User Benefits Application News Fig. 1 Polarity-Extended Liquid Chromatography-Mass Spectrometry System Take 50 mL of environmental water sample into a 50 mL centrifuge tube, add 50 mg of EDTA, mix well, and place it in a - 80°C refrigerator for overnight freezing. Put the sample in a freeze dryer and dry it in vacuum until the sample is evaporated to dryness. Add 10 mL of methanol (containing 0.1% formic acid) solution and shake for 2 minutes. Blow the sample to nearly dry with nitrogen at room temperature. Add 1 mL of 50% methanol aqueous solution, shake for 2 minutes, pass through a 0.22 µm filter membrane, and measure on the instrument. Sample Preparation Introduction Polarity-Extended Liquid Chromatography combines hydrophilic interaction chromatography and reversed-phase chromatography to achieve simultaneous analysis of polar and weakly polar compounds. This method can detect 501 emerging contaminants, offering a broad scope in terms of both quantity and variety. The water sample is concentrated using freeze-drying method, effectively recovering both polar and non-polar compounds are recovered well. Rapid Screening of 501 Emerging Contaminants in Water Using Polarity-Extended Liquid Chromatography-Mass Spectrometry Lisong CHEN Polarity-Extended Liquid Chromatography; Nexera XS Liquid Chromatograph Mass Spectrometry LCMS-8050 Series Shimadzu(China) Co.LTD In recent years, water pollution has become increasingly serious, and the problem of emerging pollutants in water has attracted much attention. Traditional analytical methods for emerging pollutants mainly focus on 1-2 categories of emerging pollutants and pay less attention to strongly polar pollutants. The Polarity-Extended Liquid Chromatography, through patented technologies such as polarity splitting, online dilution, and dual gradients, combines a Hilic column (first dimension) and a C18 column (second dimension), which can broaden the polarity analysis range of compounds and achieve simultaneous analysis of polar and weakly polar compounds. Fig. 1 shows the Polarity-Extended Liquid Chromatography-Mass Spectrometry System. In this paper, a method for rapid screening of 501 emerging pollutants was established by using Shimadzu's Polarity- Extended Liquid Chromatography-Mass Spectrometry System. This method solves the problem of simultaneous analysis of strong and weak polar pollutants. System : Polarity-Extended Liquid Chromatography Columns : 1 st dimention: Waters Cortecs HILIC(150 mm × 2.1 mm I.D., 2.7 μm) 2 nd dimention: Shim-pack GIST-HP C18-AQ ( 150 mm × 2.1 mm I.D., 3 μm ) Mobile phase : A: 0.1% FA, B: Acetonitrile(1 st dimention), A: 2mM ammonium acetate aqueous solution (containing 0.01% formic acid), B: Acetonitrile(2 nd dimention), Column Temparature : 40 ˚C Injection volume : 5 μL FCV Valve Position : 0 (0 min) → 1 (1.6 min) →0 (10 min) System : LCMS-8050 Series (ESI) Nebulizing gas : 3 L/min Drying gas : 10 L/min Heating gas : 10 L/min DL temp : 200 ˚C Heat block temp : 400 ˚C Interface temp : 300 ˚C Table 1 Analysis Conditions of PELC and LCMS-8050 Series Table 2 Time program of 1 st dimention liquid chromatography Time(min) Unit B% Flow rate(mL/min) 0.00 Pump 97 0.4 2.50 Pump 97 0.4 3.50 Pump 60 0.4 6.50 Pump 55 0.4 6.80 Pump 35 0.4 9.50 Pump 20 0.4 9.60 Pump 97 0.4 28.00 Controller Stop Fig. 2 Flow diagram of Polarity-Extended Liquid Chromatography System 1) Analytical Condition The analytical conditions for HPLC and MS are shown in Table 1. The time program of two dimentions are shown in Table 2 and 3. Flow diagram of Polarity-Extended Liquid Chromatography System are shown in Fig. 2. Table 3 Time program of 2 nd dimention liquid chromatography Time(min) Unit B% Flow rate(mL/min) 0.00 Pump 5 0.3 1.70 Pump 5 0.3 1.80 Pump 5 1.5 9.70 Pump 5 1.5 9.80 Pump 5 0.3 10.00 Pump 5 0.3 10.20 Pump 45 0.3 22.50 Pump 65 0.3 24.50 Pump 95 0.3 26.50 Pump 95 0.3 26.60 Pump 5 0.3 28.00 Controller Stop
