User Benefits Application News The AI algorithm of LabSolutions MD can automatically optimize separation conditions, greatly reducing the workload involved in LC method development. By optimizing the separation under isocratic elution, method transfer between different systems becomes easier. Automatic optimization under isocratic elution can efficiently convert existing gradient methods to isocratic ones. Automatic Optimization of Separation Conditions by AI Algorithm -Optimization under Isocratic Elution- Shinichi Fujisaki Software for Efficient Method Development Ultra High Performance Liquid Chromatograph In the typical LC method development, the process begins with “preparation” which includes mobile phase preparation, column installation, and creation of analysis schedules, then analysis is started. After that, the acquired data is analyzed and “preparation” for the subsequent analysis is carried out, followed by starting the next analysis again. The method development progresses by repeating these processes, but in addition to the significant time required to repeatedly create analysis schedules, expertise in chromatography is necessary to explore optimal conditions based on data analysis. In other words, typical method development requires “human intervention”. Therefore, eliminating human involvement and automating such method development processes would be desirable to improve labor efficiency. In this article, a case study is presented in which a mixture of low-molecular-weight compounds was used as a model sample, and conditions meeting the resolution criteria were automatically explored using the dedicated software for supporting method development “LabSolutions MD”. For gradient elution, method transfer between different systems while maintaining the separation pattern requires appropriate adjustment of various LC parameters, and achieving an exact match of the separation pattern before and after transfer can be challenging. In this article, optimization of separation conditions under isocratic elution was carried out to facilitate smoother method transfer. Introduction Analytical conditions and target compounds are listed in Table 1. In this study, a mixture of five low-molecular-weight compounds was used as a model sample, prepared in two patterns with different combinations of compounds. Resolution and final peak elution time criteria were set for these samples, and separation conditions meeting these criteria were automatically explored using LabSolutions MD. Analytical Conditions and Target Compounds AI algorithm repeatedly searches for optimal conditions Fig. 1 Workflow for Automatic Optimization of Separation Conditions by LabSolutions MD Initial setting Initial analysis Condition search Correction analysis Conditions optimized Fig. 1 shows the workflow for automatic optimization of separation conditions using LabSolutions MD. This software is equipped with a unique AI algorithm that automatically explores conditions satisfying the set criteria by iteratively performing “improvement of separation conditions by AI (condition search)” and “analysis under the improved conditions (correction analysis).” The criteria can be set for the resolution of any peak and the elution time of the last peak. In this study, the minimum resolution was set to 1.5 and the elution time of the last peak to less than 5 minutes, and separation conditions meeting these criteria were automatically explored, taking into account the reduction of analysis time. Optimization was performed under isocratic elution (Fig. 2). Automatic Optimization of Separation Conditions Table 1 Analytical Conditions and Target Compounds Mobile phase Pump A : 0.1% formic acid in water Pump B : Acetonitrile Sample 1 : (A) Antipyrine, (B) Benzoic acid, (C) Salicylic acid, (D) Hydrocortisone, (E) Furosemide Sample 2 : (A) Antipyrine, (B) Benzoic acid, (C) Furosemide, (D) Naproxen, (E) Probenecid System : Nexera TM X3 Column : Shim-pack Scepter TM C18-120 *1 (100 mm × 3.0 mm I.D., 1.9 µm) Analytical conditions B Conc. :5%(0 min)→80%(X~X+1 *2 min) →5%(X+1~X+5 min) Column Temp. :40 °C Flow rate :0.7 mL/min Injection Vol. :5 µL Detection :254 nm (SPD-M40, STD cell) *1 P/N : 227-31013-03 *2 X = 2, 3, 4, 5, 6 (for sample 1), X = 3, 4, 5, 6, 7 (for sample 2) Criteria of automatic optimization Minimal resolution :1.5 Time of last eluting peak :< 5 min Gradient mode :Isocratic Fig. 2 Setting for Automatic Optimization
Application News www.shimadzu.com/an/ Shimadzu Corporation © Shimadzu Corporation, 2025 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. First Edition: Nov. 2025 01-01087-EN Nexera, LabSolutions and Shim-pack Scepter are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries. The results of automatic optimization of separation conditions are shown in Fig. 3. After initial analyses were performed using five gradient conditions (Table 1), the AI algorithm automatically optimized the separation of all compounds through correction analyses. For Sample 1, isocratic elution at 33% enabled all compounds to achieve a resolution of 1.5 or higher while keeping the retention time of the last peak (compound E) within 5 minutes. Similarly, for Sample 2, isocratic elution at 45% allowed all compounds to reach a resolution of 1.5 or higher, with the retention time of the last peak (compound E) also within 5 minutes. Automatic optimization of separation conditions under isocratic elution using the AI algorithm of LabSolutions MD was applied to a mixture of low-molecular-weight compounds as a model sample. As a result, conditions meeting the criteria (“minimum resolution: 1.5” and “elution time of the last peak: less than 5 minutes”) were successfully explored automatically. By using LabSolutions MD, separation conditions satisfying the set criteria can be automatically identified, which is expected to significantly reduce the workload of method development. In addition, optimization under isocratic elution facilitates method transfer between different systems. While this article introduces a case of automatic optimization of separation conditions, LabSolutions MD also supports the entire workflow of method development, including the screening and robustness evaluation phases. For further details, please refer to the Technical Report “Efficient Method Development Based on Analytical Quality by Design with LabSolutions MD Software (C190-E284)”. Conclusion Fig. 3 Results of Automatic Optimization for Sample 1 (left) and Sample 2 (right) (The blue line indicates the gradient conditions) B.Conc B.Conc B.Conc B.Conc B.Conc B.Conc B.Conc B.Conc B.Conc B.Conc B.Conc = 33% AU 0 - 0.4 - 0 - 0.4 - 0 - 0.4 - 0 - 0.4 - 0 - 0.4 - AU 0 - 0.4 - 0 - 0.4 - 0 - 0.4 - 0 - 0.4 - 0 - 0.4 - % -100 - 50 -100 - 50 -100 - 50 -100 - 50 -100 - 50 % -100 - 50 -100 - 50 -100 - 50 -100 - 50 -100 - 50 3 min 5 2 4 4 min 2 6 AU 0 - 0.4 - AU 0 - 0.4 - 2 min 6 0 4 2 min 6 0 4 - 0 - 50 % - 100 - 0 - 50 % - 100 Initial Analysis Initial Analysis Condition Search & Correction Analysis ( Optimized under isocratic elution ) 1 5 3 1 5 3 7 7 B.Conc = 45% (A) (B) *11.1 (C) *3.4 (D) *2.7 (E) *9.2 (A) (B) *6.8 (C) *4.7 (D) *18.5 (E) *6.8 (A) Antipyrine, (B) Benzoic acid, (C) Furosemide, (D) Naproxen, (E) Probenecid Automatic Optimization for Sample 1 Automatic Optimization for Sample 2 * Resolution * Resolution Condition Search & Correction Analysis ( Optimized under isocratic elution ) (A) Antipyrine, (B) Benzoic acid, (C) Salicylic acid, (D) Hydrocortisone, (E) Furosemide
Please fill out the survey Related Products Some products may be updated to newer models. Method Development System Automatic Optimization of Gradient Conditions with… Nexera Series Ultra High Performance Liquid C… Ultra High Performance Liquid Chromatograph Related Solutions Pharmaceutical and Biopharmaceutical Small Molecule Pharmaceutical Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)