No. SSI-LCMS-151 Keywords: Per- and Polyfluorinated Alkyl Substances, PFAS, Perfluorinated Compounds, PFCs, PFOA, PFOS, Triple Quad, EPA Draft 1633, EPAM1633, aqueous samples, water, wastewater ■ Abstract This application note demonstrates that the Shimadzu LCMS-8050 meets and exceeds the Quality Assurance and Quality Control criteria and performance specified in Environmental Protection Agency (EPA) draft method 1633 for the analysis of Per- and Polyfluoroalkyl Substances (PFAS) in environmental samples. All analytes were reliably quantitated at or less than the Limit of Quantitation (LOQ) reported in the EPA draft method. This work proves the suitability of the Shimadzu LCMS-8050 for accurate and robust analysis of PFAS in routine laboratories in accordance with present demands in sensitivity and throughput. The method presented here can be optimized for other matrices or additional compounds. ■ Introduction The EPA is continuing to standardize methods for the analysis of PFAS in environmental samples using Liquid Chromatography with Triple Quadrupole Spectrometry (LC- MS/MS). Shimadzu offers a full line of Ultra-Fast LC-MS/MS (UFMS TM ) systems for quantitating PFAS at environmentally relevant concentrations. EPA Draft Method 1633 (EPAM1633) 1 is the latest method proposed by the EPA in 2022 for the quantitation of targeted PFAS in aqueous, solid, biosolid, and tissue samples. This application note, part of the comprehensive suite of Shimadzu’s vetted solutions for the analysis of PFAS 2 , demonstrates the performance of the Shimadzu LCMS-8050 for the quantitation of PFAS in accordance with EPAM1633. The LCMS-8050 was selected for this work because of its known dependability to deliver the sensitivity and robustness expected by modern environmental laboratories for the routine analysis of PFAS using methods requiring sample preparation by Solid Phase Extraction (SPE), like EPAM1633, or with a large-volume injection (e.g., EPA 8327, ASTM 8421-22). ■ Experimental Approach and Instrumentation This application note describes the analysis of the 40 target PFAS compounds, 23 Extracted Internal Standards (EIS), and 7 Non-Extracted Internal Standards (NIS) included in EPAM1633 in reagent water and wastewater using the Shimadzu LCMS-8050. PFAS types and acronyms used in EPAM1633 and This work are listed in Table 1. All standards used in the work presented here were purchased from Wellington Laboratories (PFAC-MXF, PFAC-MXG, PFAC-MXH, PFAC-MXI, PFAC-MXJ, MPFAC-HIF-ES, MPFAC-HIF-IS) as native mixes. User Benefits ◆ Limits of Quantitation listed in EPA Draft 1633 were readily achieved with the Shimadzu LCMS-8050, a system that delivers high-quality data, throughput, and affordability. ◆ The robust LCMS-8050 performs well with complex environmental matrices, such as wastewater samples used in this study. ◆ This efficient instrument is field upgradeable to higher sensitivity instruments, which makes it a versatile and future-proof solution for PFAS analysis. Liquid Chromatography Mass Spectrometry Analysis of Per-and Polyfluoroalkyl Substances (PFAS) using the LCMS-8050 Triple Quadrupole Mass Spectrometer According to EPA Draft Method 1633 Om Shrestha 1 , Hiroyuki Niwa 2 , Kazuhiro Kawakami 2 , Landon Wiest 1 , Marfil-Vega Ruth 1 1 Shimadzu Scientific Instruments, Inc., 2 Shimadzu Corporation
