User Benefits Application News Hydrogen is expected to be a new power and heat source to reduce greenhouse gases. Hydrogen is an easily available substance that can be produced from the electrolysis of water in our daily lives. However, when used in fuel cells, it is necessary to control impurities in hydrogen gas because certain impurities introduced during the manufacturing process deteriorate the electrode and electrolyte membrane, reducing the performance of the fuel cell or shortening its life. This section presents an example of analysis of inorganic halogens, formic acid, ammonia, and formaldehyde in hydrogen based on the standard values described in ISO 14687 Grade D 1) . Introduction This analysis meets the standard values of ISO14687 Grade D. Simultaneous analysis of inorganic halogens and formic acid is possible. Since organic solvents are not used for extraction during pretreatment, environmental impact is reduced. Analysis of Inorganic Halogens, Formic Acid, Ammonia, and Formaldehyde as Impurities in Fuel Cell Grade Hydrogen Based on ISO14687 Grade D Hiroyuki Aikawa, Jiang Yujing, Ayano Tanabe Ion Chromatograph, High Performance Liquid Chromatograph ISO14687 Grade D specifies impurities in hydrogen that affect the quality of fuel cell vehicles. Table 1 lists the applicable ingredients, standard values, concentrations (μg/L) converted by ion chromatograph (IC) / high-performance liquid chromatograph (HPLC), and equipment used. Hydrogen Impurity Specifications in ISO14687 Grade D Ingredients of interest Specified value (μmol/mol) IC/HPLC Converted concentration (μg/L) Equipment used Inorganic halogen (hydrogen chloride) *1 0.05 28 Suppressor IC Formic acid *2 0.2 92 Ammonia 0.1 26 Non-suppressor IC Formaldehyde 0.2 141 HPLC Table 1 Substances and specifications in ISO14687 Grade D *1 The value of hydrogen chloride is shown as a representative value. *2 Deleted in the 2025 edition A suppressor IC (Fig. 1, left) was used for inorganic halogen analysis. Since IC can selectively measure ions in a sample, it is less affected by contaminants than other methods, and inorganic halogens can be measured with high accuracy. Three components were simultaneously measured: chloride ion (Cl) as hydrochloric acid, bromide ion (Br) as hydrogen bromide, and formic acid (HCOOH). Table 2 shows the analytical conditions. Fig. 2 shows the chromatogram of each 50 μg/L standard sample. Fig. 3 and Table 3 show the calibration curves for the concentration range of 25 μg/L to 1000 μg/L. As shown in Fig. 3 and Table 3, the linearity was greater than 0.9999 in all cases. Table 4 shows the repeatability for 7 replicates at concentrations near the lower limit of quantification. Analysis of Inorganic Halogens, Formic Acid, and Ammonia by IC Fig. 1 Equipment used Suppressor IC (HIC-ESP) Non-suppressor IC (HIC-NS) HPLC(Nexera lite) System : HIC-ESP Column : Shim-pack IC-SA2 (250 mm × 4.0 mm I.D., 9 µm) *1 Guard Column : Shim-pack IC-SA2(G) (10 mm × 4.6 mm I.D., 9 µm) *2 Mobile Phase : 12 mM NaHCO3 + 0.6 mM Na2CO3 Flow Rate : 0.8 mL/min Column Temp. : 30 ˚C Injection Vol. : 100 μL Vial : 4 mL, Polypropylene *3 Detection : Conductivity Table 2 Analysis conditions (inorganic halogens) Fig. 2 Chromatogram of inorganic halogens (50 μg/L each) Fig. 3 Calibration curves for inorganic halogens 0.0 5.0 10.0 min 0 1.0 µS/cm HCOOH Br Cl 0 250 500 750 Concentration (μg/L) 0 5000 10000 15000 20000 25000 30000 Area 0 250 500 750 Concentration (μg/L) 0 10000 20000 30000 40000 50000 60000 70000 Area r 2 = 0.99998 Formic acid Chloride ion r 2 = 0.99994 r 2 = 0.99995 Bromide ion 0 250 500 750 Concentration (μg/L) 0 5000 10000 15000 20000 25000 30000 Area *1 P/N : 228-38983-91, *2 P/N : 228-38983-92, *3 P/N : 228-31537-91 Conc. of Calibration Curves µg/L Peak Area of Formic acid Peak Area of Chloride ion Peak Area of Bromide ion STD1 20 487 1267 501 STD2 50 1373 3192 1392 STD3 100 2822 6449 2831 STD4 500 15410 32837 14289 STD5 1000 31127 66830 29066 correlation coefficient (r 2 ) 0.99998 0.99994 0.99995 Table 3 Concentration of calibration curves and peak area for inorganic halogens
