>82.95 6.628 0.9978 PFMPA 228.95>85.00 ---- 3.656 0.9981 PFMBA 278.95>85.00 ---- 5.279 0.9979 3:3 FTCA 241.00>177.00 241.00>117.00 3.804 0.9717 5:3 FTCA 341.00>237.00 341.00>217.00 6.375 0.9945 7:3 FTCA 441.00>317.00 441.00>337.00 7.752 0.9964 FHUEA 357.00>293.00 ---- 6.472 0.9962 FOUEA 456.95>393.00 ---- 7.704 0.9973 HQ-115 279.90>146.95 279.90>210.90 7.259 0.9988 13C4-PFBA_Surr 217.00>172.00 ---- 3.023 0.9982 13C5-PFPeA_Surr 268.00>223.00 ---- 4.726 0.9976 13C5-PFHxA_Surr 318.00>273.00 318.00>120.00 6.026 0.9972 13C4-PFHpA_Surr 367.00>322.00 ---- 6.806 0.9994 13C8-PFOA_Surr 421.00>376.00 ---- 7.45 0.9959 13C9-PFNA_Surr 472.00>427.00 ---- 8.108 0.9947 13C6-PFDA_Surr 519.00>474.00 ---- 8.81 0.9989 13C7-PFUnA_Surr 570.00>525.00 ---- 9.541 0.9979 13C2-PFDoA_Surr 614.95>569.95 ---- 10.285 0.9968 13C2-PFTreA_Surr 714.95>669.95 ---- 11.755 0.9952 13C8-PFOSA_Surr 505.95>77.95 ---- 11.077 0.9983 D3-NMeFOSAA_Surr 573.00>418.95 ---- 8.697 0.9933 D5-NEtFOSAA_Surr 589.00>418.95 ---- 9 0.9976 D3-NMeFOSA_Surr 515.00>219.00 515.00>168.90 13.548 0.9993 D5-NEtFOSA_Surr 531.00>219.00 531.00>168.90 14.131 0.9983 D7-NMeFOSE_Surr 623.05>59.00 ---- 13.206 0.9957 D9-NEtFOSE_Surr 639.10>59.00 ---- 13.807 0.9994 13C3-HFPO-DA_Surr 287.00>169.00 284.90>185.00 6.363 0.9921 13C2-4:2FTS_Surr 329.00>308.95 329.00>80.90 5.678 0.9943 13C2-6:2FTS_Surr 428.95>408.95 428.95>80.90 7.147 0.9903 13C2-8:2FTS_Surr 528.95>508.95 528.95>80.90 8.425 0.9956 13C8-PFOS_Surr 506.95>79.95 506.95>98.95 9.274 0.9966 13C3-PFBS_Surr 301.95>79.95 301.95>98.95 6.17 0.9953 13C3-PFHxS_Surr 401.95>79.95 401.95>98.95 7.782 0.9921 Fig. 5 Calibration curve and midpoint chromatogram for PFOA
Fig. 6 Calibration curve and midpoint chromatogram for PFOS Fig. 7 Calibration curve and midpoint chromatogram for PFPrA Fig. 8 Calibration curve and midpoint chromatogram for NEtFOSE
These data are well within the 70 -130 % recovery and ≤ 30 %RSD limits of the method. Recovery and Repeatability Stu dy Recovery and repeatability (Table 5) were evaluated in reagent water and wastewater, each spiked four times at the concentration indicated. Recovery was calculated after subtracting the native PFAS found in the unspiked sample matrices. Table 5 Recovery and Repeatability in Reagent Water and Wastewater Compound Spike Concentration (ppt) Reagent Water % Recovery Reagent Water %RSD (n=4) Wastewater % Recovery Wastewater %RSD (n=4) PFTreA 160 110 3.76 119 2.71 PFTriA 160 109 2.08 79.9 4.82 PFDoA 160 104 4.33 107 4.6 PFUnA 160 113 5.53 105 2.49 PFDA 160 113 2.67 102 4.5 PFNA 160 113 6.71 107 2.34 PFOA 160 111 7.52 112 5.7 PFHpA 160 116 4.13 108 4.46 PFHxA 160 114 6.83 115 3.41 PFPeA 320 106 4.37 108 2.47 PFBA 640 107 0.55 108 2.06 PFDS 160 112 8.89 112 3.73 PFNS 160 113 3.72 116 5.58 PFOS 160 110 3.83 122 4.02 PFHpS 160 115 6.09 102 6.91 PFHxS 160 113 5.93 113 13.15 PFPeS 160 124 4.94 119 9.49 PFBS 160 109 4.59 114 5.12 PFOSA 160 101 2.44 100 4.59 8:2FTS 640 113 6.24 103 4.97 6:2FTS 640 119 3.2 107 3.3 4:2FTS 640 121 0.7 100 2.47 NEtFOSAA 160 113 7.5 89.4 10.15 NMeFOSAA 160 111 13.35 88.0 7.42 PFDoS 160 106 5.23 108 9.31 NMeFOSA 160 102 3.68 91.5 5.59 NEtFOSA 160 100 0.73 90.5 3.33 NMeFOSE 1600 97.2 0.34 93.6 0.94 NEtFOSE 1600 96.8 0.9 93.5 1.39 HFPO-DA 160 109 2.35 112 9.66 ADONA 160 110 1.01 104 4.22 9Cl-PF3ONS 160 111 2.01 111 1.92 11Cl-PF3OUdS 160 112 3.59 111 2.78 PFPrA 800 108 1.8 105 0.77 NFDHA 160 107 9.13 110 4.16 PFEESA 160 112 4.71 115 4.45 PFMPA 160 106 2.37 102 5.83 PFMBA 160 113 7.07 115 2.55 3:3 FTCA 160 87.1 20.57 94.8 14.52 5:3 FTCA 800 95.7 6.91 90.4 4.22 7:3 FTCA 800 92.5 2.56 88.8 2.73 FHUEA 160 99.1 3.91 95.3 2.16 FOUEA 160 102 4.2 97.2 3.5 HQ-115 160 112 2.56 111 0.78 Surrogates 13C4-PFBA_Surr 320 102 1.7 96.4 2.88 13C5-PFPeA_Surr 160 108 6.03 96.4 3.36 13C5-PFHxA_Surr 80 110 3.7 101 5.35 13C4-PFHpA_Surr 80 104 6.43 105 5.99 13C8-PFOA_Surr 80 107 11.32 100 10.19 13C9-PFNA_Surr 40 98.2 13.76 92.2 17.29