Table 1 Target analytes, EIS, and NIS included in this method Type Name Type Name Target PFBA EIS 13C4-PFBA Target PFPeA EIS 13C5-PFPeA Target PFHxA EIS 13C5-PFHxA Target PFHpA EIS 13C4-PFHpA Target PFOA EIS 13C8-PFOA Target PFNA EIS 13C9-PFNA Target PFDA EIS 13C6-PFDA Target PFUnA EIS 13C7-PFUnA Target PFDoA EIS 13C2-PFDoA Target PFTrDA EIS 13C2-PFTeDA Target PFTeDA EIS 13C3-PFBS Target PFBS EIS 13C3-PFHxS Target PFPeS EIS 13C8-PFOS Target PFHxS EIS 13C2-4:2FTS Target PFHpS EIS 13C2-6:2FTS Target PFOS EIS 13C2-8:2FTS Target PFNS EIS 13C8-PFOSA Target PFDS EIS D3-NMeFOSA Target PFDoS EIS D5-NEtFOSA Target 4:2FTS EIS D3-NMeFOSAA Target 6:2FTS EIS D5-NEtFOSAA Target 8:2FTS EIS D7-NMeFOSE Target PFOSA EIS D9-NEtFOSE Target NMeFOSA EIS 13C3-HFPO-DA Target NEtFOSA NIS 13C3-PFBA Target NMeFOSAA NIS 13C2-PFHxA Target NEtFOSAA NIS 13C4-PFOA Target NMeFOSE NIS 13C5-PFNA Target NEtFOSE NIS 13C2-PFDA Target HFPO-DA NIS 18O2-PFHxS Target ADONA NIS 13C4-PFOS Target 9Cl-PF3ONS -- TDCA Target 11Cl-PF3OUdS Target 3:3 FTCA Target 5:3 FTCA Target 7:3 FTCA Target PFEESA Target PFMPA Target PFMBA Target NFDHA Standards and Calibration Curve Preparation The native PFAS mixtures PFAC-MXF, PFAC-MXG, PFAC- MXH, PFAC-MXI, PFAC-MXJ, and methanol were mixed in a ratio of 20:10:10:10:12.5:37.5 for preparing Target Standard Solution I. Target standard solution II was prepared by diluting Target Standard Solution I 10 times with methanol. Native PFAS Mixtures MPFAC-HIF-ES, MPFAC-HIF, and methanol were mixed in a ratio of 15:15:270 for preparing the Internal standard solution. Calibration standards were prepared by mixing Target Standard Solution I, Target Standard Solution II, Internal Standard Solution, and methanol to achieve the desired concentrations listed in EPA draft method 1633 and Table 2 (targets: 0.2 – 250 ng/mL; EIS: 1.25 – 25 ng/mL; NIS: 1.25 – 5 ng/mL). Sample Preparation 500 mL of wastewater effluent or ultra-pure water were collected into pre-cleaned polyethylene bottles and weighed. 25 µL of EIS stock solution was added to each sample. Pre-conditioning of the SPE cartridges with 0.1% methanolic ammonium, methanol, and reagent water was performed according to the product manual (Inert Sep mini MA-2: 5010-27235, 280 mg). Samples were loaded onto the SPE cartridges at 5 mL/min. Sample bottles were rinsed with 10 mL of reagent water and those rinses were loaded onto the SPE cartridges. The cartridge was dried with nitrogen gas and the analytes were eluted with 5 mL of 0.1% methanolic ammonium. 25 µL of NIS was added to each extract.
Instrument and operational conditions The LC-MS/MS analysis was performed using a Shimadzu Nexera ultrahigh pressure liquid chromatography (UHPLC) system coupled with a triple quad LCMS-8050. A delay column was used in this work as the essential modification of hardware for minimizing possible PFAS background contamination from LC and solvents 3 . A description of the LC-MS/MS parameters is included in Table 3. Sample-to-sample run time of 20 minutes includes the re-equilibration