■ Introduction Endogenous steroids are essential to the regulation of several metabolic pathways including energy metabolism, stress, and fertility. Dried blood spots (DBS) (Fig. 1) offer an alternative to conventional venipuncture blood collection by allowing less invasive sample collection at home. A challenge for the lab is the small sample size collected from the dried blood spot resulting in low concentrations of endogenous analytes. To assist with accurate quantitation of select steroids in this biological matrix, Dispersive Pipette XTRaction (DPX) technology was utilized for increased sensitivity and selectivity by reducing matrix interference. A Shimadzu LCMS-8060NX triple quadrupole mass spectrometry with its sensitive quantitation capability was used to analyze four endogenous steroids (Fig. 2). Optimized source conditions and MRM transitions on the mass spectrometer were used to detect clinically relevant concentration limits in matrix. Correlation between traditional immunoassay assays and this newly developed LC-MS/MS method was determined using eight unique human serum samples. Fig. 1: Two types of DBS cards. Left card produces ~30 µL of serum; right produces ~6 µL serum. Fig. 2: Four steroids quantitated in this LC-MS/MS method. ■ Experimental Materials and Sample Preparation A six-point calibration curve with three QC levels were prepared from commercially available steroid standards (Cerilliant, TX) spiked in stripped human serum (Golden West, CA). Correlation samples were prepared by charging blank ADx cards with serum from venous drawn samples. The ADx sample cards were dried overnight and stored in the aluminum Ziploc sealed shipping bags provided with the cards. Thirty microliter aliquots of each calibrator and QC were plated, while the ADx card samples were paper punched into culture tubes. All samples were spiked with internal standard and submerged in a water:methanol solution and incubated on a plate- rocker for 25 min. The samples were then transferred to a well plate and the reconstituted protein was digested with 0.5% formic acid in 50% methanol in water prior to dSPE. Testosterone Progesterone Cortisol DHEAS LCMS-8060NX Liquid Chromatograph Mass Spectrometer Quantitation of Endogenous Steroids in Serum Using Dried Blood Spot Serum Separator Card and Triple Quadrupole Mass Spectrometry Vikki Johnson Shimadzu Scientific Instruments Application News User Benefits The Shimadzu LCMS-8060NX provides clinically relevant sensitivity and selectivity for quantifying four steroids from serum separator cards. The small sample volume obtained from serum separator cards simplifies the processes of sample collection, storage, and transportation. Increased throughput is achieved by simultaneously analyzing multiple analytes in a single run, compared to traditional ELISA, which requires separate assays for each analyte.
Each sample was exposed to DPX XTR tips containing 20 mg Supel™ Swift HLB (60 µm) (DPX170357) sorbent to isolate the target analytes from the matrix and further remove interferences using the Bind- Wash-Elute protocol as shown in Fig. 3. DPX’s patented technology provides dispersive SPE in a pipette tip that improves recovery by allowing the loose sorbent to mix more efficiently with the sample solution and increases the available surface area of the sorbent for more adequate sample binding than other traditional SPE products. Fig. 3: DPX Sample Preparation Procedure. Bind and wash solution: 50% methanol in water. Elution solution: acetonitrile. Analytical Methods Standards and samples were analyzed using the LCMS-8060NX triple quadrupole mass spectrometer coupled to Nexera HPLC equipped with a pump, autosampler, column oven as shown in Fig. 4. Two column stationary phases (biphenyl and C18) were tested for evaluation. The biphenyl column was used for initial evaluation for correlation samples. The C18 column resulted in less matrix interferences with minimal gradient adjustments for spiked serum calibration and QC samples. The analytical conditions for both columns tested for this evaluation along with MS/MS conditions are listed in Table 1. Fig. 4: LCMS-8060NX configuration. The IonFocus design improves signal intensity with higher gas flows and higher effective temperatures. The probe position can be moved further from the MS inlet without losing signal intensity. Table 1: HPLC and MS/MS Conditions [LC Conditions 1] [LC Common Conditions] Column: Shim-pack Velox C18 (100 mm x 3.0 mm I.D., 2.7 µm Mobile Phase A: 0.1 mM ammonium fluoride in water Gradient Program: B 50% (0-0.1 min); B 95% (0.1-4 min); B 95% (4.0-5 min); B 50% (5.01-7 min) Mobile Phase B: Methanol Flow Rate: 0.6 mL/min [MS/MS Conditions] Column Temperature: 40 o C Ionization: ESI (+) and (-) mode Injection Volume: 10 µL Mode: MRM [LC Conditions 2] Interface Voltage: +3.0 