User Benefits
Application
News
Fig. 1 Chromatogram of Mixed Standard Sample of 8 Compounds (100 ppm)
As representative compounds used in electrolytes, the 8 types
of carbonic ester compounds listed in Table 2 were prepared.
These compounds were diluted with dichloromethane, and 8-
compound mixed standard samples were prepared for the 6
points of 10 mg/L, 25 mg/L, 50 mg/L, 100 mg/L, 200 mg/L, and
500 mg/L. Calibration curves were prepared and repeatability
was confirmed using these standard samples.
Analysis of Standard Sample
Table 1 Analysis Conditions
Table 2 Types of Carbonic Esters and Additives Used
Diffusion of lithium ion batteries, which are used in electric
vehicles (EVs) and power storage systems, has increased in recent
years because this type of battery achieves high speed
charging/discharging and long battery life. Lithium ion batteries
consist of electrodes, separators, and an electrolyte, but among
these components, the composition and purity of the carbonic
ester and additives in the electrolyte are critical for maintaining
the battery quality and performance. These compounds can be
analyzed with a gas chromatograph (GC).
The lithium ion battery market has expanded rapidly in recent
years, and analysis of the electrolytes used in batteries is
indispensable as a key element of quality control. In quality
control, it is necessary to analyze as many samples as possible in a
limited laboratory space. Under these conditions, one particularly
effective option is the Shimadzu Brevis GC-2050, which boasts
outstanding analysis performance combined with a space-saving
design.
This Application News article introduces an example in which
several types of carbonic esters and additives which are
frequently used in electrolytes were analyzed with a Brevis GC-
2050+FID detector. In this analysis, N
2
was used as the carrier gas.
Introduction
Carbonic esters and additives in lithium ion battery electrolytes can be analyzed with high separation and good repeatability by
using N
2
as the carrier gas.
The Brevis GC-2050+FID is the optimum instrument for routine analysis thanks to its space-saving design.
Analysis of Carbonic Esters and Additives
in Lithium Ion Battery Electrolytes
Misato Ishimoto
Brevis™ GC-2050 Gas Chromatograph
Table 1 shows the analysis conditions. A PTFE syringe (P/N 221-
74469) was used.
Analysis Conditions
Model : Brevis GC-2050
Inj. temp. : 250 ˚C
Inj. mode : Split 30:1
Carrier gas : N
2
, constant linear velocity (25 cm/s)
Column : SH-I-5MS (P/N 221-75940-30)
(30 m × 0.25 mm I.D., 0.25 μm)
Column temp. : 40 ˚C (3 min) → 10 ˚C/min → 160 ˚C (5 min)
Detector : FID
FID temp. : 250 ˚C
Makeup gas : N
2
, 24 mL/min
H2 flow : 32 mL/min
Air flow : 200 mL/min
Compound name Abbreviation
1 Dimethyl carbonate DMC
2 Ethyl methyl carbonate EMC
3 Vinylene carbonate VC
4 Diethyl carbonate DEC
5 Fluoroethylene carbonate FEC
6 Ethylene carbonate EC
7 Propylene carbonate PC
8 1,3-propanesultone PS
Fig. 1 shows the total chromatogram for the 100 ppm standard
sample.
Chromatogram Results of 8-Compound
Mixed Standard Samples
1 DMC
2 EMC
3 VC
4 DEC
5 FEC
6 EC
7 PC
8 PS
1
2
3
4
5
6
7
8
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0 min
Fig. 2 shows the calibration curve results for the respective
compounds. All compounds showed satisfactory linearity, with
R
2
= 0.999 or higher. Table 3 shows the peak area repeatability
(%RSD) for 5 repeated analyses (n = 5) at 10 mg/L. Good
repeatability results were obtained for all compounds.
Calibration Curve Results of 8-Compound
Mixed Standard Samples
EMC
R
2
>0.9998
0 100 200 300 400 Conc.
0
25000
50000
75000
100000
125000
150000
Area
DMC
R
2
>0.9999
0 100 200 300 400 Conc.
0
10000
20000
30000
40000
50000
60000
70000
80000
Area
DEC
R
2
>0.9998
0 100 200 300 400 Conc.
0
25000
50000
75000
100000
125000
150000
175000
Area
VC
R
2
>0.9998
0 100 200 300 400 Conc.
0
10000
20000
30000
40000
50000
60000
70000
80000
90000
100000
Area