for both the delay and analytical columns after final wash out with concentrated acetonitrile to flush the column, remove background residual contaminants, and restore column performance before starting the next run. All compound parameters, including precursor ion, product ions, and collision energies, were optimized using flow injection analysis (i.e., bypassing the analytical column) using LabSolutions TM software. Optimized retention times and precursor and product ions for multiple reaction monitoring (MRMs) are listed in Table 3. Table 3 Chromatography and mass spectrometer conditions Parameter Value LCMS Shimadzu LCMS-8050 Analytical column Shim-pack TM GIST-HP C18 2.1 × 50 mm, 3 µm Delay column Shim-pack GIST C18 3.0 × 50 mm, 5 µm Column oven temp 40 º C Injection volume 2 µL Mobile phase A: 2 mM Ammonium Acetate in 5 % (v/v) Acetonitrile in reagent water B: Acetonitrile Gradient flow rate 0.4 mL/ Min Run time 20 minutes Nebulizing gas flow 3 L/ Min Heating gas flow 15 L/Min Interface temperature 190 º C Desolvation line temperature 200 º C Heat block temperature 300 º C Drying gas flow 5 L/ min Acquisition cycle time 16 min Total MRMs 72 Table 2 Retention time, precursor ion, product ions, and calibration range used in this method. Type Name RT Precursor Ion Product Ion-1 Product Ion-2 Concentration CS1 (ng/mL) Concentration CS7 (ng/mL) Target PFBA 2.024 213.0 169.0 N/A 0.8 250 Target PFPeA 3.57 263.0 219.0 69.0 0.4 125 Target PFHxA 4.906 313.0 269.0 119.0 0.2 62.5 Target PFHpA 7.708 363.0 319.0 169.0 0.2 62.5 Target PFOA 8.64 413.0 369.0 169.0 0.2 62.5 Target PFNA 9.119 463.0 419.0 219.0 0.2 62.5 Target PFDA 9.511 513.0 469.0 219.0 0.2 62.5 Target PFUnA 9.901 563.0 519.0 269.0 0.2 62.5 Target PFDoA 10.461 613.0 569.0 319.0 0.2 62.5 Target PFTrDA 11.345 663.0 619.0 168.9 0.2 62.5 Target PFTeDA 12.237 713.0 669.0 168.9 0.2 62.5 Target PFBS 4.841 299.0 80.0 99.0 0.2 62.5 Target PFPeS 7.766 349.0 80.0 99.0 0.2 62.5 Target PFHxS 8.776 399.0 80.0 99.0 0.2 62.5 Target PFHpS 9.294 449.0 80.0 99.0 0.2 62.5 Target PFOS 9.709 499.0 80.0 99.0 0.2 62.5 Target PFNS 10.2 549.0 80.0 99.0 0.2 62.5 Target PFDS 10.956 598.9 80.0 99.0 0.2 62.5 Target PFDoS 12.524 698.9 80.0 99.0 0.2 62.5 Target 4:2FTS 4.404 327.0 307.0 80.9 0.8 250 Target 6:2FTS 8.398 427.0 407.0 80.9 0.8 250 Target 8:2FTS 9.323 527.0 507.0 80.9 0.8 250 Target PFOSA 11.438 498.0 78.0 478.0 0.2 62.5 Target NMeFOSA 13.178 512.0 219.0 169.0 0.2 62.5 Target NEtFOSA 13.475 526.0 219.0 169.0 0.2 62.5 Target NMeFOSAA 9.491 570.0 419.0 483.0 0.2 62.5 Target NEtFOSAA 9.647 584.0 419.0 526.0 0.2 62.5 Target NMeFOSE 13.016 616.0 59.0 N/A 2 625 Target NEtFOSE 13.305 630.0 59.0 N/A 2 625 Target HFPO-DA 5.693 285.0 169.0 185.0 0.8 250
