User Benefits
Application
News
Hinoki (Japanese cedar) has been used as a timber alongside
cedar for a long time. It is easy to be processed and highly durable,
making it an important resource in construction. Additionally,
Hinoki contains various aromatic components, suggesting the
potential for people to feel a sense of preference and comfort in
wooden structures. Shimadzu Corporation is involved in Kyoto
Prefecture Model Forest Movement, contributing to the
maintenance of "Shimadzu Corporation Forest" and aiming to
contribute to the community and preserve biodiversity. The
Hinoki oil and diffuser made from thinned wood from this forest
won the Brands & Communication Design category of the
prestigious international design award "Red Dot Design Award
2021" in Germany. This is the first time our company has won in
this category.
In this application, five types of Hinoki oils, including Hinoki oil
made by our company and Hinoki oils from different regions,
were analyzed using the detector splitting system between MS
and FID. The results from MS were compared using eMSTAT
Solution to analyze the composition of the components.
Additionally, the quantification of δ-Cadinene, an important
aromatic component that is a sesquiterpene was conducted using
FID, which showed high calibration linearity over a wide
concentration range.
Furthermore, with a compact width of approximately 63 cm, the
device configuration of Brevis GC-2050 + GCMS-QP2050 not only
contributes to space-saving in the lab but also demonstrates high
analytical performance.
Introduction
The detector splitting system allows for simultaneous acquisition of MS and FID data in a single measurement.
Even complex samples can undergo simplified statistical analysis using the eMSTAT Solution
TM
.
The combination of the compact Brevis
TM
GC-2050 and GCMS-QP2050 enables space-saving in the laboratory.
Analysis of Essential Oil Using
GC-MS/FID Detector Splitting System
Kei Usui
GC-MS GCMS-QP2050
Table 1 System Configuration
In quality control (QC) of fragrances like Hinoki oil, where a wide
concentration range is typically expected, GC-FID, which provides
detector response over a broad range, is commonly used. By
comparing the chromatogram shapes of product samples
obtained during the production process with approved reference
samples, the component profile is confirmed. However, in the
case of Hinoki oil containing numerous components, qualitative
analysis of each peak becomes challenging with an FID detector.
Therefore, by using the detector splitting system between MS and
FID, it is possible to obtain data from both MS and FID
simultaneously in a single measurement. This allows for the usual
confirmation of chromatogram shapes with FID and enables
qualitative analysis with MS when unknown components are
detected.
Essential Oil Analysis Using GC-MS/FID
Column
FID
GC Model : Brevis GC-2050 / AOC
TM
-30i
MS Model : GCMS-QP2050 (TMP exhaust: 255 L/sec)
Injection Port : SPL
Column : SH-I-5Sil MS (P/N
:
221-75954-30)
(30 m
×
0.25 mm I.D., df= 0.25 µm)
Detector splitting : SMI FLOW DEVICE 2-Way Splitter with APC
(P/N : 221-88100-43)
Software : Labsolutions
TM
GCMS
eMSTAT Solution
GC
Injection Temperature : 250℃
Flow Control Mode : Pressure (He 230 kPa )
Initial Flow Rate : 3.24 mL/min
Purge Flow : 5 mL/min
Injection Volume : 0.8 µL
Split ratio : 10
Column Oven Temp. Program : 80℃→ 5℃/min → 230℃
AUX-APC : He 40 kPa
Resistance tube : 1.7 m
×
0.15 mm I.D. (MS)
2.0 m
×
0.25 mm I.D. (FID)
Split ratio : MS: FID of 1:3.2
Initial Flow Rate : 1.66 mL/min (MS)
5.29 mL/min (FID)
MS
Ion Source Temperature : 230℃
Interface Temperature : 300℃
Measurement Mode : Scan
Scan Range (m/z) : 45-500
Ionization Mode : EI
Event Time : 0.3 s
FID
Detector Temperature : 300℃
Makeup Gas : N2 24 mL/min
Detector Gas : H2 32 mL/min
、
Air 200 mL/min
AUX-APC
MS
Detector splitting
Table 2 Analysis Conditions
Fig. 1 Detector Splitting System
The system configuration and analysis conditions are shown in
Table 1 and 2, respectively In this analysis, SMI (Smart Micro Inert)
FLOW DEVICE 2-Way Splitter for detector splitting was used (Fig.
1). The SMI FLOW DEVICE is a capillary flow device for analysis
with features such as low leakage, low dead volume, low heat
capacity, and inertness. This allows for the analysis of highly
adsorptive components. We adjusted the inner diameter of the
resistors used for MS/FID detector Splitting, as well as the pressure
of the AUX-APC (Digital Flow Controller), to set the split ratio at
approximately 1:3 for MS:FID. By using eMSTAT Solution for
analysis, we visually grasped the differences in component
characteristics of each Hinoki oil and identified the compounds
contributing to those differences.
System Configuration and Analysis
Conditions