Supercritical Fluid Chromatography (SFC) is a separation
technique that employs supercritical carbon dioxide as the
primary mobile phase. Owing to its high diffusion coefficient and
low viscosity, supercritical carbon dioxide allows high flow-rate
delivery at lower column backpressure compared to Liquid
Chromatography (LC), enabling shorter analysis time. It also
provides excellent separation performance for structurally related
compounds. However, since supercritical carbon dioxide has a
polarity similar to that of hexane, it often lacks sufficient elution
strength on its own. Therefore, organic solvents (modifiers) such
as methanol, acetonitrile, 2-propanol, and ethanol are commonly
added to the mobile phase to adjust polarity. By appropriately
selecting and combining these modifiers, the interaction
between analytes and the stationary phase can be finely tuned,
resulting in diverse separation patterns. Because SFC and LC
exhibit different retention behaviors, compounds that are difficult
to separate by LC may be effectively separated by SFC.
Accordingly, combining both techniques in method development
can improve the efficiency of searching for optimal separation
conditions. On the other hand, both column and mobile phase
selection have a strong influence on retention behavior. To obtain
optimal separation, it is ideal to examine multiple types of
columns and mobile phases. However, this process is highly labor-
and time-intensive. This study demonstrates efficient
optimization of separation conditions using Nexera UC/s
(LC/SFC switching system), which allows LC and SFC analyses to
Introduction
Because LC and SFC have different separation characteristics, SFC has the potential to separate compounds that are difficult to
resolve by LC.
Both LC and SFC analyses can be performed on a single system.
By utilizing mobile phase and column switching valves, LabSolutions MD enables automated screening of mobile phases and
columns.
Diversification of Separation Selectivity Using
Supercritical Fluid Chromatography
-Efficient Search for Optimal Separation Conditions by Combining LC and SFC-
Shinichi Fujisaki
Supercritical Fluid Chromatography System / Ultra High Performance Liquid Chromatograph
be performed on a single instrument, in combination with the
LabSolutions MD, a dedicated software for supporting method
development. Column and mobile phase screening were
conducted using a mixed sample containing six small-molecule
pharmaceutical compounds as a model sample.
Nexera UC/s enables both LC and SFC analyses on a single system
(flow diagram: Fig. 1). The system is configured by adding a
supercritical CO₂ delivery pump and backpressure regulator unit
to a standard LC system. By switching the ON/OFF status of the
delivery pump and backpressure, both LC and SFC analyses can
be carried out. The organic solvent delivery pumps, autosampler,
column oven, and detector can be shared between LC and SFC,
enabling space-saving and cost-effective operation. Existing LC
systems can also be upgraded. In combination with a PDA
(photodiode array) detector and the single quadrupole mass
spectrometer LCMS-2050, the system allows obtained mass data
to be used for identification of known compounds and estimation
of unknown impurities. When controlled by LabSolutions MD, up
to 12 columns and 7 types of organic solvents (modifiers) can be
automatically switched, enabling comprehensive screening for
both LC and SFC and significantly reducing labor in method
development.
Features of LC/SFC Switching System
Fig. 1 LC/SFC Switching System (Nexera UC/s)
(Green: SFC flow path, Blue: LC flow path, Orange: Common flow path for LC and SFC)
CO
2
pump
(SFC)
CO₂ cylinder
Organic solvents
pump (LC/SFC) Autosampler
Column oven A (SFC)
PDA
detector
Back pressure
regulator
Aqueous solvents
pump (LC) Column oven B (LC)
Make-up
pump
4 solvents
switching valve
7 solvents
switching valve
Single MS
Shut-off
valve
Manifold
Column
switching valve
Column
switching valve
Manifold