Type Name RT Precursor Ion Product Ion-1 Product Ion-2 Concentration CS1 (ng/mL) Concentration CS7 (ng/mL) Target ADONA 8.209 377.0 251.0 85.0 0.8 250 Target 9Cl-PF3ONS 10.032 530.9 351.0 353.0 0.8 250 Target 11Cl-PF3OUdS 11.728 630.9 451.0 453.0 0.8 250 Target 3:3 FTCA 2.937 241.0 177.0 117.0 1 312.5 Target 5:3 FTCA 6.076 341.0 237.0 217.0 5 1562.5 Target 7:3 FTCA 8.952 441.0 317.0 337.0 5 1562.5 Target PFEESA 5.943 315.0 135.0 83.0 0.4 125 Target PFMPA 2.681 229.0 85.0 N/A 0.4 125 Target PFMBA 3.916 279.0 85.0 N/A 0.4 125 Target NFDHA 4.715 295.0 201.0 85.0 0.4 125 EIS 13C4-PFBA 2.025 217.0 172.0 172.0 10 10 EIS 13C5-PFPeA 3.568 268.0 223.0 223.0 5 5 EIS 13C5-PFHxA 4.904 318.0 273.0 120.0 2.5 2.5 EIS 13C4-PFHpA 7.707 367.0 322.0 322.0 2.5 2.5 EIS 13C8-PFOA 8.64 421.0 376.0 376.0 2.5 2.5 EIS 13C9-PFNA 9.119 472.0 427.0 427.0 1.25 1.25 EIS 13C6-PFDA 9.51 519.0 474.0 474.0 1.25 1.25 EIS 13C7-PFUnA 9.899 570.0 525.0 525.0 1.25 1.25 EIS 13C2-PFDoA 10.458 615.0 570.0 570.0 1.25 1.25 EIS 13C2-PFTeDA 12.236 715.0 670.0 670.0 1.25 1.25 EIS 13C3-PFBS 4.835 302.0 80.0 99.0 2.5 2.5 EIS 13C3-PFHxS 8.775 402.0 80.0 99.0 2.5 2.5 EIS 13C8-PFOS 9.71 507.0 80.0 98.9 2.5 2.5 EIS 13C2-4:2FTS 4.405 329.0 309.0 80.9 5 5 EIS 13C2-6:2FTS 8.397 429.0 409.0 80.9 5 5 EIS 13C2-8:2FTS 9.323 529.0 509.0 80.9 5 5 EIS 13C8-PFOSA 11.438 506.0 78.0 78.0 2.5 2.5 EIS D3-NMeFOSA 13.174 515.0 219.0 168.9 2.5 2.5 EIS D5-NEtFOSA 13.467 531.0 219.0 168.9 2.5 2.5 EIS D3-NMeFOSAA 9.489 573.0 419.0 419.0 5 5 EIS D5-NEtFOSAA 9.645 589.0 419.0 419.0 5 5 EIS D7-NMeFOSE 12.997 623.1 59.0 59.0 25 25 EIS D9-NEtFOSE 13.282 639.1 59.0 59.0 25 25 EIS 13C3-HFPO-DA 5.688 287.0 169.0 185.0 10 10 NIS 13C3-PFBA 2.023 216.0 172.0 N/A 5 5 NIS 13C2-PFHxA 4.905 315.0 270.0 119.0 2.5 2.5 NIS 13C4-PFOA 8.637 417.0 172.0 N/A 2.5 2.5 NIS 13C5-PFNA 9.118 468.0 423.0 N/A 1.25 1.25 NIS 13C2-PFDA 9.51 515.0 470.0 N/A 1.25 1.25 NIS 18O2-PFHxS 8.775 403.0 84.0 N/A 2.5 2.5 NIS 13C4-PFOS 9.708 503.0 80.0 99.0 2.5 2.5 -- TDCA 8.401 498.3 124.0 80.0 ■ Results Calibration was performed for all targeted PFAS using a seven-point calibration curve and following the recommendations included in EPA draft method 1633; concentrations ranged from 0.2 – 250 ng/mL for targets, 1.25 – 25 ng/mL for EISs, and 1.25 – 5 ng/mL as shown in Table 3. The LOQs for each target compound are listed in Table 3 and equivalent to CS1; all LOQs meet the concentration listed in EPAM1633. Fig. 1 shows the MRM transitions from CS1; this figure demonstrates the separation and peak shape of targets at the lowest concentration standard included in the calibration curve. Table 4 summarizes signal-to-noise (S/N), accuracy, and instrument linearity for the initial calibration, and the instrument detection limit (IDL) for all target compounds. The S/N of CS1 ranged from 8 to >30,000; this demonstrates that lower concentrations of the targeted PFAS could be easily measured with the LCMS-8050. Accuracy for all target compounds in CS1 ranged between 80% and 116%, exceeding results reported and accepted in EPAM1633. The instrument linearity for the calibration curve was evaluated by calculating the relative standard error (RSE); RSE for all target compounds was ≤20% in accordance with EPAM1633. The IDLs and %RSDs at the IDLs for each target are shown in Table 4. The IDLs ranged between 1.9 ng/mL (7:3 FTCA) and 0.01 ng/mL (multiple analytes), approximately one order of
