Table 3 shows the ICP-AES analysis results of transition metals in
electrolytes. Because the electrolyte samples were measured after
dilution, Table 3 lists both the measured values of the diluted
solutions and the corresponding concentrations in original
electrolytes. The analytical conditions are described later in Table
7.
Figure 6 compares the concentrations of Ni, Mn, and Co in
electrolytes under different cutoff voltages (#1, #2, #3), while
Figure 7 presents comparisons under different ambient
temperatures (#1, #4, #5). For the cutoff voltage conditions, the
detected amounts of transition metals increased as the conditions
became more severe. In contrast, under the different ambient
temperature conditions, the detected amounts were below 1
ppm in all cases.
Electrolyte analysis results (ICP-AES)
Element Ni Mn Co
Wavelength (nm) 231.604 257.610 238.892
Instrument
Quantitation Limit
*1
(10σ) (mg/L)
0.004 0.0002 0.001
Measured value
of diluted
solutions (mg/L)
#1 0.004 0.00024 N.D.
*2
#2 1.18 0.0237 0.264
#3 0.913 0.197 0.215
#4 0.014 0.00191 0.0025
#5 0.004 0.00022 N.D.
Concentration in
electrolytes
*3
(mg/kg)
#1 0.30 0.016 N.D.
#2 179 3.59 40.0
#3 380 82.1 89.6
#4 0.72 0.098 0.13
#5 0.25 0.015 N.D.
Table 3 ICP-AES analysis results of transition metals in electrolytes
*1 Instrument quantitation limit = 10 × σ (standard deviation of the blank
solution) × calibration curve slope.
*2 N.D.: below the quantitation limit.
*3 Concentration in electrolytes = measured value of diluted solutions ×
dilution factor. Refer to Table 7 for dilution factors.
Fig.6 Comparison of Ni, Mn, and Co concentrations in electrolytes
(by cutoff voltage)
Fig.7 Comparison of Ni, Mn, and Co concentrations in electrolytes
(by ambient temperature)
Table 4 lists electrolyte degradation products
2),3)
. Figure 8 shows
example chromatograms of EDMP and DEDOHC detected under
Condition #3. The peak area values of Compounds A–I obtained in
SIM mode are summarized in Table 5. The analysis conditions are
described later in Table 8.
Figures 9 and 10 compare the peak areas of each compound for
different cutoff voltages and different ambient temperatures,
respectively. For the cutoff voltage conditions, the detected
amounts of all compounds increased markedly, and for the
ambient temperature conditions, compounds G, H, and I also
showed an increasing trend.
Electrolyte analysis results (GC-MS)
Symbol Compounds (abb.)
Quantitative
ion (m/z)
Qualitative
ion (m/z)
A
Dimethyl phosphorofluoridate
(DMFP)
90 80
B
Ethyl methyl phosphorofluoridate
(EMPF)
115 127, 97
C Diethyl phosphorofluoridate (DEPF) 101 129, 113
D
Ethyl dimethyl phosphate
(EDMP)
127 110, 109
E
Diethyl methyl phosphate
(DEMP)
141 113, 95
F
Triethyl phosphate
(TEP)
155 127, 99
G
Dimethyl-2,5-dioxahexane
dicarboxylate (DMDOHC)
102 91, 59
H
Ethylmethyl-2,5-dioxahexane
dicarboxylate (EMDOHC)
116 91, 89
I
Diethyl-2,5-dioxahexane
dicarboxylate (DEDOHC)
103 90, 89
12.1 12.2
0.0
0.5
1.0
1.5
(x100,000)
109.00
110.00
127.00
14.1 14.2
0.0
1.0
2.0
3.0
4.0
5.0
(x100,000)
89.05
90.00
103.00
Fig.8 Chromatograms of compounds detected from Condition #3 by SIM-
mode (right: EDMP, left: DEDOHC)
Table 4 Electrolyte degradation products list identified by GC-MS
Symbol #1 #2 #3 #4 #5
A - 21,221 125,609 - -
B - 261,555 1,804,650 - -
C - 371,846 4,745,557 - -
D - 38,925 352,338 - -
E - 6,784 249,905 - -
F - 30,207 612,945 - -
G - 17,494 26,039 331 592
H 2,351 231,407 363,546 3.209 7,179
I 5,168 364,041 503,875 6,885 14,370
Table 5 Peak area values for each compounds