Application News Ammonia was determined as ammonium ion (NH 4 ) using a non- suppressor IC (Fig. 1, center). Table 5 shows the analytical conditions, Fig. 4 shows the chromatogram of a standard sample at a concentration near the specified value (25 μg/L), Fig. 5 and Table 6 show the calibration curve prepared in the concentration range of 2.5 μg/L to 200 μg/L, and Table 7 shows the repeatability at a concentration near the detection limit (10 μg/L). Both inorganic halogens and ammonia can be analyzed in the same apparatus by switching the mobile phase and column. System : HIC-NS Column : Shim-pack IC-C4 (150 mm × 4.6 mm I.D., 7 μm) *1 Guard Column : Shim-pack IC-GC4 (10 mm × 4.6 mm I.D., 7 μm) *2 Mobile Phase : 2.5 mM Methane sulfonic acid Flow Rate : 1.0 mL/min Column Temp. : 40 ˚C Injection Vol. : 100 μL Vial : 4 mL, Polypropylene *3 Detection Conductivity Table 5 Analysis conditions (ammonium ion) Fig. 5 Calibration curve for ammonium ion NH 4 Fig. 4 Chromatogram of ammonium ion (25 μg/L) 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 min 0 1.0 2.0 3.0 4.0 5.0 µS/cm Table 7 Repeatability of ammonium ion (10 μg/L, n=7) Retention time (min) Peak area Concentration (μg/L) 1 4.36 1315 9.99 2 4.36 1355 10.3 3 4.37 1380 10.5 4 4.37 1284 9.74 5 4.37 1292 9.80 6 4.37 1246 9.43 7 4.36 1318 10.0 Average 4.36 1313 9.97 %RSD 0.04 3.41 3.61 Formaldehyde (HCHO) may adsorb on the catalyst of a fuel cell and interfere with hydrogen generation. We used HPLC (Fig.1, right) to measure formaldehyde. DNPH derivatization was performed according to the pretreatment protocol shown in Fig. 6 and measured by HPLC. Table 8 shows the analytical conditions, Fig. 7 shows the chromatogram of a standard sample at a concentration near the specified value (100 μg/L), Fig. 8 and Table 9 shows the calibration curve prepared at a concentration range of 10 μg/L to 50 μg/L, and Table 10 shows the repeatability at a concentration near the lower limit of quantification (10 μg/L). Analysis of Formaldehyde by HPLC System : Nexera lite Column : Shim-pack GIST-HP-C18-HP (150 mm × 3.0mm I.D., 3 μm) *1 Mobile Phase : Water/Acetonitrile = 1 : 1 Flow Rate : 1.0 mL/min Column Temp. : 40 ˚C Injection Vol. : 10 μL Vial : 1.5 mL, Glass *2 Detector : UV 360 nm Table 8 Analytical conditions (formaldehyde) Fig. 7 Chromatogram of formaldehyde (100 μg/L) 0.0 1.0 2.0 3.0 4.0 min 0.0 5.0 10.0 15.0 µS/cm DNPH-HCHO Fig. 6 Pretreatment protocol for DNPH derivatization Fig. 8 Calibration curve for formaldehyde Pretreatment Each Standard & Sample 100 mL 10 mL 0.2 mL Phosphoric acid (1+4) 0.5 mL 1 g/L 2,4-DNPH in Acetonitrile Standing for 20 min HPLC *1 P/N : 228-41616-91, *2 P/N : 228-59900-91, *3 P/N : 228-31537-91 *1 P/N : 227-30040-05, *2 P/N : 227-34500-02 Table 4 Repeatability of inorganic halogens (10 μg/L, n=7) Formic acid Concentration (μg/L) Chloride ion Concentration (μg/L) Bromide ion Concentration (μg/L) 1 10.1 9.48 9.12 2 10.4 9.66 9.33 3 10.6 9.56 8.47 4 10.6 9.48 9.11 5 10.7 9.89 9.90 6 10.8 9.74 8.77 7 11.1 9.68 9.84 Average 10.6 9.60 9.20 %RSD 2.91 1.55 5.68 0.0 25.0 50.0 75.0 100.0 125.0 150.0 175.0 Concentration (µg/L) 0 2500 5000 7500 10000 12500 15000 17500 20000 22500 25000 Area r 2 = 0.9993 Conc. of Calibration Curve µg/L Peak Area STD1 2.5 481 STD2 10 1135 STD3 25 3281 STD4 50 6328 STD5 100 12272 STD6 200 25711 correlation coefficient (r 2 ) 0.9993 Table 6 Concentration of Calibration Curves and Peak Area for ammonium ion 0.0 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0 45.0 Conc. (µg/L) 0 1000 2000 3000 4000 5000 6000 7000 8000 9000 10000 Area r 2 = 0.9991