magnitude lower than the LOQs. Fig. 1 MRM transitions of all PFAS (target, EIS, NIS) at the lowest calibration standard (CS1) concentration. Table 4 Method performance summary of neat standards using the Shimadzu LCMS-8050. Name CS1 Average S/N (n=7) %Accuracy CS1 (n=7) RSE IDL ng/mL (n=7) %RSD IDL (n=7) PFBA 12.36 102.68 3.64 0.16 5.77 PFPeA 18.73 105.25 2.98 0.18 12.52 PFHxA 25.93 106.18 4.75 0.13 21.55 PFHpA 7.85 102.14 4.77 0.11 15.03 PFOA 16.56 103.79 8.42 0.08 11.19 PFNA 26.59 105.89 7.06 0.10 15.65 PFDA 32.95 105.47 9.36 0.07 11.79 PFUnA 44.23 104.42 7.51 0.10 15.78 PFDoA 56.80 99.14 6.67 0.15 20.17 PFTrDA 47.26 109.92 6.73 0.11 16.01 PFTeDA 145.08 86.49 7.07 0.08 10.89 PFBS 106.61 104.01 8.32 0.12 23.82 PFPeS 57.58 116.00 8.97 0.12 16.54 PFHxS 287.07 103.13 9.56 0.18 31.19 PFHpS 96.74 100.94 10.68 0.07 10.80 PFOS 12.97 104.09 11.42 0.21 50.62 PFNS 32.95 113.14 10.88 0.12 16.77 PFDS 101.25 94.83 11.37 0.18 34.16 PFDoS 62.67 109.76 6.70 0.17 19.79 4:2FTS 28989.63 83.92 15.20 0.24 9.70 6:2FTS 24990.49 80.44 15.82 0.50 26.45 8:2FTS 34754.78 108.43 10.40 0.32 12.78 PFOSA 304.11 113.56 11.92 0.22 35.66 NMeFOSA 107.66 110.25 10.66 0.14 19.22 NEtFOSA 537.58 107.01 9.04 0.12 15.94 NMeFOSAA 111.13 102.36 5.69 0.09 15.71 NEtFOSAA 18.72 112.22 10.59 0.18 34.30 NMeFOSE 150.32 92.74 6.52 0.85 11.82 NEtFOSE 158.18 101.23 6.01 0.38 5.63 HFPO-DA 626.89 100.05 4.20 0.22 9.04 ADONA 2149.52 101.71 3.84 0.13 5.14 9Cl-PF3ONS 486.71 94.21 4.70 0.16 6.11 11Cl-PF3OUdS 671.33 111.72 9.08 0.35 14.03 3:3 FTCA 3924.23 104.45 8.49 0.31 12.02 5:3 FTCA 196.34 112.09 6.11 0.84 5.14 7:3 FTCA 1179.93 111.16 5.90 1.89 11.20 PFEESA 87.20 105.06 3.19 0.11 7.91 PFMPA 476.40 104.14 2.67 0.09 6.87 PFMBA 38489.86 113.02 7.73 0.14 10.19 NFDHA 2035.17 85.56 8.37 0.24 17.57
Repeatability was evaluated at CS1 (n=7), CS4 (n=6) and CS7 (n=6). In CS1, the lowest calibration standard, 75% of the targeted compounds showed a %RSD of less than 20%; the %RSD of 12.5% of the targets ranged between 20% and 30%, and %RSD was >30% for the remaining 12.5% of the targets (5 compounds). The %RSD for all targets in CS4 and CS7 (i.e., the mid-point and high point of the calibration curve, respectively) was <20%, with most compounds showing a %RSD of less than 10%. Fig. 2 The %RSDs of target compounds from CS4 (n=6) and CS7 (n=6). Two types of aqueous samples were analyzed following the extraction procedure outlined in EPAM1633: ultrapure reagent water (n=5) and wastewater (n=7). Both sample types were spiked at concentrations equal to those in CS1. Precision and accuracy are normally evaluated by spiking the samples with a mid-level concentration. CS1 was used instead to better assess the method performance at a more challenging concentration. Table 5 shows the %recoveries of the EISs in the ultrapure reagent water and wastewater samples. In ultrapure reagent water, the recoveries from all EISs were within 50% and 150%, except D5-NEtFOSA(48%). The recoveries of the EIS in wastewater were slightly lower for all compounds than in ultrapure reagent water. The %recoveries of the EIS in both types of samples exceed the reported ranges in EPAM1633. Fig. 3 shows the recovery of the target compounds spiked at CS1 concentrations in ultrapure reagent water (n=5) and wastewater (n=3), with error bars displaying the %RSD. The %recovery and % RSD for all the targeted PFAS in ultra-pure water and wastewater, except for 3:3 FTCA, are within