Application News 97% 2% 1% 0~0.3% 0.3~0.5% 0.5~1% 95% 3% 2% 0~0.3% 0.3~0.5% 0.5~1% Fig. 8 RSD Distribution Diagram of Retention time for Repeated Injections (Left: High Concentration, Right: Low Concentration) Recovery A mixture of emerging pollutant standard substances was added to pure water samples to investigate the recovery rate during the screening process. Under the current pretreatment method, 76% of the compounds have a recovery rate better than 50%. 84% 13% 3% 0~5% 5~10% 10~20% 79% 16% 5% 0~5% 5~10% 10~20% Fig. 7 RSD Distribution Diagram of Peak Area for Repeated Injections (Left: High Concentration, Right: Low Concentration) Chromatograms of 501 compounds Fig. 3 Chromatograms of 501 compounds Retention of strongly polar compounds Traditional analytical ideas for emerging pollutants use C18 columns for analysis. However, the separation ability of C18 columns for strongly polar target compounds is relatively weak. These substances hardly have any retention on C18 columns, and problems such as solvent effect and matrix effect are prone to occur. In this method, a Hilic column is first used, which has a better retention and separation effect on strongly polar compounds. Then, medium-polar and weak-polar compounds are introduced into the C18 column in the second dimension for separation, realizing the full separation of strongly polar to weak - polar compounds. Fig. 4 shows MRM of some strongly polar compounds. Fig. 4 MRM of some strongly polar compounds. Characteristics of PELC in this topic 3.40e6 Q 122.10>58.10 (+) 保留时间 (min) 5.0 5.5 6.0 6.5 0.0e0 2.0e6 4.74e6 Q 586.30>163.30 (+) 保留时间 (min) 7.5 8.0 8.5 9.0 0.0e0 2.0e6 4.0e6 chlormequat amikacin Coverage of types of emerging pollutants This method collects 501 emerging pollutants, including PFASs, PPCPs, hormones, preservatives, insecticides, fungicides, organophosphorus flame retardants, anilines, herbicides, plant growth regulators, etc. The propotion of number of each type of pollutants in this method is listed in Fig. 5. 1, 10% 2, 22% 3, 22% 4, 1% 5, 10% 6, 2% 7, 17% 8, 11% 9, 2% 10, 3% 1、PFAS 2、PPCPs 3、Hormones 4、Flame retardant 5、Herbicide 6、Preservatives 7、Pesticide 8、Fungicide 9、Anilines 10、Plant growth regulator Fig. 5 The proportion of the number of each type of pollutant in this method 1.78e5 245:TIC(-) 保留时间 (min) 0 2 4 6 8 10 12 14 16 18 20 22 24 26 0.0e0 5.0e5 1.0e6 1.5e6 2.0e6 2.5e6 3.0e6 3.5e6 4.0e6 1.24e6 113:TIC(+) 保留时间 (min) 0 2 4 6 8 10 12 14 16 18 20 22 24 26 0.0e0 5.0e5 1.0e6 1.5e6 2.0e6 2.5e6 3.0e6 3.5e6 4.0e6 4.5e6 5.0e6 5.5e6 6.0e6 1.24e6 113:TIC(+) 保留时间 (min) 4.50 4.75 5.00 5.25 5.50 5.75 6.00 6.25 6.50 6.75 7.00 7.25 7.50 7.75 8.00 8.25 8.50 8.75 9.00 9.25 9.50 9.75 0.0e0 5.0e5 1.0e6 1.5e6 2.0e6 2.5e6 3.0e6 3.5e6 4.0e6 PFASs Fig. 6 Chromatogram of PFASs and PPCPs PPCPs Method validation Linearity External standard method was used for quantification. The linear correlation coefficient R of the calibration curves of 501 compounds was ≥0.994. Repeatability Two concentration levels of high and low concentration points were selected for repeatability investigation. Six consecutive injections were made. For the vast majority of compounds, the RSD of peak area was less than 10%, and the RSD of retention time was less than 0.3%. Detailed results are shown in Fig. 7 and Fig. 8. Fig. 3 shows the chromatograms of 501 compounds in this method. Typical chromatograms of high concern substances This method includes 46 PFASs and 112 PPCPs. The chromatograms of PFASs and PPCPs are as follows. Fig. 6 shows the chromatograms of PFASs and PPCPs.