Application News www.shimadzu.com/an/ Shimadzu Corporation © Shimadzu Corporation, 2025 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See https://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. First Edition: Dec. 2025 Revision A: Mar. 2026 01-01071A-EN As shown in Fig. 9, hydrogen from a G2 grade hydrogen cylinder was bubbled with ultrapure water in an impinger to absorb impurities. Two impingers were installed in the order of 1 st impinger and 2 nd impinger, and dissolved inorganic halogen, formic acid, ammonia, and formaldehyde were analyzed by analyzing water from each impinger. Table 11 shows the collection conditions. Fig. 10 shows the chromatograms of anions (Formic acid, chloride, and bromide), Fig. 11 shows the chromatograms of cations (ammonia), and Fig. 12 shows the chromatograms of formaldehyde. No impurities were detected from either 1 st impinger or 2 nd impinger. Measurement of Hydrogen Gas 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 min 0 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 uS/cm 25 μg/L Standard Solution 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 min -0.1 -0.05 0 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4 µS/cm Each 50 μg/L standard solution Flow Rate : 500 mL/min Sampling Time : 30 min Water Amount in Impinger : 40 mL for each Material of Impinger : Perfluoroalkoxy alkanes 100mL volume Table 11 Conditions for collecting hydrogen gas impurities Fig. 9 Collection of hydrogen gas impurities Fig. 10 Chromatograms of hydrogen gas sample (inorganic halogen) Fig. 11 Chromatograms of hydrogen gas sample (ammonium ion) Impinger 1st 2nd HCOOH Br Cl NH 4 Hydrogen Cylinder Vent Ultrapure Water 0 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 min 0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 mAu Conclusion Using an ion chromatograph and HPLC, we analyzed inorganic halogens, formic acid, ammonia, and formaldehyde, which are impurities in hydrogen, and confirmed that the results meet ISO 14687 Grade D standards. Inorganic halogens, formic acid, and ammonia can be analyzed using the same equipment by switching the analysis conditions. Simple pretreatment and a unified analysis procedure allow measurement without hesitation. Fig. 12 Chromatograms of hydrogen gas sample (formaldehyde) <References> 1) ISO 14687 Hydrogen fuel quality-product specification (2025) <Related Applications> 1. Analysis of Ammonia as An Impurity in Fuel Cell Grade Hydrogen According to ISO 14687 Grade D Using A Gas Generator, Application News No. 01-01070 Blank solution 25 μg/L Standard Solution 2nd Impinger 1st Impinger DNPH-HCHO Nexera and Shim-pack are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries. 2nd Impinger 1st Impinger 2nd Impinger 1st Impinger Conc. of Calibration Curve µg/L Peak Area STD1 10 2947 STD2 20 4772 STD3 50 9523 correlation coefficient (r 2 ) 0.9991 Table 9 Concentration of Calibration Curves and Peak Area for formaldehyde Table 10 Repeatability of formaldehyde (10 μg/L, n=7) Retention time (min) Peak area Concentration (μg/L) 1 3.23 2810 8.65 2 3.23 2877 9.05 3 3.23 2888 9.13 4 3.23 2865 8.98 5 3.23 2892 9.15 6 3.23 2773 8.42 7 3.23 2792 8.54 Average 3.23 2842 8.84 %RSD 0.00 1.73 0.30 Ultrapure Water
Please fill out the survey Related Products Some products may be updated to newer models. Anion Suppressor Ion Chromatograph Nexera Series Ultra High Performance Liquid C… Ultra High Performance Liquid Chromatograph Related Solutions Clean Energy Fuel Cells Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)