acceptable ranges for the analysis of PFAS in environmental samples when spiked at the lowest standard of the calibration curve. Table 5 The %recoveries of EIS in ultra-pure water and wastewater samples. EIS % Recovery CS1 Spike in UPW (n=5) % Recovery CS1 Spike in WW (n=7) 13C4-PFBA 94.09 32.78 13C5-PFPeA 96.48 85.57 13C5-PFHxA 99.69 93.64 13C4-PFHpA 100.51 91.56 13C8-PFOA 104.16 93.28 13C9-PFNA 107.61 98.75 13C6-PFDA 104.58 89.06 13C7-PFUnA 97.69 72.58 13C2-PFDoA 82.12 58.55 13C2-PFTeDA 65.92 42.09 13C3-PFBS 96.04 92.31 13C3-PFHxS 105.44 100.97 13C8-PFOS 109.04 99.40 13C2-4:2FTS 86.72 82.76 13C2-6:2FTS 93.63 83.95 13C2-8:2FTS 93.09 84.67 13C8-PFOSA 72.53 85.68 D3-NMeFOSA 52.33 54.40 D5-NEtFOSA 47.86 44.46 D3-NMeFOSAA 102.61 79.32 D5-NEtFOSAA 91.70 61.64 D7-NMeFOSE 56.78 49.43 D9-NEtFOSE 53.77 45.35
■ Summary and Conclusions The Shimadzu LCMS-8050 was evaluated for its ability to analyze PFAS in wastewater samples in accordance with EPAM1633. The results demonstrated the excellent performance of the LCMS-8050 for key quality control and performance parameters defined in the draft method and indicate that lower detection limits are easily achievable. ■ References 1. United States Environmental Protection Agency, Office of Water. 3rd Draft Method 1633 Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous, Solid, Biosolids, and Tissue Samples by LC/MS/MS, 2022. https://nepis.epa.gov/Exe/ZyPURL.cgi?Dockey=P101687F.txt (accessed June 1, 2023) 2. https://www.ssi.shimadzu.com/products/liquid-chromatograph-mass-spectrometry/lc-ms-system/lcms-ms- ms-method-package-pfas/index.html 3. https://www.ssi.shimadzu.com/sites/ssi.shimadzu.com/files/pim/pim_document_file/ssi/technical/ technical_reports/16761/2023_PFAS_Essentials_111523.pdf Fig. 3 %Recovery of CS1 in ultrapure reagent water and wastewater.
LCMS-8050 LCMS-8040 LCMS-8060NX LCMS-2020 Q-TOF LCMS-9030/9050 LCMS-8045 LCMS-2050 Copyright © 2023 Shimadzu Corporation and/or its affiliates. All rights reserved. Shimadzu Corporation www.shimadzu.com/an/ Shimadzu Scientific Instruments, Inc. www.ssi.shimadzu.com Founded in 1875, Shimadzu Corporation, a leader in the development of advanced technologies, has a distinguished history of innovation built on the foundation of contributing to society through science and technology. Established in 1975, Shimadzu Scientific Instruments (SSI), the American subsidiary of Shimadzu Corporation, provides a comprehensive range of analytical solutions to laboratories throughout North, Central, and parts of South America. SSI maintains a network of ten regional offices strategically located across the United States, with experienced technical specialists, service and sales engineers situated throughout the country, as well as applications laboratories on both coasts. For information about Shimadzu Scientific Instruments and to contact your local office, please visit our Web site at www.ssi.shimadzu.com LabSolutions, UFMS and Shim-pack are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries. 02-SSI-LCMS-151A First Edition: Sep. 2023 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “ ® ”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. Revision A: Feb.2024