Application
News
www.shimadzu.com/an/
Shimadzu Corporation
© Shimadzu Corporation, 2024
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
See http://www.shimadzu.com/about/trademarks/index.html for details.
Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not
they are used with trademark symbol “TM” or “”.
Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
to change without notice.
First Edition: Nov. 2024 01-00803-EN
X-ray Fluorescence Spectra of Measured
Elements
Conclusion
ALTRACE is a trademark of Shimadzu Corporation or its affiliated companies in Japan and/or other countries.
Fig. 3 shows a comparison of the X-ray fluorescence spectra of the
1 ppm standard sample and blanks of As and Cd acquired with
ALTRACE and EDX-7000 (excerpted from Application News No.
X260). In the case of As, a clear difference between the added
samples and the blank can be detected by both instruments, but
for Cd, only ALTRACE can detect a clear difference in the peaks.
The greatly improved sensitivity of ALTRACE can be understood
from the peak shape, particularly in the measurements of Cd, Sn
and other high energy band elements.
Excellent results were obtained for both the calibration curves
and the analysis results of powdered milk. EDXRF is useful for
production process and quality control, as it is possible to
measure samples directly in powder form or with only simple
sample preparation, and thanks to its simple instrument
handling and excellent repeatability, there are virtually no
individual differences in results and judgments.
The sensitivity of ALTRACE for heavy metal elements was
substantially improved by installing a high power X-ray tube,
enabling highly accurate management in a shorter time,
thereby contributing to improved throughput.
<Related Application News>
1. Screening Analysis of Trace Heavy Elements in Powdered Milk by
EDXRF, Application News No.X260
Limit of Detection
Table 5 shows the limits of detection calculated from the
theoretical statistical variation. The data for the EDX-7000 are
excerpts from Application News No. X260. Sensitivity equal to or
higher than that of the EDX-7000 can be obtained with ALTRACE,
even when the measurement time is greatly shortened to about
1/10 of the conventional level.
Table 3 Accuracy of Calibration Curves [ppm]
Element
33As 80Hg 82Pb 48Cd 50Sn
Accuracy 0.08 0.08 0.11 0.12 0.11
ALTRACE
Fig. 3 X-ray Fluorescence Spectra (Blue: As, Cd 1 ppm, Red: Blank)
Measurement time: ALTRACE (As 600 s, Cd 600 s)
EDX-7000 (As 3600 s, Cd 7200 s)
Table 4 shows the analysis conditions.
Analysis Conditions
Table 4 Analysis Conditions
Instrument ALTRACE
Elements As, Cd, Sn, Hg, Pb
Analysis group Calibration curve method
Detector SDD
X-ray tube Rh target
Tube voltage
50 [kV] (As, Hg, Pb)
65 [kV] (Cd, Sn)
Tube current Auto [μA]
Primary filter #5 (As, Hg, Pb), #1 (Cd, Sn)
Atmosphere Air
Integration time 600 [s] x 2 Ch
Dead time Max. 40 [%]
EDX-7000
Table 5 Limits of Detection [ppm]
Element
33As 80Hg 82Pb 48Cd 50Sn
Analytical line AsKα HgLα PbLβ1 CdKα SnKα
Instrument ALTRACE
LOD
*1
0.029 0.069 0.059 0.075 0.131
Measurement time 600 s 600 s
Instrument EDX-7000
LOD 0.047 0.069 0.074 0.237 0.573
Measurement time 3600 s 7200 s
*1 LOD: Limit of detection; 3 times the theoretical statistical variation
obtained from the calibration curve.
Results of Screening Analysis
Assuming a solution prepared by dissolving 13 g of powdered
milk in hot water to make up a total of 100 g is used as the
standard concentration when ingested, the value obtained by
multiplying the standard values in Table 1 by the dilution factor
of 7.69 (100÷13) is set as the reference value in the powder
form. The result is “OK” if the total of the quantitative value and
error (judgment value) is less than the reference value.
Table 6 shows the result of a screening analysis of sample 4 in
Table 2. From this result, it can be understood that determination
at the 1 ppm level is sufficiently possible with ALTRACE.
Table 6 Result of Screening Analysis of Standard Sample (1 ppm)
Heavy metal, etc.
(analyte element)
Arsenic
(As)
Total
mercury
(Hg)
Lead
(Pb)
Cadmium
(Cd)
Tin
(Sn)
Judgment OK Not OK OK OK OK
①
Quantitative value 1.00 1.01 0.97 1.02 1.00
②
Standard deviation σm 0.020 0.020 0.017 0.022 0.044
③
Calibration curve accuracy σc 0.08 0.08 0.11 0.12 0.11
④
Error 2
×
√(
②
2
+
③
2
) 0.158 0.162 0.215 0.237 0.232
⑤
Determination value
①
+
④
1.16 1.17 1.19 1.26 1.23
⑥
Reference value 3.8 0.76 2.3 1.5 77
Note: Error was obtained using the following expression:
Error = × σ
2
+ σ
2
: With coverage factor set to 2
σ
: Calibration curve accuracy
σ
: Measurement repeatability
(standard deviation)
Measurement time: 600 s Measurement time: 3600 s
Measurement time: 600 s Measurement time: 7200 s