Application
News
www.shimadzu.com/an/
Shimadzu Corporation
© Shimadzu Corporation, 2025
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
See http://www.shimadzu.com/about/trademarks/index.html for details.
Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not
they are used with trademark symbol “TM” or “”.
Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
to change without notice.
0
1
2
3
4
5
6
7
8
9
10
0 10 20 30 40 50 60 70 80 90 100
Force (N)
Displacement (mm)
Room Temperature
50℃
70℃
First Edition: Jul. 2025 01-00944-EN
AGS and TRAPEZIUM are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries.
Test Results
Fig. 4 shows the force-displacement curve. The dash-dotted
curve is at room temperature, dotted curve is at 50 °C, and real
curve is at 70 °C. The tests were conducted three times at each
temperature whereas the shown curve is one representative
specimen of the three.
The average values of adhesion force are shown in Table 3.
Adhesion forces were calculated from the average force
between 20 mm and 80 mm of the displacement.
These results show that when the temperature gets higher,
adhesion force becomes larger.
Conclusion
Peel tests in a thermostatic chamber were conducted using an
AGS-V and TCE-N300A for Lithium iron phosphate used as
cathode sheet of lithium-ion batteries.
From the test results, it was confirmed that the adhesion force
between the cathode active material and the aluminum foil
increased as the temperature increased.
This measurement can easily evaluate the temperature
dependence of mechanical characteristics of battery cathodes.
This method is expected to be applied to solve problems for
improving battery performance.
Fig. 4 Test Results
Table 3 Adhesion Strength at each temperature (Average of n = 3)
Temperature Adhesion Strength (N)
Room Temperature 0.753
50 °C 0.903
70 °C 8.07
Range for calculating
test force average
Table 2 Test Conditions
Cathode Active Material: Lithium Iron Phosphate
Temperature: Room temp., 50 °C, 70 °C
Test Speed: 5 mm/sec
Disp. Origin: Force 0.05 N
Target Value: Stroke 100 mm
Number of Samples: 3
Test Conditions
Table 2 shows test conditions. In this test, the force and its
temperature dependence were evaluated. The temperature
conditions included room temperature, 50 °C, and 70 °C.
50 °C
70 °C