User Benefits Application News Fig. 1 Compact model GC Brevis™ GC-2050 (Left), and Headspace sampler HS-20NX (Right) Figure 1 shows the new compact model GC, the Brevis GC-2050, used for this measurement. Table 1 shows the Instrument configuration and analysis conditions used in this measurement. The analysis column used was a 624 column, which is a medium-polarity column of the cyano-propyl-phenyl type designed to separate polar solvents like NMP and its similar components. The headspace method maintains a vial containing the sample at a constant temperature to measure the gas in the vapor phase or gas-solid equilibrium state. Even for samples containing solid samples or compounds with low volatility, GC measurements can be performed simply by sealing a certain amount of sample in a vial. By setting the headspace vial at relatively high temperature, the experiment examined the amount of NMP in the NCM cathode under the assumption that most of the remaining NMP in the cathode is in the vapor phase inside the vial. Residual NMP analysis in cathode Prepared NMP acetone solutions of 10, 100, and 1000ppm as standard solutions, and injected 2 µL each into 20 mL headspace vials. Conducted measurements five times in a row, and the relative standard deviations and calibration curves obtained are shown in Table 2 and Figure 2. The density of NMP was calculated as 1.03 g/cm 3 . Lithium-ion batteries (LiB) consist of cathode and anode, electrolyte, separator, etc. The electrodes are formed by coating a slurry of active material, conductive additives, binders, etc., mixed in an organic solvent, on a metal film (current collector) and then dried. N-Methyl-2-pyrrolidone (NMP) is known as an organic solvent used in solvent-based slurries, and especially for quality control of the cathode, the residual amount of NMP in the cathode is examined during the drying process. In this application, a simple analysis method for residual NMP in the cathode of lithium-ion battery NCM (nickel-cobalt- manganese ternary material) using GC-FID with headspace method is introduced. In addition, results of qualitative analysis of other solvents remaining in the NCM cathode using GC-MS and comparison of residual solvent amounts in five different cathodes with different drying processes are also presented. Introduction Analyze residual NMP and similar solvent structures in the cathode of lithium-ion batteries. Optimize drying conditions in the cathode manufacturing process and quantify the amount of residual solvents for quality control. Minimize equipment footprint by combining Brevis GC-2050 and GCMS-QP2050. Residual NMP and Solvent Analysis in Lithium-ion Battery Cathode Arisa Suzuki, Yu Nagao GC Brevis TM GC-2050 / HS-20 NX GC-MS GCMS-QP2050 / HS-20 NX Model : Brevis GC-2050 Detector : FID (Flame Ionization Detector) Column : SH-I-624Sil MS (0.32 mm I.D.×30m, d.f.= 1.8 µm) P/N:227-36077-01 Column Temp. : 100˚C – 5˚C /min – 180˚C (3 min) Injection Mode : Split 20 Carrier Gas : Linear velocity 30 cm/sec (He) Detector Temp. : 250˚C FID H 2 Flow Rate : 32 mL/min FID Make up Flow Rate : 24 mL/min (N 2 ) FID Air Flow Rate : 200 mL/min Table 1 Analysis of NMP in the cathode Instrument configuration and analysis conditions GC Model : HS-20 NX USTL (Ultra Short Transfer Line) Oven Temp. : 180˚C Sample Line Temp. : 200˚C Transfer Line Temp. : 200˚C Vial Equilibrating Time : 30 min Vial Pressurizing Time : 1.0 min Vial Pressure : 65 kPa (N 2 ) Loading Time : 0.5 min Load Equilib. time : 0.1 min Needle Flush Time : 1.0 min HS STD ppm NMP ng Area (n=5) Area RSD% 10 20.6 311 3.18 100 206 3485 1.19 1000 2060 38003 1.94 NMP(ng) Area R2>0.9999 Sample preparation and Result Table 2 Result of standard solution Fig. 2 Calibration curve of NMP 40000 30000 20000 10000 0 0 500 1000 1500 2000
Application News Residual solvent Analysis in cathode Considering the differences in residual amounts depending on the cutting location, three cut samples were measured, and using the calibration curve shown in Figure 2 for quantification, the NMP detected from the cathode was found to be an average of 33.5 ng. Comparing this to the mass of the sealed cathode, a residual amount of 2.76 ppm (w/w) was calculated (Table 3). 0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 min STD 100ppm Cathode NMP 10.0 10.5 11.0 11.5 12.0 min Expanded view NMP No. Cathode sheet mg Detected NMP ng Residual NMP ppm(w/w) 1 11.30 32.30 3.12 2 12.51 34.40 2.75 3 12.79 30.85 2.41 Average 33.5 2.76 Model : Nexis GC-2030 Column : SH-I-624Sil MS (0.32 mm I.D.×60m, d.f.