13C6-PFDA_Surr 40 106 5.74 96.5 8.59 13C7-PFUnA_Surr 40 98.6 6.01 90.8 6.68 13C2-PFDoA_Surr 40 96.1 4.31 94.9 5.43 13C2-PFTreA_Surr 40 98.9 10.44 119 9.46 13C8-PFOSA_Surr 80 92.2 2.22 91.4 6.54 D3-NMeFOSAA_Surr 160 98.4 3.95 90.6 6.56 D5-NEtFOSAA_Surr 160 95.0 3.55 81.3 3.26 D3-NMeFOSA_Surr 80 89.3 12.1 82.6 8.57 D5-NEtFOSA_Surr 80 90.5 8.83 84.2 8.94 D7-NMeFOSE_Surr 800 88.8 0.49 85.4 1.68 D9-NEtFOSE_Surr 800 89.6 0.97 85.9 1.49 13C3-HFPO-DA_Surr 320 103 3.19 101 8.25 13C2-4:2FTS_Surr 160 116 3.15 96.2 6.88 13C2-6:2FTS_Surr 160 119 2.25 97.3 5.65 13C2-8:2FTS_Surr 160 104 0.73 97.4 7.53 13C8-PFOS_Surr 80 103 9.45 105 9.61 13C3-PFBS_Surr 80 121 5.06 94.0 6.33 13C3-PFHxS_Surr 80 116 9.99 108 6.65 Conclusion This application news demonstrates the analysis of 44 PFAS and 24 surrogate compounds in non-potable water by ASTM D8421 using the Shimadzu LCMS-8060NX Liquid Chromatography Mass Spectrometer (LC/MS/MS). Chromatographic conditions were optimized to achieve excellent peak shape, even for the earliest eluting compounds, such as PFPrA and PFBA. The highly sensitive Shimadzu LCMS-8060NX easily exceeds method performance criteria of the ASTM method and provides testing laboratories with highly accurate and reliable, repeatable results for PFAS in wastewater samples. References 1. ASTM Test Method D8241 Determination of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous Matrices by Co- solvation followed by Liquid Chromatography Tandem Mass Spectrometry (LC/MS/MS) 2. ASTM Standard D8272-19 -Standard Guide for Development and Optimization of D19 Chemical Analysis Methods Intended for EPA Compliance Reporting / ASTM International / West Conshohocken / PA / 2020 / 10.1520/D8272-19/ https://www.astm.org/ 3. Lipps, W., ASTM D8421-22 Standard Test Method for Determination of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous Matrices by Co-solvation followed by Liquid Chromatography Tandem Mass Spectrometry (LC/MS/MS), Shimadzu Whitepaper, August 2023.
Copyright © 2023 Shimadzu Corporation and/or its affiliates. All rights reserved. Shimadzu Corporation www.shimadzu.com/an/ Shimadzu Scientific Instruments, Inc. www.ssi.shimadzu.com Founded in 1875, Shimadzu Corpora�on, a leader in the development of advanced technologies, has a dis�nguished history of innova�on built on the founda�on of contribu�ng to society through science and technology. Established in 1975, Shimadzu Scien�fic Instruments (SSI), the American subsidiary of Shimadzu Corpora�on, provides a comprehensive range of analy�cal solu�ons to laboratories throughout North, Central, and parts of South America. SSI maintains a network o�en regional offices strategically located across the United States, with experienced technical specialists, service and sales engineers situated throughout the country, as well as applica�ons laboratories on both coasts. For information about Shimadzu Scien�fic Instruments and to contact your local office, please visit our Web site at www.ssi.shimadzu.com LCM S-8050 LCM S-8040 LCM S-8060NX LCM S-2020 Q-TOF LCMS -9030/9050 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “ ® ”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. LCM S-8045 LCM S-2050 LabSolutions, Shim-pack and Shim-pack Sceptor are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries. 02-SSI-LCMS-150A First Edition: Sep. 2023 Revision A: Feb.2024
Please fill out the survey Related Products Some products may be updated to newer models. LCMS-8060NX Liquid Chromatograph Mass Spectrometer LCMS-TQ RX Series Triple Quadrupole LC- MS/MS Triple Quadrupole LC-MS/MS Related Solutions Environment PFAS Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)