Application News www.shimadzu.com/an/ Shimadzu Corporation Copyright © 2024 Shimadzu Corporation and/or its affiliates. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. Shimadzu (China) Co., Ltd www.shimadzu.com.cn First Edition: Dec. 2024 03-LCMSMS-919-EN Conclusion In this study, a method for detecting emerging pollutants in water by combining the Polarity-Extended Liquid Chromatography and LCMS - 8050 series coupling system was established. This method combines hydrophilic interaction chromatography and reversed - phase chromatography to achieve the simultaneous separation of polar and non - polar compounds with a single injection. The detection of 501 emerging pollutants was achieved within 28 minutes. The linearity and repeatability of the method meet the quantitative requirements. After pretreatment operations such as freeze - drying, re - dissolution, and nitrogen blowing of the samples, the recovery rates for most compounds were satisfactory. This method has been successfully used for the rapid screening of new pollutants in tap water and sewage. The combination of Polarity-Extended Liquid Chromatography and mass spectrometry enables the one-injection analysis of compounds with different polarities without significantly increasing analysis time, making it suitable for the rapid screening of emerging pollutants in water. Analysis of actual samples Tap water and sewage were selected respectively. After the water samples were processed, they were analyzed on the instrument. The screening results in Fig. 9 show that pollutants were detected in both tap water and sewage samples. 5.49e3 Q 207.00>89.00 (+) 保留时间 (min) 1.5 2.0 0.0e0 2.0e3 4.0e3 5.04e5 Q 183.10>98.80 (+) 保留时间 (min) 12.5 13.0 13.5 14.0 0.0e0 2.0e5 4.0e5 4.36e4 Q 292.00>211.10 (+) 保留时间 (min) 12.0 12.5 13.0 0.0e0 2.0e4 4.0e4 7.34e3 Q 250.00>169.10 (+) 保留时间 (min) 12.0 12.5 13.0 13.5 0.0e0 5.0e3 1.99e4 Q 223.10>126.05 (+) 保留时间 (min) 12.5 13.0 13.5 0.0e0 1.0e4 2.0e4 5.73e3 Q 190.00>163.00 (+) 保留时间 (min) 12.5 13.0 13.5 0.0e0 2.5e3 5.0e3 Fig. 9 Pollutants detected in tap water and sewage aldicarb Triethyl phosphate thiamethoxam clothianidin acetamiprid tricyclazole 2.27e5 Q 122.10>58.10 (+) 保留时间 (min) 5.0 5.5 6.0 0.0e0 1.0e5 2.0e5 chlormequat chloride 1.85e5 Q 237.00>194.05 (+) 保留时间 (min) 13.0 13.5 14.0 14.5 0.0e0 1.0e5 carbamazepine 8.75e4 Q 278.85>186.00 (+) 保留时间 (min) 12.0 12.5 13.0 13.5 0.0e0 5.0e4 Sulfamethazine <References> 1) Rui Guo, Qisheng Zhong, Jiaqi Liu, Peiming Bai, Zongpeng Wang, Jieling Kou, Polarity-extended liquid chromatography- triple quadrupole mass spectrometry for simultaneous hydrophilic and hydrophobic metabolite analysis. Analytica Chimica Acta, Volume 1277, 2023, 341655 Nexera is a trademark of Shimadzu Corporation or its affiliated companies in Japan and/or other countries. Retention Time Retention Time Retention Time Retention Time Retention Time Retention Time Retention Time Retention Time Retention Time
Please fill out the survey Related Products Some products may be updated to newer models. Nexera series Ultra High Performance Liquid Chromatograph LCMS-8050 High Performance Liquid Chromatograph Triple Quadr… Shim-pack GIST Series HPLC Column Solutions Related Solutions Environment PFAS Water Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)