= 1.8 µm) P/N:221-75963-60 Column Temp. : 100˚C – 5˚C/min – 180˚C (15 min) Injection Mode : Split 20 Carrier Gas : Linear velocity 40 cm/sec (He) GC Model : HS-20 NX (Trap model) Mode : Loop mode Oven Temp. : 180 ˚C Sample Line Temp. : 200 ˚C Transfer Line Temp. : 200 ˚C Vial Equilibrating Time : 30 min Vial Pressurizing Time : 1.0 min Vial Pressure : 65 kPa (N 2 ) Loading Time : 0.5 min Load Equilib. time : 0.1 min Needle Flush Time : 1.0 min HS Model : GCMS-QP2050 (TMP exhaust: 255 L/sec) Ion Source Temp. : 200˚C Interface Temp. : 200˚C Acquisition Mode : Scan Event Time : 0.3 sec m/z Range : 35 – 500 MS Following this, the cathode sheet was cut into 10 mm squares, with the electrode surface facing upwards for easier solvent evaporation, and sealed in 20 mL headspace vials. The obtained chromatograms and an expanded view of the NMP elution position are shown in Figure 3. It was found that trace amounts of residual NMP were detected. Table 3 Quantification of residual NMP Fig. 3 Chromatogram of cathode and enlarged view of NMP elution position Fig. 4 Gas chromatograph Nexis TM GC-2030 and Mass Spectrometer GCMS-QP2050 and Headspace sampler HS-20NX(Trap model) Table 4 Analysis of residual solvent in the cathode Instrument configuration and analysis conditions Fig. 5 MS spectrum at retention time 13.9 min (top) and the MS spectrum of the hit NMP by library search (bottom) When examining the chromatogram in Figure 3, it is evident that residual solvents other than NMP are present. In cases where unknown residual solvents or peaks detected vary by product, qualitative analysis of compounds by mass spectrometry is useful in quality control during the cathode manufacturing process. The compact model GCMS-QP2050 (Figure 4) is a user-friendly gas chromatograph mass spectrometer that is easy to maintain and can be retrofitted to existing GC systems. Table 4 shows the device configuration and analysis conditions used for qualitative analysis of residual solvents in NCM cathode. Here, we use the expandable Nexis TM GC-2030 in the front GC, but the qualitative results do not change depending on the GC model. Similar to the analysis of NMP, the cathode sheet is cut into 10 mm squares, sealed in a 20 mL headspace vial, and the obtained TIC chromatogram is shown in Figure 6. Five main components (A to E) can be identified, and spectral searches using the NIST library revealed compounds similar to NMP as candidate compounds for each peak (Table 5). Based on the MS chromatogram of m/z 99 zoomed in on Figure 6 and the MS spectrum in Figure 5, it is considered that NMP elutes around 13.9 minutes. The separation characteristics of the 624 column suggest that NMP and similar compounds to NMP are well separated.
Application News www.shimadzu.com/an/ Shimadzu Corporation © Shimadzu Corporation, 2024 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. First Edition: Aug. 2024 01-00738-EN Comparison of residual solvents with cathodes drying processes Brevis and Nexis are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries. Residual solvent levels were compared for five different NCM cathodes with varying drying processes using the instrument method in Table 1 with headspace GC-FID. The chromatogram comparisons obtained are shown in Figure 7, and the comparison of the area values of major organic solvents A to E is shown in Figure 8. It was confirmed that the amount of residual organic solvents remaining in the cathode varies depending on the drying conditions. By using headspace GC-FID, it is possible to easily analyze residual NMP in the cathode and quantify residual solvents without the need for preprocessing. Additionally, with headspace GC-MS, qualitative analysis of residual organic solvents can also be performed. Using compact models such as Brevis GC-2050 or GCMS-QP2050 allows for minimization of experimental space and detailed analysis of solvent analysis for each drying process. This can be applied to quality control and research and development in the cathode manufacturing process. <Acknowledgment> We would like to express our sincere gratitude to Dainen material Co., Ltd. for their significant cooperation in providing cathodes and analyzing experimental results for the development of this application. <Related Applications> 1. Analysis of battery electrolytes and N-methyl-2-pyrrolidone (NMP) via headspace GC-FID- Application News No.05-SCA- 180-049-EN 2. Analysis of Carbonic Esters and Additives in Lithium Ion Battery Electrolytes- Application News No.01-00708-EN 13.0 13.5 14.0 14.5 15.0 min 99.00(100.00) TIC NMP 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0 min A B C D E Cathode No. Drying process ˚C Time min 1 80 10 2 80 30 3 80 60 4 120 30 5 80 + Vacuum(120) 30 + One night A B C D E 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 min cathode 5 cathode 4 cathode 3 cathode 2 cathode 1 0.0 500.0 1000.0 1500.0 2000.0 2500.0 3000.0 3500.0 4000.0 4500.0 A B C D E Area/mg Candidate compounds CAS No. similarity A N-Methylmaleimide 930 - 88 - 1 97 B 1,3,4-Trimethyl-2-pyrazoline 14044 - 41 - 8 84 C 3-Pyrrolin-2-one, 1-methyl- 13950 - 21 - 5 91 D 2,5-Pyrrolidinedione, 1-methyl- 1121 - 07 - 9 94 E Caprolactam 105 - 60 - 2 80 Fig. 6 TIC chromatogram in headspace-GC-MS, along with candidate compounds from library search for major peaks. Top right: Expanded view of NMP elution position and the m/z 99 MS chromatogram. Fig. 7 Comparison of chromatograms and drying conditions for five different NCM cathodes with different drying processes Fig.8 Comparison of residual solvent levels from five different NCM cathodes with different drying processes (n=3 average) Table 5 Similarity search results Summary
Related Products Some products may be updated to newer models. Brevis™ GC-2050 Gas Chromatograph GCMS-QP2050 Gas Chromatograph Mass Spectrometer SH-I-624Sil MS Related Solutions Electronics Electronic Component Clean Energy Lithium-ion Battery Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)