kV/-3.0 kV Column: Shim-pack Velox Biphenyl (100 mm x 3.0 mm I.D., 3.0 µm Focus Voltage 2 kV Gradient Program: B 30% (0 min); B 55% (0.0-1 min); B 80% (1.0-4.5 min); B 80% (4.5 min-6.3 min); B 98% (6.30-6.31 min); B 98% (6.31 min-7.31 min); B 30% (7.32 min-8.7 min) Nebulizing gas flow: 3 L/min Flow Rate: 0.8 mL/min Drying gas flow: 15 L/min Column Temperature: 50 o C Heating gas flow: 25 L/min Injection Volume: 20 µL DL Temp: 250 o C Heat Block Temp: 500 o C MRM Transitions The quantitative and qualitative MRM transitions used for analysis are shown in Table 2. Table 2: MRM transitions Analyte Ret. Time, (min), biphenyl Ret. Time, (min), C18 Quant. Transition Collision Energy (V) Qual. Transition Collision Energy (V) Polarity DHEAS-d5 2.023* 2.323* 372.10>98.0 13.0 --- --- - DHEAS 2.029* 2.401* 367.05>97.05 13.0 --- --- - Cortisol-d5 3.021 2.136 367.20>121.10 -23.0 --- --- + Cortisol 3.034 2.142 363.20>105.10 -21.0 363.20 -23.0 + Testosterone-d3 4.824 3.120 292.20>109.10 -23.0 --- --- + Testosterone 4.834 3.125 289.20>109.10 -23.0 289.20>97.10 -22.0 + Progesterone-d9 6.608 3.787 324.30>113.15 -11.0 --- --- + Progesterone 6.690 3.812 315.20>109.05 -11.0 315.20>97.10 -17.0 + *adequate column conditioning should be completed to avoid retention time shifting for DHEAS
■ Results Calibrators and QCs Representative calibration curves and chromatograms for calibrator 1 using 30 µL serum aliquots for biphenyl and C18 column phases are show in Fig. 5 and Fig. 6. Correlation of coefficient (r 2 ) values were above 0.99 for all analytes and percent accuracy was between 80-120 % for all calibration and QC levels. Intra-day (n=6) and Inter-day (n=21, 4 days) for replicate QC injection were less than 8% RSD as shown in Table 3. Fig. 5: Calibration curves and chromatogram for calibrator 1 on C18 column. Fig. 6: Calibration curves and chromatogram for calibrator 1 on biphenyl column.
Table 3: Intra- and inter-day precision for each QC level on C18 column. Analyte Sample ID Intra-day %RSD (n=6) Inter-day %RSD (n=21, 4 days) Cortisol LowQC 2.46 2.26 MidQC 1.15 1.31 HighQC 3.04 2.49 DHEAS LowQC 2.52 2.3 MidQC 2.64 2.38 HighQC 3.4 3.85 Testosterone LowQC 4.84 5.39 MidQC 1.81 1.94 HighQC 3.65 2.97 Progesterone LowQC 5.46 7.2 MidQC 2.98 2.22 HighQC 2.61 3.28 Correlation Correlation comparing immunoassay results for eight human venous drawn samples to this newly developed quantitative method is shown in Table 4. Each sample was charged with 30 µL of serum on ADx cards. The percent difference for all analytes compared to immunoassay results were less than 30% for all analytes in all samples. A representative sample chromatogram is shown in Fig. 7. Table 4: Correlation comparing LCMS results from charged ADx DBS cards vs. immunoassay Fig. 7: Chromatogram for Sample B on biphenyl column ■ Conclusion A highly sensitive and accurate LC-MS/MS method was developed for the quantification of four steroids on ADx DBS cards, with calibration ranges for all compounds within published reference ranges. The sensitivity of LC-MS/MS analysis using the small sample volume from DBS has been comparable to classical immunoassay for all analytes. This method has shown that fingerstick collection on DBS cards could potentially be a substitute to conventional venipuncture blood collection.
Copyright © 2024 Shimadzu Corporation and/or its affiliates. All rights reserved. SHIMADZU Scientific Instruments Inc. www.ssi.shimadzu.com Shimadzu Corporation www.shimadzu.com/an/ Founded in 1875, Shimadzu Corporation, a leader in the development of advanced technologies, has a distinguished history of innovation built on the foundation of contributing to society through science and technology. Established in 1975, Shimadzu Scientific Instruments (SSI), the American subsidiary of Shimadzu Corporation, provides a comprehensive range of analytical solutions to laboratories throughout North, Central, and parts of South America. SSI maintains a network of ten regional offices strategically located across the United States, with experienced technical specialists, service and sales engineers situated throughout the country, as well as applications laboratories on both coasts. For information about Shimadzu Scientific Instruments and to contact your local office, please visit our Web site at www.ssi.shimadzu.com LCMS-8050RX LCMS-8060RX LCMS-2020 Q-TOF LCMS-9030/9050 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. LCMS-8045RX LCMS-2050 02-SSI-LCMS-152-EN First Edition: September 2023, Revision A: July 2024
Please fill out the survey Related Products Some products may be updated to newer models. LCMS-TQ RX Series Triple Quadrupole LC-MS/MS LCMS-8060NX Liquid Chromatograph Mass Spectrometer Related Solutions Clinical Research and Forensics Clinical Research Forensic Analysis Blood Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)