Application
News
www.shimadzu.com/an/
Shimadzu Corporation
Copyright © 2024 Shimadzu Corporation and/or its affiliates. All rights reserved.
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
See http://www.shimadzu.com/about/trademarks/index.html for details.
Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not
they are used with trademark symbol “TM” or “”.
Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
to change without notice.
SHIMADZU Europa GmbH,
www.shimadzu.eu
First Edition: July. 2024 05-SCA-280-108-EN
AOC, LabSolutions Insight and Nexis are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries.
Results and Discussion
To analyze the phosphate species the SCAN mode on the GCMS
was used, which enabled the identification of the different
analytes based on their spectra. Since most compounds were
not available in the most common libraries the spectra from the
available scientific publications were used as a reference [1].
The corresponding retention times, used m/z traces and the
detected areas of the different compounds are summarized in
the Table 1.
Figure 3 Mass spectra of the detected fluorophosphates in the aged LIB, DMFP
(left), EMFP (middle), DEFP (right)
Figure 4 GCMS chromatogram of the TMP, EDMP and DEMP in a new LIB
(left) and after 1500 cycles (right)
Table 1 Chemical structures and the specific m/z traces for identification
Conclusion
The presented application demonstrated the suitability of
GCMS to investigate the aging of the LIBs using phosphate-
based degradation products. This investigation is based on
SCAN measurements. Here, a SCAN/SIM mode would be
beneficial offering the possibility to detect the main
compounds (carbonated, additives) and the degradation
products in one measurement. A further improvement would
be a dedicated SIM method for the mentioned phosphates with
the highest selectivity and sensitivity.
Compound
Chemical
structure
m/z for
SIM
Ret. Time
[min]
Peak area
New
battery
After
1500
cycles at
45 °C
Dimethyl
fluorophosph
ate (DMFP)
97, 98,
128
3.83 15304 11406
Ethyl methyl
fluorophosph
ate (EMFP)
97, 115,
127, 141
5.64 7015 14619
Diethyl
fluorophosph
ate (DEFP)
101, 113,
129
8.05 2136 3426
Trimethyl
phosphate
(TMP)
140, 110,
109, 95
9.12 N.D 3952
Ethyl dimethyl
phosphate
(EDMP)
153, 139,
127, 110,
109, 96,
95
10.11 N.D 1028
Diethyl methyl
phosphate
(DEMP)
141, 113 10.77 N.D 588
Triethyl
phosphate
(TEP)
155, 127,
109, 99
N.D N.D N.D
The fluorinated species could be detected in both, the new and
the aged battery. The content of EMFP and DEFP in the aged
battery is significantly higher in comparison to the new one,
and at the same time the content of DMFP is slightly lower. The
reason for this result is the very fast formation of fluorinated
compounds and their dependence on many external factors
(sample preparation, time of storage, moisture). As mentioned
above, the fluorinated compounds are more suitable to analyze
the quality of the manufactured electrolyte, e.g. to investigate
the destruction due to storage conditions like moisture levels. In
contrast to the fluorinated compounds, the tri-alkylated
phosphates are formed significantly slower and are usually not
detectable in a fresh electrolyte, offering the possibility to
investigate their formation as an indicator of battery aging. The
obtained GCMS chromatograms and the corresponding MS
spectra are shown in Figure 3.
As it is shown in Figure 4, TMP, EDMP and DEMP could be
detected in the electrolyte after 1500 charge cycles, TEP was not
observed in this experiment, nevertheless TEP might be
detected in a sample with a more advanced aging.
Acknowledgment
We would like to thank Münster Electrochemical Energy
Technology (MEET), Münster, Germany for great support in
producing this application.
Literature
[1] W. Weber, V. Kraft, M. Grützke, R. Wagner, M. Winter, S. Nowak,
Identification of alkylated phosphates by gas chromatography–mass
spectrometric investigations with different ionization principles of at
thermally aged commercial lithium-ion battery electrolyte, 2015, 1394, 128-
136.